ASTM E1781 E1781M-2013 Standard Practice for Secondary Calibration of Acoustic Emission Sensors《声发射传感器二级校准的标准实施规程》.pdf
《ASTM E1781 E1781M-2013 Standard Practice for Secondary Calibration of Acoustic Emission Sensors《声发射传感器二级校准的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1781 E1781M-2013 Standard Practice for Secondary Calibration of Acoustic Emission Sensors《声发射传感器二级校准的标准实施规程》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1781 08E1781/E1781M 13Standard Practice forSecondary Calibration of Acoustic Emission Sensors1This standard is issued under the fixed designation E1781;E1781/E1781M; the number immediately following the designation indicatesthe year of original adoption or, in the case of revision, the
2、 year of last revision. A number in parentheses indicates the year of lastreapproval. A superscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope Scope*1.1 This practice covers requirements for the secondary calibration of acoustic emission (AE) sensors. The
3、 secondarycalibration yields the frequency response of a sensor to waves of the type normally encountered in acoustic emission work. Thesource producing the signal used for the calibration is mounted on the same surface of the test block as the sensor under testing(SUT). Rayleigh waves are dominant
4、under these conditions; the calibration results represent primarily the sensors sensitivity toRayleigh waves. The sensitivity of the sensor is determined for excitation within the range of 100 kHz to 1 MHz. Sensitivity valuesare usually determined at frequencies approximately 10 kHz apart. The units
5、 of the calibration are volts per unit of mechanicalinput (displacement, velocity, or acceleration).1.2 UnitsThe values stated in either SI units or inch-pound units are to be regarded as the standard. The values given inparentheses are for information only. stated in each system may not be exact eq
6、uivalents; therefore, each system shall be usedindependently of the other. Combining values from the two systems may result in non-conformance with the standards.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the
7、user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E114 Practice for Ultrasonic Pulse-Echo Straight-Beam Contact TestingE494 Practice for Measuring Ultrasonic Velo
8、city in MaterialsE1106 Test Method for Primary Calibration of Acoustic Emission SensorsE1316 Terminology for Nondestructive Examinations3. Terminology3.1 DefinitionsRefer to Terminology E1316, Section B, for terms used in this practice.3.2 Definitions of Terms Specific to This Standard:3.2.1 referen
9、ce sensor (RS)a sensor that has had its response established by primary calibration (also called secondarystandard transducer) (see Method E1106).3.2.2 secondary calibrationa procedure for measuring the frequency or transient response of an AE sensor by comparisonwith an RS.3.2.3 test blocka block o
10、f homogeneous, isotropic, elastic material on which a source, an RS, and a SUT are placed forconducting secondary calibration.4. Significance and Use4.1 The purpose of this practice is to enable the transfer of calibration from sensors that have been calibrated by primarycalibration to other sensors
11、.1 This practice is under the jurisdiction ofASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.04 on Acoustic EmissionMethod.Current edition approved Dec. 15, 2008June 1, 2013. Published January 2009June 2013. Originally approved in 1996. Last previous
12、 edition approved in 19982008 asE1781 - 98E1781 - 08.(2003)1. DOI: 10.1520/E1781-08.10.1520/E1781_E1781M-13.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the stand
13、ards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accuratel
14、y, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Driv
15、e, PO Box C700, West Conshohocken, PA 19428-2959. United States15. General Requirements5.1 Units for CalibrationSecondary calibration produces the same type of information regarding a sensor as does primarycalibration (Method E1106). An AE sensor responds to motion at its front face. The actual stre
16、ss and strain at the front face of amounted sensor depends on the interaction between the mechanical impedance of the sensor (load) and that of the mounting block(driver); neither the stress nor the strain is amenable to direct measurement at this location. However, the free displacement thatwould o
17、ccur at the surface of the block in the absence of the sensor can be inferred from measurements made elsewhere on thesurface. SinceAE sensors are used to monitor motion at a free surface of a structure and interactive effects between the sensor andthe structure are generally of no interest, the free
18、 motion is the appropriate input variable. It is therefore required that the units ofcalibration shall be volts per unit of free displacement or free velocity, that is, volts per metreunit or volt seconds per metre.unit.5.2 The calibration results may be expressed, in the frequency domain, as the st
19、eady-state magnitude and phase response of thesensor to steady-state sinusoidal excitation or, in the time domain, as the transient response of the sensor to a delta function ofdisplacement.5.3 Importance of the Test Block MaterialThe specific acoustical impedance (c) of the test block is an importa
20、nt parameterthat affects calibration results. Calibrations performed on blocks of different materials yield sensor sensitivities that are verydifferent. For example, a sensor that has been calibrated on a steel block, if calibrated on a glass or aluminum block, may havean average sensitivity that is
21、 50 % of the value obtained on steel and, if calibrated on a polymethyl methacrylate block, may havean average sensitivity that is 3 % of the value obtained on steel.35.3.1 For a sensor having a circular aperture (mounting face) with uniform sensitivity over the face, there are frequencies atwhich n
22、ulls in the frequency response occur. These nulls occur at the zeroes of the first order Bessel function, J1 (ka), where k= 2pif/c,f = frequency, c = the Rayleigh speed in the test block, and a = the radius of the sensor face.3 Therefore, calibrationresults depend on the Rayleigh wave speed in the m
23、aterial of the test block.5.3.2 For the reasons outlined in 5.3 and 5.3.1, all secondary calibration results are specific to a particular material; a secondarycalibration procedure must specify the material of the block.46. Requirements of the Secondary Calibration Apparatus6.1 Basic SchemeA prototy
24、pe apparatus for secondary calibration is shown in Fig. 1. A glass-capillary-break device or othersuitable source device (A) is deployed on the upper face of the steel test block (B). The RS (C) and the SUT (D) are placed at equaldistances from the source and in opposite directions from it. Because
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME1781E1781M2013STANDARDPRACTICEFORSECONDARYCALIBRATIONOFACOUSTICEMISSIONSENSORS 声发 传感器 二级 校准 标准 实施

链接地址:http://www.mydoc123.com/p-529445.html