ASTM E1582-2017 Standard Test Method for Temperature Calibration of Thermogravimetric Analyzers《热重分析仪温度校准的标准试验方法》.pdf
《ASTM E1582-2017 Standard Test Method for Temperature Calibration of Thermogravimetric Analyzers《热重分析仪温度校准的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1582-2017 Standard Test Method for Temperature Calibration of Thermogravimetric Analyzers《热重分析仪温度校准的标准试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1582 17Standard Test Method forTemperature Calibration of Thermogravimetric Analyzers1This standard is issued under the fixed designation E1582; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisio
2、n. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method describe the temperature calibration ofthermogravimetric analyzers over the temperature range from25 to 1500 C a
3、nd is applicable to commercial and custom-built apparatus. This calibration may be accomplished by theuse of either melting point standards or magnetic transitionstandards.1.2 The weight change curve in thermogravimetry resultsfrom a number of influences, some of which are characteristicof the speci
4、men holder assembly and atmosphere rather thanthe specimen. The variations from instrument to instrumentoccur in the point of measurement of the temperature, thenature of the material, its size and packing, the geometry andcomposition of the specimen container, the geometry anddesign of the furnace,
5、 and the accuracy and sensitivity of thetemperature sensor and displaying scales. These all contributeto differences in measured temperatures, which may exceed 20C. In addition, some sample holder assemblies will showvariations of measured temperature with sample size orheating/cooling rate, or both
6、. Since it is neither practical noradvisable to standardize sample holders or thermobalancegeometries, instruments may be calibrated by measurement ofthe deviation of a melting or magnetic (Curie Point) transitiontemperature from the standard reference temperature. Thisdeviation can be applied as a
7、correction term to subsequentmeasurements.1.3 This test method assumes that the indicated temperatureof the instrument is linear over the range defined by a two-pointcalibration and that this linearity has been verified. These twocalibration temperatures should be as close to the experimentalmeasure
8、ments to be made as possible.1.4 This test method describes two procedures for tempera-ture calibration of thermogravimetric analyzers using any typebalance. Procedure A uses melting point standards for calibra-tion. Procedure B uses magnetic transition standards forcalibration.1.5 The data generate
9、d by these procedures can be used tocorrect the temperature scale of the instrument by either apositive or negative amount using either a one- or two-pointtemperature calibration procedure.1.6 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in
10、thisstandard.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations
11、 prior to use.1.8 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization Techni
12、calBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E473 Terminology Relating to Thermal Analysis and Rhe-ologyE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test MethodE967 Test Method for Temperature Calibration of Differen-tial Sca
13、nning Calorimeters and Differential Thermal Ana-lyzersE1142 Terminology Relating to Thermophysical PropertiesE2040 Test Method for Mass Scale Calibration of Thermo-gravimetric Analyzers3. Terminology3.1 DefinitionsTechnical terms used in this document aredefined in Terminologies E473 and E1142, incl
14、uding Celsius,Curie temperature, derivative, Kelvin, magnetictransformation, onset point, thermogravimetry, and thermo-gravimetric analyzer.1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on Calo-rimetry
15、and Mass Loss.Current edition approved Oct. 1, 2017. Published October 2017. Originallyapproved in 1993. Last previous edition approved in 2014 as E1582 14. DOI:10.1520/E1582-17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer service at serviceastm.org.
16、For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis
17、international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT
18、) Committee.13.1.1 magnetic reference temperature, nthe observed tem-perature at which a change in the magnetic properties of amaterial in a magnetic field produces an apparent mass change.This temperature is read from the dynamic thermogravimetrycurve as the point of intersection of the extrapolate
19、d highertemperature portion of the base line with a tangent drawn to thepoint of greatest slope of apparent mass-change curve. Thistemperature most closely represents the Curie Point, that pointon the mass change curve where the magnetic effect of thestandard material has disappeared completely (see
20、 Fig. 1).3.1.1.1 DiscussionThe position of the magnet and thedesign of the instrument will affect the direction of the masschange.4. Summary of Practice4.1 This test method provides a set of different proceduressince thermogravimetric apparatus is often of significantlydiffering design.4.2 Calibrati
21、on of Analyzers Using Magnetic TransitionStandards:4.2.1 In this procedure, the apparent mass change of one ormore of the magnetic transition standards is obtained under thenormal operating conditions of the instrument. The extrapo-lated endpoint temperature (see Fig. 1) is determined andcompared wi
22、th the established transition temperature for thematerial. The difference provides an adjustment or calibrationthat may be applied to the temperature scale of the instrument.4.2.2 The apparent mass change of the magnetic transitionmaterials is caused by the magnetic (ferromagnetic) to non-magnetic (
23、paramagnetic) transition in the presence of a mag-netic field.4.3 Calibration of Analyzers That Have SimultaneousThermogravimetry-Differential Scanning Calorimeter orThermogravimetry-Differential Thermal Analysis CapabilityThese instruments may be calibrated using melting temperaturestandards follow
24、ing Practice E967.5. Significance and Use5.1 Thermogravimetric analyzers are used to characterize abroad range of materials. In most cases, one of the desiredvalues to be assigned in thermogravimetric measurements isthe temperature at which significant changes in specimen massoccur. Therefore, the t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME15822017STANDARDTESTMETHODFORTEMPERATURECALIBRATIONOFTHERMOGRAVIMETRICANALYZERS 分析 温度 校准 标准 试验 方法

链接地址:http://www.mydoc123.com/p-528984.html