ASTM E1770-1995(2006) Standard Practice for Optimization of Electrothermal Atomic Absorption Spectrometric Equipment《电热原子吸收光谱仪器优化的标准实用规程》.pdf
《ASTM E1770-1995(2006) Standard Practice for Optimization of Electrothermal Atomic Absorption Spectrometric Equipment《电热原子吸收光谱仪器优化的标准实用规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1770-1995(2006) Standard Practice for Optimization of Electrothermal Atomic Absorption Spectrometric Equipment《电热原子吸收光谱仪器优化的标准实用规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1770 95 (Reapproved 2006)Standard Practice forOptimization of Electrothermal Atomic AbsorptionSpectrometric Equipment1This standard is issued under the fixed designation E 1770; the number immediately following the designation indicates the year oforiginal adoption or, in the case of
2、revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the optimization of electrothermalatomic absorption spectrometers and the
3、checking of spectrom-eter performance criteria.1.2 The values stated in SI units are to be regarded as thestandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate
4、safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E50 Practices for Apparatus, Reagents, and Safety Consid-erations for Chemical Analysis of Metals, Ores, andRelated MaterialsE 876 Practice for Use of Statis
5、tics in the Evaluation ofSpectrometric Data3E 1184 Practice for Electrothermal (Graphite Furnace)Atomic Absorption AnalysisE 1452 Practice for Preparation of Calibration Solutions forSpectrophotometric and for Spectroscopic Atomic Analy-sis33. Significance and Use3.1 This practice is for optimizing
6、the parameters used inthe determination of trace elements in metals and alloys by theelectrothermal atomic absorption spectrometric method. It alsodescribes the practice for checking the spectrometer perfor-mance. The work is expected to be performed in a properlyequipped laboratory by trained opera
7、tors and appropriatedisposal procedures are to be followed.4. Apparatus4.1 Atomic Absorption Spectrometer with ElectrothermalAtomizer, equipped with an appropriate background corrector,a signal output device such as a video display screen (VDS), adigital computer, a printer or strip chart recorder,
8、and anautosampler.4.2 Grooved Pyrolytic Graphite-Coated Graphite Tubes,conforming to the instrument manufacturers specifications.4.3 Pyrolytic Graphite Platforms, Lvov design, fitted to thetubes specified in 4.2.4.4 Pyrolytic Graphite-Coated Graphite Tubes, platform-less, conforming to the instrumen
9、t manufacturers specifica-tions.4.5 Radiation Source for the AnalyteA hollow cathodelamp or electrodeless discharge lamp is suitable.NOTE 1The use of multi-element lamps is not generally recom-mended, since they may be subject to spectral line overlaps.4.6 For general discussion of the theory and in
10、strumentalrequirements of electrothermal atomic absorption spectromet-ric analysis, see Practice E 1184.5. Reagents5.1 Purity and Concentration of ReagentsThe purity andconcentration of common chemical reagents shall conform toPractices E50. The reagents should be free of or containminimal amounts (
11、0.01 g/g) of the analyte of interest.5.2 Magnesium Nitrate Solution 2 g/L Mg(NO3)2Dissolve 0.36 6 0.01 g high-purity Mg(NO3)26H2O in about50 mL of water, in a 100-mL beaker, and transfer the solutioninto a 100-mL volumetric flask. Dilute to mark with water andmix. Store in polypropylene or high-dens
12、ity polyethylenebottle.5.3 Calibration SolutionsRefer to the preparation of cali-bration solutions in the relevant analytical method for thedetermination of trace elements in the specific matrix. Calibra-tion solution S0represents the calibration solution containingno analyte; S1the least concentrat
13、ed calibration solution; S2thecalibration solution with the next highest concentration;1This practice is under the jurisdiction of ASTM Committee E01 on AnalyticalChemistry for Metals, Ores, and Related Materials and is the direct responsibility ofSubcommittee E01.20 Fundamental Practices.Current ed
14、ition approved June 1, 2006. Published June 2006. Originallyapproved in 1995. Last previous edition approved in 2001 as E 1770 95 (2001).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume i
15、nformation, refer to the standards Document Summary page onthe ASTM website.3Withdrawn.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.through Sk, the most concentrated calibration solution. Alsorefer to Practice E 1452.5.4 Matrix Mo
16、difiersRefer to the relevant analyticalmethod for the determination of trace elements in the specificmatrix.6. Initial Checks and Adjustments6.1 Turn on power, cooling water, gas supplies, and fumeexhaust system.6.2 Open the furnace to inspect the tube and contacts.Replace graphite components, if we
17、ar or contamination isevident. Inspect windows and clean or replace as required.6.2.1 New graphite contacts or new tubes should be condi-tioned prior to use, in accordance with the heating programrecommended by the manufacturer.6.2.1.1 In the absence of manufacturers recommendations,a conditioning p
18、rogram for a graphite furnace is shown inTable 1.7. Radiation Source7.1 Install and operate hollow cathode lamps or electrode-less discharge lamps in accordance with the manufacturersinstructions.7.2 After the manufacturers prescribed warm-up time, thesignal from the radiation source should not devi
19、ate by morethan 0.5 % from the maximum value (that is, by not more than0.002 absorbance units) over a period of 15 min. Significantlygreater fluctuations are usually indicative of a faulty lamp orpower supply.8. Spectrometer Parameters8.1 Wavelength, as specified by the appropriate procedure.8.2 Sli
20、t Width, as recommended by the manufacturer. Wheretwo slit height settings are available, select the shorter height.8.3 Background Correction:8.3.1 Zeeman Background Correction System:8.3.1.1 Ensure that the poles of the magnet are clean andsecurely tightened.8.3.1.2 If necessary, set the optical te
21、mperature sensor inaccordance with the instrument manufacturers instructions.8.3.2 Continuum Background System:8.3.2.1 Select the background correction option and allowlamps to stabilize for 30 min. Verify that the energies of theanalyte lamp and the deuterium lamp are balanced withintolerances reco
22、mmended by the manufacturer.8.3.2.2 If necessary, set the optical temperature sensor inaccordance with the instrument manufacturers recommenda-tion.8.3.3 To check the performance of the background correc-tion system, measure the atomic background absorbance of 20L of 2 g/L magnesium nitrate solution
23、 at a wavelength in the200 to 250 nm region (for example, Bi 223.1 nm) using a drytemperature of 120C, a pyrolysis temperature of 950C, andan atomization temperature of 1800C. A large backgroundsignal should be observed with no over-or under-correction ofthe atomic signal.NOTE 2In general, Zeeman sy
24、stems should compensate for back-ground levels as high as 1.0 to 1.5 absorbance units. A continuumcorrection system should be able to correct for the broad-band backgroundabsorbance up to 0.5 to 0.6 absorbance units.8.4 AutosamplerCheck operation of the autosampler. Payparticular attention to the co
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME177019952006STANDARDPRACTICEFOROPTIMIZATIONOFELECTROTHERMALATOMICABSORPTIONSPECTROMETRICEQUIPMENT

链接地址:http://www.mydoc123.com/p-529423.html