1、Designation: E 1935 97 (Reapproved 2008)Standard Test Method forCalibrating and Measuring CT Density1This standard is issued under the fixed designation E 1935; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revis
2、ion. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers instruction for determining thedensity calibration of X- and g-ray computed tomography (CT)systems and f
3、or using this information to measure materialdensities from CT images. The calibration is based on anexamination of the CT image of a disk of material withembedded specimens of known composition and density. Themeasured mean CT values of the known standards are deter-mined from an analysis of the im
4、age, and their linear attenu-ation coefficients are determined by multiplying their measuredphysical density by their published mass attenuation coeffi-cient. The density calibration is performed by applying a linearregression to the data. Once calibrated, the linear attenuationcoefficient of an unk
5、nown feature in an image can be measuredfrom a determination of its mean CT value. Its density can thenbe extracted from a knowledge of its mass attenuation coeffi-cient, or one representative of the feature.1.2 CT provides an excellent method of nondestructivelymeasuring density variations, which w
6、ould be very difficult toquantify otherwise. Density is inherently a volumetric propertyof matter. As the measurement volume shrinks, local materialinhomogeneities become more important; and measured valueswill begin to vary about the bulk density value of the material.1.3 All values are stated in S
7、I units.1.4 This standard does not purport to address the safetyconcerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety andhealth practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Docu
8、ments2.1 ASTM Standards:2E 1316 Terminology for Nondestructive ExaminationsE 1441 Guide for Computed Tomography (CT) ImagingE 1570 Practice for Computed Tomographic (CT) Exami-nation3. Terminology3.1 Definitions:3.1.1 The definitions of terms relating to CT, that appear inTerminology E 1316 and Guid
9、e E 1441, shall apply to theterms used in this test method.3.2 Definitions of Terms Specific to This Standard:3.2.1 density calibrationcalibration of a CT system foraccurate representation of material densities in examinationobjects.3.2.2 effective energythe equivalent monoenergetic en-ergy for a po
10、lyenergetic CT system. Thus, the actual, polyen-ergetic CT system yields the same measured attenuationcoefficient for an examination object as a theoretical, monoen-ergetic CT system at the effective energy.3.2.3 phantoma part or item being used to calibrate CTdensity.3.2.4 examination objecta part
11、or specimen being sub-jected to CT examination.4. Basis of Application4.1 The procedure is generic and requires mutual agreementbetween purchaser and supplier on many points.5. Significance and Use5.1 This test method allows specification of the densitycalibration procedures to be used to calibrate
12、and performmaterial density measurements using CT image data. Suchmeasurements can be used to evaluate parts, characterize aparticular system, or compare different systems, provided thatobserved variations are dominated by true changes in objectdensity rather than by image artifacts. The specified p
13、roceduremay also be used to determine the effective X-ray energy of aCT system.5.2 The recommended test method is more accurate and lesssusceptible to errors than alternative CT-based approaches,because it takes into account the effective energy of the CTsystem and the energy-dependent effects of th
14、e X-ray attenu-ation process.5.3 This (or any) test method for measuring density is validonly to the extent that observed CT-number variations arereflective of true changes in object density rather than image1This test method is under the jurisdiction of ASTM Committee E07 onNondestructive Testing a
15、nd is the direct responsibility of Subcommittee E07.01 onRadiology (X and Gamma) Method.Current edition approved Dec. 15, 2008. Published December 2008. Originallyapproved in 1997. Last previous edition approved in 2003 as E 1935 - 97(2003).2For referenced ASTM standards, visit the ASTM website, www
16、.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.artif
17、acts. Artifacts are always present at some level and canmasquerade as density variations. Beam hardening artifacts areparticularly detrimental. It is the responsibility of the user todetermine or establish, or both, the validity of the densitymeasurements; that is, they are performed in regions of t
18、heimage which are not overly influenced by artifacts.5.4 Linear attenuation and mass attenuation may be mea-sured in various ways. For a discussion of attenuation andattenuation measurement, see Guide E 1441 and PracticeE 1570.6. Apparatus6.1 Unless otherwise agreed upon between the purchaserand sup
19、plier, the density calibration phantom shall be con-structed as follows (see Fig. 1):6.1.1 A selection of density standards bracketing the rangeof densities of interest shall be chosen. For best results, thematerials should have known composition and should bephysically homogeneous on a scale compar
20、able to the spatialresolution of the CT system. It is a good idea to radiographi-cally verify homogeneity and to independently verify chemicalcomposition. All materials should be manufactured to repro-ducible standards. Solids should be readily machinable and notsusceptible to surface damage.6.1.2 O
21、ne or more cylinders of each density standard shallbe machined or prepared, or both. Selecting cylinders overrectangles reduces the uncertainties and streaks that sharpcorners have on volumetric determination and verificationmethods. The cylinders should be large enough that the meanCT number corres
22、ponding to each standard can be computedover a hundred or more uncorrupted (see 8.1.3) pixels but smallenough relative to the dimensions of the host disk that radialeffects are minimal.6.1.3 The physical density of each density standard shall bedetermined empirically by weighing and measuring the sp
23、eci-mens as accurately as possible. It is a good idea to indepen-dently verify the measured densities using volumetric displace-ment methods.6.1.4 The mass attenuation coefficient, /r, at the effectiveenergy of the system (see 8.3) shall be determined from areference table. For compounds, /r can be
24、obtained by takingthe weighted sum of its constituents, in accordance with thefollowing equation:m5 /r5(iwi/r!i(1)where:wi= the weight fraction of the ith elemental component.6.1.5 For each density standard, the measured density, r,shall be multiplied by its corresponding mass attenuationcoefficient
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME193519972008STANDARDTESTMETHODFORCALIBRATINGANDMEAUSRINGCTDENSITY 校准 测量 计算机 断层 摄影 密度 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-529857.html