1、Designation: E 1647 09Standard Practice forDetermining Contrast Sensitivity in Radiology1This standard is issued under the fixed designation E 1647; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A numbe
2、r in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the design and material selection ofa contrast sensitivity measuring gauge used to determine theminimum change in mater
3、ial thickness or density that may beimaged without regard to spatial resolution limitations.1.2 This practice is applicable to transmitted-beam radio-graphic and radioscopic imaging systems utilizing X-ray andgamma ray radiation sources.1.3 The values stated in inch-pound units are to be regardedas
4、standard. The SI units given in parentheses are for informa-tion only.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the
5、 applica-bility of regulatory limitations prior to use. For specific safetystatements, see NIST/ANSI Handbook 114 Section 8, Code ofFederal Regulations 21 CFR 1020.40 and 29 CFR 1910.96.2. Referenced Documents2.1 ASTM Standards:2B 139/B 139M Specification for Phosphor Bronze Rod,Bar, and ShapesB 150
6、/B 150M Specification for Aluminum Bronze Rod,Bar, and ShapesB 161 Specification for Nickel Seamless Pipe and TubeB 164 Specification for Nickel-Copper Alloy Rod, Bar, andWireB 166 Specification for Nickel-Chromium-Iron Alloys(UNS N06600, N06601, N06603, N06690, N06693,N06025, N06045, and N06696)* a
7、nd Nickel-Chromium-Cobalt-Molybdenum Alloy (UNS N06617) Rod, Bar, andWireE 747 Practice for Design, Manufacture and MaterialGrouping Classification of Wire Image Quality Indicators(IQI) Used for RadiologyE 1000 Guide for RadioscopyE 1025 Practice for Design, Manufacture, and MaterialGrouping Classif
8、ication of Hole-Type Image Quality Indi-cators (IQI) Used for RadiologyE 1255 Practice for RadioscopyE 1316 Terminology for Nondestructive ExaminationsE 1411 Practice for Qualification of Radioscopic SystemsE 2002 Practice for Determining Total Image Unsharpnessin RadiologyE 2445 Practice for Qualif
9、ication and Long-Term Stabilityof Computed Radiology Systems2.2 Federal Standards:321 CFR 1020.40 Safety Requirements for Cabinet X-raySystems29 CFR 1910.96 Ionizing Radiation2.3 NIST/ANSI Standards:NIST/ANSI Handbook 114 General Safety Standard forInstallations Using Non-Medical X-ray and SealedGam
10、ma Ray Sources, Energies to 10 MeV42.4 Other Standard:5EN 462 5 Duplex Wire Image Quality IndicatorEN 130681 Radioscopic Testing-Part 1: Qualitative Mea-surement of Imaging Properties3. Terminology3.1 DefinitionsDefinitions of terms applicable to this testmethod may be found in Terminology E 1316.4.
11、 Summary of Practice4.1 It is often useful to evaluate the contrast sensitivity of apenetrating radiation imaging system separate and apart fromspatial resolution measurements. Conventional image qualityindicators (IQIs), such asTest Method E 747 wire and PracticeE 1025 plaque IQIs, combine the cont
12、rast sensitivity andresolution measurements into an overall performance figure ofmerit, other methods such as included in Practice E 2002 do1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommittee E07.01 onRadiology (X
13、 and Gamma) Method.Current edition approved July 1, 2009. Published August 2009. Originallyapproved in 1994. Last previous edition approved in 2003 as E 1647 03.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book o
14、f ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from U.S. Government Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http:/www.access.gpo.gov.4Available fromAmerican Society for N
15、ondestructiveTesting (ASNT), P.O. Box28518, 1711 Arlingate Ln., Columbus, OH 43228-0518, http:/www.asnt.org.5Available from British Standards Institute (BSI), 389 Chiswick High Rd.,London W4 4AL, U.K., http:/www.bsi-.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocke
16、n, PA 19428-2959, United States.not address contrast specifically. Such figures of merit are oftennot adequate to detect subtle changes in imaging systemperformance. For example, in a high contrast image, spatialresolution can degrade with almost no noticeable effect uponoverall image quality. Simil
17、arly, in an application in which theimaging system provides a very sharp image, contrast can fadewith little noticeable effect upon the overall image quality.These situations often develop and may go unnoticed until thesystem performance deteriorates below acceptable image qual-ity limits.5. Signifi
18、cance and Use5.1 The contrast sensitivity gauge measures contrast sensi-tivity independent of the imaging system spatial resolutionlimitations. The thickness recess dimensions of the contrastsensitivity gauge are large with respect to the spatial resolutionlimitations of most imaging systems. Four l
19、evels of contrastsensitivity are measured: 4 %, 3 %, 2 %, and 1 %.5.2 The contrast sensitivity gauge is intended for use inconjunction with a high-contrast resolution measuring gauge,such as the EN 462 5 Duplex Wire Image Quality Indicator.Such gauges measure spatial resolution essentially independe
20、ntof the imaging systems contrast sensitivity. Such measure-ments are appropriate for the qualification and performancemonitoring of radiographic and radioscopic imaging systemswith film, realtime devices, Computed Radiography (CR) andDigital Detector Arrays (DDA).5.3 Radioscopic/radiographic system
21、 performance may bespecified by combining the measured contrast sensitivity ex-pressed as a percentage with the spatial resolution expressed inmillimeters of unsharpness. For the EN 462 5 spatial resolu-tion gauge, the unsharpness is equal to twice the wire diameter.For the line pair gauge, the unsh
22、arpness is equal to thereciprocal of the line-pair/mm value. As an example, animaging system that exhibits 2 % contrast sensitivity andimages the 0.1 mm EN 462 5 paired wires (equivalent toimaging 5 line-pairs/millimeter resolution on a line-pair gauge)performs at a 2 %0.2 mm sensitivity level.Astan
23、dard methodof evaluating overall radioscopic system performance is givenin Practice E 1411 and in EN 130681 and for CR it can befound in Practice E 2445.6. Contrast Sensitivity Gauge Construction and MaterialSelection6.1 Contrast sensitivity gauges shall be fabricated in accor-dance with Fig. 1, usi
24、ng the dimensions given in Table 1, Table2, and Table 3.6.2 The gauge shall preferably be fabricated from theexamination object material. Otherwise, the following materialselection guidelines are to be used:6.2.1 Materials are designated in eight groupings, in accor-dance with their penetrating radi
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME16472009STANDARDPRACTICEFORDETERMININGCONTRASTSENSITIVITYINRADIOSCOPY 射线 检验 测定 灵敏度 标准 实施 规程 PDF

链接地址:http://www.mydoc123.com/p-529135.html