1、Designation: E1840 96 (Reapproved 2014)Standard Guide forRaman Shift Standards for Spectrometer Calibration1This standard is issued under the fixed designation E1840; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers Raman shift values for commonliquid and solid chemicals that can be used for wavenumbercalibration of Ra
3、man spectrometers. The guide does notinclude procedures for calibrating Raman instruments. Instead,this guide provides reliable Raman shift values that can be usedas a complement to low-pressure arc lamp emission lineswhich have been established with a high degree of accuracyand precision.1.2 The va
4、lues stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 Some of the chemicals specified in this guide may behazardous. It is the responsibility of the user of this guide toconsult material safety data sheets and other pertinent infor-mati
5、on to establish appropriate safety and health practices anddetermine the applicability of regulatory limitations prior totheir use.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establi
6、sh appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E131 Terminology Relating to Molecular SpectroscopyE1683 Practice for Testing the Performance of ScanningRaman Spectrometers3. Terminology3.1
7、 DefinitionsTerminology used in this guide conformsto the definitions set forth in Terminology E131.4. Significance and Use4.1 Wavenumber calibration is an important part of Ramananalysis.The calibration of a Raman spectrometer is performedor checked frequently in the course of normal operation ande
8、ven more often when working at high resolution. To date, themost common source of wavenumber values is either emissionlines from low-pressure discharge lamps (for example,mercury, argon, or neon) or from the non-lasing plasma lines ofthe laser. There are several good compilations of these well-estab
9、lished values (1-8).3The disadvantages of using emissionlines are that it can be difficult to align lamps properly in thesample position and the laser wavelength must be knownaccurately.With argon, krypton, and other ion lasers commonlyused for Raman the latter is not a problem because lasingwavelen
10、gths are well known. With the advent of diode lasersand other wavelength-tunable lasers, it is now often the casethat the exact laser wavelength is not known and may bedifficult or time-consuming to determine. In these situations itis more convenient to use samples of known relative wave-number shif
11、t for calibration. Unfortunately, accurate wavenum-ber shifts have been established for only a few chemicals. Thisguide provides the Raman spectroscopist with average shiftvalues determined in seven laboratories for seven pure com-pounds and one liquid mixture.5. Raman Shift Standards5.1 Reagents an
12、d MethodologyRaman shifts were mea-sured in seven laboratories for the following eight materials:1This guide is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.08 on Raman Spectroscopy.Current edition
13、approved May 1, 2014. Published June 2014. Originaly approvedin 1996. Last previous edition approved in 2007 as E1840 96(2007). DOI:10.1520/E1840-96R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMSta
14、ndards volume information, refer to the standards Document Summary page onthe ASTM website.3The boldface numbers in parentheses refer to a list of references at the end ofthe text.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1Compou
15、nd SourceNaphthalene Mallinckrodt41,4-Bis(2-methylstyryl)benzene (a laser dye) Aldrich5Sulfur Aldrich550/50 (v/v) toluene/acetonitrile Mallinckrodt44-Acetamidophenol6Aldrich5Benzonitrile Aldrich5Cyclohexane Mallinckrodt4Polystyrene Aldrich55.1.1 The eight materials were selected to cover a widewaven
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME184019962014STANDARDGUIDEFORRAMANSHIFTSTANDARDSFORSPECTROMETERCALIBRATION 分光计 校准 用拉曼 位移 标准 指南 PDF

链接地址:http://www.mydoc123.com/p-529617.html