ASTM E1933-1999a(2010) Standard Test Methods for Measuring and Compensating for Emissivity Using Infrared Imaging Radiometers《用红外线成像辐射计测量和补偿辐射率的标准试验方法》.pdf
《ASTM E1933-1999a(2010) Standard Test Methods for Measuring and Compensating for Emissivity Using Infrared Imaging Radiometers《用红外线成像辐射计测量和补偿辐射率的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1933-1999a(2010) Standard Test Methods for Measuring and Compensating for Emissivity Using Infrared Imaging Radiometers《用红外线成像辐射计测量和补偿辐射率的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1933 99a (Reapproved 2010)Standard Test Methods forMeasuring and Compensating for Emissivity Using InfraredImaging Radiometers1This standard is issued under the fixed designation E1933; the number immediately following the designation indicates the year oforiginal adoption or, in the c
2、ase of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods cover procedures for measuring andcompensating for emissivity when mea
3、suring the surfacetemperature of a specimen with an infrared imaging radiom-eter.21.2 The values stated in SI units are to be regarded as thestandard.1.3 These test methods may involve use of equipment andmaterials in the presence of heated or electrically-energizedequipment, or both.1.4 This standa
4、rd does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 AST
5、M Standards:3E1316 Terminology for Nondestructive Examinations3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 reflected temperaturethe temperature of the energyincident upon and reflected from the measurement surface ofthe specimen.3.1.2 surface-modifying materialany tape, spr
6、ay, paint orthe like that is used to change the emissivity of the specimensurface.3.2 See also Terminology E1316.4. Summary of Test Method4.1 Two test methods are given for measuring the emissivityof a specimen surface, the contact thermometer method and thenon-contact thermometer method.4.2 A test
7、method is also given for compensating for theerror produced by emissivity using the computer built into aninfrared imaging radiometer.5. Significance and Use5.1 The emissivity of a specimen can cause surface tem-perature measurement errors. Two test methods are providedfor measuring and compensating
8、 for this error source.5.2 These test methods can be used in the field or laboratory,using commonly available materials.5.3 These test methods can be used with any infraredradiometers that have the required computer capabilities.6. Interferences6.1 Contact Thermometer MethodContact thermometerscan a
9、ct as heat sinks and change the temperature of thespecimen.6.2 Noncontact Thermometer Method:6.2.1 The use of surface-modifying materials can changethe heat transfer properties and temperature of the specimen.Any such errors can be minimized by applying surface-modifying materials to the smallest ar
10、ea that satisfies themeasurement accuracy requirements of the radiometer andinfrared thermographer.6.2.2 Before the surface-modifying material is applied to anarea of the specimen adjacent to the area where the emissivityis to be measured (as directed in 8.2.4), errors can beminimized by viewing the
11、 imager display to ensure that bothareas have the same temperature.6.2.3 When removing a surface-modifying material, as di-rected in 8.2.7, errors can be minimized by ensuring that thesurface is returned to its original condition.6.3 Both test methods require the specimen to be at atemperature that
12、is at least 10C warmer or cooler than theambient temperature. Potential errors can be minimized byensuring the stability of the temperature difference between thespecimen and the ambient temperature during the test.Also, the1These test methods are under the jurisdiction of ASTM Committee E07 onNonde
13、structive Testing and is the direct responsibility of Subcommittee E07.10 onSpecialized NDT Methods.Current edition approved June 1, 2010. Published November 2010. Originallyapproved in 1997. Last previous edition approved in 2005 as E1933 - 99a (2005)1.DOI: 10.1520/E1933-99AR10.2These test methods
14、are adapted from the Guideline for Measuring andCompensating for Reflected Temperature, Emittance and Transmittance developedby the Infraspection Institute, 425 Ellis Street, Burlington, NJ 08016.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at
15、 serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.emissivity measurement accuracy can be increased b
16、y increas-ing this temperature difference.6.4 The emissivity of a specimen may be specific to thetemperature of the specimen and the spectral waveband of theinfrared imaging radiometer used to make the measurement.Therefore, the temperature of the specimen and the spectralwaveband of the radiometer
17、should be noted along with themeasured emissivity value.6.5 These test methods are valid only for specimens that areopaque in the waveband of the infrared imaging radiometer.6.6 As the emissivity of a specimen decreases, its reflectiv-ity increases. Careful consideration and avoidance of potentialer
18、ror sources, including the precise determination of reflectedtemperature in 8.1.3 and 8.2.3, is required to accuratelymeasure the emissivity values of specimens having loweremissivities. For materials with emissivities less than 0.5,radiometric temperature measurements and emissivity mea-surements m
19、ay have a high likelihood of error.7. Apparatus7.1 Calibrated Infrared Imaging Radiometer, with a built-incomputer that allows the infrared thermographer to inputreflected temperatures and emissivity values.7.2 Tripod, or device to support the infrared imaging radi-ometer.7.3 A natural or induced me
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME19331999A2010STANDARDTESTMETHODSFORMEASURINGANDCOMPENSATINGFOREMISSIVITYUSINGINFRAREDIMAGINGRADIOMETERS

链接地址:http://www.mydoc123.com/p-529851.html