ASTM E1834-1996(2002) Standard Test Method for Determination of Lead in Nickel Alloys by Electrothermal Atomic Absorption Spectrometric Method《电热原子吸收光谱法测定镍合金中铅的标准试验方法》.pdf
《ASTM E1834-1996(2002) Standard Test Method for Determination of Lead in Nickel Alloys by Electrothermal Atomic Absorption Spectrometric Method《电热原子吸收光谱法测定镍合金中铅的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1834-1996(2002) Standard Test Method for Determination of Lead in Nickel Alloys by Electrothermal Atomic Absorption Spectrometric Method《电热原子吸收光谱法测定镍合金中铅的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1834 96 (Reapproved 2002)Standard Test Method forDetermination of Lead in Nickel Alloys by ElectrothermalAtomic Absorption Spectrometric Method1This standard is issued under the fixed designation E 1834; the number immediately following the designation indicates the year oforiginal ad
2、option or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of lead innickel and nickel
3、 alloys in the concentration range of0.00005 % to 0.001 % by electrothermal atomic absorptionspectrometry.NOTE 1If this test method is used to test materials having contentsless than 0.0001 % lead, users in different laboratories will experiencemore than the usual 5 % risk that their results will di
4、ffer by more than50 % relative error.1.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory lim
5、itations prior to use. For specific hazardsassociated with the use of this practice see Practices E 50.2. Referenced Documents2.1 ASTM Standards:2E 50 Practices for Apparatus, Reagents, and Safety Precau-tions for Chemical Analysis of Metals, Ores, and RelatedMaterialsE 1184 Practice for Electrother
6、mal (Graphite Furnace)Atomic Absorption AnalysisE 1452 Practice for Preparation of Calibration Solutions forSpectrophotometric and Spectroscopic Atomic AnalysisE 1601 Practice for Conducting an Interlaboratory Study toEvaluate the Performance of an Analytical MethodE 1770 Practice for Optimization o
7、f Electrothermal AtomicAbsorption Spectrometric Equipment2.2 ISO Standards:ISO Standard 11437Nickel AlloysDetermination ofTrace-Element ContentElectrothermal Atomic Absorp-tion Spectrometric MethodPart 2: Determination of LeadContent3ISO Guide 5725Accuracy, Trueness, and Precision ofMeasurements, Me
8、thods and Results33. Summary of Test Method3.1 The sample is dissolved in a mixture of nitric acid,hydrofluoric acid, and water. The sample and a nickel-ammonium phosphate matrix modifier are injected into theelectrothermal atomizer of an atomic absorption spectrometer,which is provided with a backg
9、round corrector. Measurementof the absorbance is made at a wavelength of 283.3 nm. Thelead concentration is determined from a calibration curveestablished with nickel solutions containing a known amountof lead.4. Significance and Use4.1 This test method is intended to be used for the determi-nation
10、of trace levels of lead in nickel and nickel alloys. It isassumed that the procedure will be performed by trainedanalysts capable of performing common laboratory practicesskillfully and safely. It is expected that the work will beperformed in a properly equipped laboratory and proper wastedisposal p
11、rocedures will be followed.5. Apparatus5.1 Atomic Absorption Spectrometer, with electrothermalatomizer, equipped with an appropriate background corrector, asignal output device, such as a video display screen, a digitalcomputer, a printer or strip chart recorder, and an autosampler.5.2 Pyrolytically
12、 Coated Graphite Tubes, grooved, con-forming to the instrument manufacturers specifications.5.3 Pyrolytic Graphite Platforms, Lvov design, fitted to thetubes specified in 5.2.5.4 Radiation Source, lead hollow cathode lamp, or elec-trodeless discharge lamp with an appropriate power supply.1This test
13、method is under the jurisdiction of ASTM Committee E01 onAnalytical Chemistry for Metals, Ores, and Related Materials and is the directresponsibility of Subcommittee E01.08 on Ni and Co and High Temperature Alloys.Current edition approved Oct. 10, 1996. Published December 1996.2For referenced ASTM s
14、tandards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute, 25 W. 43rd St., 4thFloor, Ne
15、w York, NY 10036.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. Instrument Parameters and Criteria6.1 The atomic absorption spectrometer and electrothermalatomizer are satisfactory if after adjustment of the instrumentsand optimi
16、zation of the furnace heating program as describedin Practice E 1770, they meet the criteria given in 6.4-6.7.6.2 Calibration solutions S0through S3prepared in accor-dance with 7.9 and summarized in Table 1 shall be used fortesting criteria.6.3 The parameters for the determination of lead and theest
17、ablishing of the instrument criteria are as follows:6.3.1 The injection volume is 20 L.6.3.2 The peak area absorbance integration measurementsshall be at a wavelength of 283.3 nm.6.4 Characteristic MassThe characteristic mass deter-mined in accordance with Practice E 1770 shall be within 20 %of that
18、 given in the manufacturers literature.6.5 PrecisionThe variability obtained from calibrationsolution S3shall not exceed 10 % of the mean absorbance ofthe same solution, and the variability of calibration solution S1shall not exceed 4 % of the mean absorbance of solution S3when determined in accorda
19、nce with Practice E 1770.6.6 Limit of DetectionThe limit of detection of lead asdescribed in Practice E 1770 shall be less than 20 pg (equiva-lent to 1.0 g/L in the test solution or 0.2 ppm in the originalsample).6.7 LinearityThe linearity of the calibration as deter-mined in Practice E 1770 shall n
20、ot be less than 0.7.7. Reagents7.1 Purity and Concentration of ReagentsThe purity andconcentration of common chemical reagents shall conform toPractices E 50. The reagents should be free of or containminimal amounts (0.01 g/g) of lead.7.2 Dissolution Acid MixtureTo 150 mL water carefullyadd 150 mL n
21、itric acid and 150 mL hydrofluoric acid. Mix andstore in plastic bottle.7.3 Pure Nickel Metal99.9 % minimum purity nickel, andshall contain less than 0.1 ppm lead.7.4 Nickel, Stock Solution (50 g/L)Dissolve 25 6 0.1gofthe pure nickel in 200 mL of nitric acid (1:1) in a 600 mLbeaker. Cool the solutio
22、n and transfer to a 500 mL volumetricflask. Dilute to the mark and mix.NOTE 2If large nickel turnings or chunks are used, gentle heating maybe required to complete the dissolution. When using carbonyl nickelpowder, the solution must be filtered to remove undissolved carbon.7.5 Nickel, Solution (5.0
23、g/L)Transfer 10.0 mL of theNickel Stock Solution to a 100 mL volumetric flask. Dilute tothe mark with water and mix.7.6 Nickel Nitrate-Ammonium Phosphate Matrix ModifierSolutionDissolve 6.0 6 0.1 g high purity ammonium dihy-drogen phosphate (NH4H2PO4) in 50 mLwater. Transfer to 100mL volumetric flas
24、k and add 20 mL of nickel solution (7.5).Dilute with water and mix. This solution shall be freshlyprepared.7.7 Lead Reference Solution (100 mg/L)Dissolve 0.100 60.001 g of lead metal (99 % minimum purity) in 20 mL nitricacid (1:1) in a 250-mL beaker. Heat to assist dissolution. Cooland transfer to 1
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME183419962002STANDARDTESTMETHODFORDETERMINATIONOFLEADINNICKELALLOYSBYELECTROTHERMALATOMICABSORPTIONSPECTROMETRICMETHOD

链接地址:http://www.mydoc123.com/p-529597.html