ASTM E1935-1997(2013) Standard Test Method for Calibrating and Measuring CT Density《校准和测量CI密度的标准试验方法》.pdf
《ASTM E1935-1997(2013) Standard Test Method for Calibrating and Measuring CT Density《校准和测量CI密度的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1935-1997(2013) Standard Test Method for Calibrating and Measuring CT Density《校准和测量CI密度的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1935 97 (Reapproved 2013)Standard Test Method forCalibrating and Measuring CT Density1This standard is issued under the fixed designation E1935; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisio
2、n. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers instruction for determining thedensity calibration of X- and -ray computed tomography (CT)systems and for
3、using this information to measure materialdensities from CT images. The calibration is based on anexamination of the CT image of a disk of material withembedded specimens of known composition and density. Themeasured mean CT values of the known standards are deter-mined from an analysis of the image
4、, and their linear attenu-ation coefficients are determined by multiplying their measuredphysical density by their published mass attenuation coeffi-cient. The density calibration is performed by applying a linearregression to the data. Once calibrated, the linear attenuationcoefficient of an unknow
5、n feature in an image can be measuredfrom a determination of its mean CT value. Its density can thenbe extracted from a knowledge of its mass attenuationcoefficient, or one representative of the feature.1.2 CT provides an excellent method of nondestructivelymeasuring density variations, which would
6、be very difficult toquantify otherwise. Density is inherently a volumetric propertyof matter. As the measurement volume shrinks, local materialinhomogeneities become more important; and measured valueswill begin to vary about the bulk density value of the material.1.3 All values are stated in SI uni
7、ts.1.4 This standard does not purport to address the safetyconcerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety andhealth practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents
8、2.1 ASTM Standards:2E1316 Terminology for Nondestructive ExaminationsE1441 Guide for Computed Tomography (CT) ImagingE1570 Practice for Computed Tomographic (CT) Examina-tion3. Terminology3.1 Definitions:3.1.1 The definitions of terms relating to CT, that appear inTerminology E1316 and Guide E1441,
9、shall apply to the termsused in this test method.3.2 Definitions of Terms Specific to This Standard:3.2.1 density calibrationcalibration of a CT system foraccurate representation of material densities in examinationobjects.3.2.2 effective energythe equivalent monoenergetic energyfor a polyenergetic
10、CT system. Thus, the actual, polyenergeticCT system yields the same measured attenuation coefficient foran examination object as a theoretical, monoenergetic CTsystem at the effective energy.3.2.3 phantoma part or item being used to calibrate CTdensity.3.2.4 examination objecta part or specimen bein
11、g sub-jected to CT examination.4. Basis of Application4.1 The procedure is generic and requires mutual agreementbetween purchaser and supplier on many points.5. Significance and Use5.1 This test method allows specification of the densitycalibration procedures to be used to calibrate and performmater
12、ial density measurements using CT image data. Suchmeasurements can be used to evaluate parts, characterize aparticular system, or compare different systems, provided thatobserved variations are dominated by true changes in objectdensity rather than by image artifacts. The specified proceduremay also
13、 be used to determine the effective X-ray energy of aCT system.5.2 The recommended test method is more accurate and lesssusceptible to errors than alternative CT-based approaches,because it takes into account the effective energy of the CTsystem and the energy-dependent effects of the X-ray attenu-a
14、tion process.1This test method is under the jurisdiction of ASTM Committee E07 onNondestructive Testing and is the direct responsibility of Subcommittee E07.01 onRadiology (X and Gamma) Method.Current edition approved June 1, 2013. Published June 2013. Originallyapproved in 1997. Last previous editi
15、on approved in 2008 as E1935 - 97 (2008).DOI: 10.1520/E1935-97R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM web
16、site.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.3 This (or any) test method for measuring density is validonly to the extent that observed CT-number variations arereflective of true changes in object density rather than imagear
17、tifacts. Artifacts are always present at some level and canmasquerade as density variations. Beam hardening artifacts areparticularly detrimental. It is the responsibility of the user todetermine or establish, or both, the validity of the densitymeasurements; that is, they are performed in regions o
18、f theimage which are not overly influenced by artifacts.5.4 Linear attenuation and mass attenuation may be mea-sured in various ways. For a discussion of attenuation andattenuation measurement, see Guide E1441 and PracticeE1570.6. Apparatus6.1 Unless otherwise agreed upon between the purchaserand su
19、pplier, the density calibration phantom shall be con-structed as follows (see Fig. 1):6.1.1 A selection of density standards bracketing the rangeof densities of interest shall be chosen. For best results, thematerials should have known composition and should bephysically homogeneous on a scale compa
20、rable to the spatialresolution of the CT system. It is a good idea to radiographi-cally verify homogeneity and to independently verify chemicalcomposition. All materials should be manufactured to repro-ducible standards. Solids should be readily machinable and notsusceptible to surface damage.6.1.2
21、One or more cylinders of each density standard shallbe machined or prepared, or both. Selecting cylinders overrectangles reduces the uncertainties and streaks that sharpcorners have on volumetric determination and verificationmethods. The cylinders should be large enough that the meanCT number corre
22、sponding to each standard can be computedover a hundred or more uncorrupted (see 8.1.3) pixels but smallenough relative to the dimensions of the host disk that radialeffects are minimal.6.1.3 The physical density of each density standard shall bedetermined empirically by weighing and measuring the s
23、peci-mens as accurately as possible. It is a good idea to indepen-dently verify the measured densities using volumetric displace-ment methods.6.1.4 The mass attenuation coefficient, /, at the effectiveenergy of the system (see 8.3) shall be determined from areference table. For compounds, / can be o
24、btained by takingthe weighted sum of its constituents, in accordance with thefollowing equation:m5 / 5(iwi/!i(1)where:wi= the weight fraction of the ith elemental component.6.1.5 For each density standard, the measured density, ,shall be multiplied by its corresponding mass attenuationcoefficient, /
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME193519972013STANDARDTESTMETHODFORCALIBRATINGANDMEASURINGCTDENSITY 校准 测量 CI 密度 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-529858.html