ASTM E1840-1996(2007) Standard Guide for Raman Shift Standards for Spectrometer Calibration《分光剂校准RAMAN转换标准的标准指南》.pdf
《ASTM E1840-1996(2007) Standard Guide for Raman Shift Standards for Spectrometer Calibration《分光剂校准RAMAN转换标准的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1840-1996(2007) Standard Guide for Raman Shift Standards for Spectrometer Calibration《分光剂校准RAMAN转换标准的标准指南》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1840 96 (Reapproved 2007)Standard Guide forRaman Shift Standards for Spectrometer Calibration1This standard is issued under the fixed designation E 1840; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of la
2、st revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers Raman shift values for commonliquid and solid chemicals that can be used for wavenumbercalibration of
3、 Raman spectrometers. The guide does notinclude procedures for calibrating Raman instruments. Instead,this guide provides reliable Raman shift values that can be usedas a complement to low-pressure arc lamp emission lineswhich have been established with a high degree of accuracyand precision.1.2 The
4、 values stated in SI units are to be regarded as thestandard.1.3 Some of the chemicals specified in this guide may behazardous. It is the responsibility of the user of this guide toconsult material safety data sheets and other pertinent infor-mation to establish appropriate safety and health practic
5、es anddetermine the applicability of regulatory limitations prior totheir use.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and deter
6、mine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 131 Terminology Relating to Molecular SpectroscopyE 1683 Practice for Testing the Performance of ScanningRaman Spectrometers3. Terminology3.1 DefinitionsTerminology used in this guide conforms
7、to the definitions set forth in Terminology E 131.4. Significance and Use4.1 Wavenumber calibration is an important part of Ramananalysis.The calibration of a Raman spectrometer is performedor checked frequently in the course of normal operation andeven more often when working at high resolution. To
8、 date, themost common source of wavenumber values is either emissionlines from low-pressure discharge lamps (for example, mer-cury, argon, or neon) or from the non-lasing plasma lines of thelaser. There are several good compilations of these well-established values (1-8).3The disadvantages of using
9、emissionlines are that it can be difficult to align lamps properly in thesample position and the laser wavelength must be knownaccurately.With argon, krypton, and other ion lasers commonlyused for Raman the latter is not a problem because lasingwavelengths are well known. With the advent of diode la
10、sersand other wavelength-tunable lasers, it is now often the casethat the exact laser wavelength is not known and may bedifficult or time-consuming to determine. In these situations itis more convenient to use samples of known relative wave-number shift for calibration. Unfortunately, accurate waven
11、um-ber shifts have been established for only a few chemicals. Thisguide provides the Raman spectroscopist with average shiftvalues determined in seven laboratories for seven pure com-pounds and one liquid mixture.5. Raman Shift Standards5.1 Reagents and MethodologyRaman shifts were mea-sured in seve
12、n laboratories for the following eight materials:Compound SourceNaphthalene Mallinckrodtfvariant41,4-Bis(2-methylstyryl)benzene (a laser dye) Aldrichfvariant5Sulfur Aldrichfvariant550/50 (v/v) toluene/acetonitrile Mallinckrodtfvariant44-Acetamidophenolfvariant6Aldrichfvariant5Benzonitrile Aldrichfva
13、riant5Cyclohexane Mallinckrodtfvariant4Polystyrene Aldrichfvariant51This guide is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.08 on Raman Spectroscopy.Current edition approved March 1, 2007. Publis
14、hed March 2007. Originalyapproved in 1996. Last previous edition approved in 2002 as E 1840 96 (2002).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards
15、Document Summary page onthe ASTM website.3The boldface numbers in parentheses refer to a list of references at the end ofthe text.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.1.1 The eight materials were selected to cover a wide
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME184019962007STANDARDGUIDEFORRAMANSHIFTSTANDARDSFORSPECTROMETERCALIBRATION 分光 校准 RAMAN 转换 标准 指南 PDF

链接地址:http://www.mydoc123.com/p-529616.html