ASTM E1421-1999(2009) Standard Practice for Describing and Measuring Performance of Fourier Transform Mid-Infrared (FT-MIR) Spectrometers Level Zero and Level One Tests《傅里叶转换中红外线(F.pdf
《ASTM E1421-1999(2009) Standard Practice for Describing and Measuring Performance of Fourier Transform Mid-Infrared (FT-MIR) Spectrometers Level Zero and Level One Tests《傅里叶转换中红外线(F.pdf》由会员分享,可在线阅读,更多相关《ASTM E1421-1999(2009) Standard Practice for Describing and Measuring Performance of Fourier Transform Mid-Infrared (FT-MIR) Spectrometers Level Zero and Level One Tests《傅里叶转换中红外线(F.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1421 99 (Reapproved 2009)Standard Practice forDescribing and Measuring Performance of FourierTransform Mid-Infrared (FT-MIR) Spectrometers: Level Zeroand Level One Tests1This standard is issued under the fixed designation E 1421; the number immediately following the designation indica
2、tes the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes two levels of tests
3、to measure theperformance of laboratory Fourier transform mid-infrared(FT-MIR) spectrometers equipped with a standard sampleholder used for transmission measurements.1.2 This practice is not directly applicable to Fourier trans-form infrared (FT-IR) spectrometers equipped with variousspecialized sam
4、pling accessories such as flow cells or reflec-tance optics, nor to Fourier transform near-infrared (FT-NIR)spectrometers, nor to FT-IR spectrometers run in step scanmode.1.2.1 If the specialized sampling accessory can be removedand replaced with a standard transmission sample holder, thenthis pract
5、ice can be used. However, the user should recognizethat the performance measured may not reflect that which isachieved when the specialized accessory is in use.1.2.2 If the specialized sampling accessory cannot be re-moved, then it may be possible to employ a modified versionof this practice to meas
6、ure spectrometer performance. The useris referred to Guide E 1866 for a discussion of how these testsmay be modified.1.2.3 Spectrometer performance tests for FT-NIR spectrom-eters are described in Practice E 1944.1.2.4 Performance tests for dispersive MIR instruments aredescribed in Practice E 932.1
7、.2.5 For FT-IR spectrometers run in a step scan mode,variations on this practice and information provided by theinstrument vendor should be used.2. Referenced Documents2.1 ASTM Standards:2E 131 Terminology Relating to Molecular SpectroscopyE 932 Practice for Describing and Measuring Performanceof Di
8、spersive Infrared SpectrometersE 1866 Guide for Establishing Spectrophotometer Perfor-mance TestsE 1944 Practice for Describing and Measuring Performanceof Laboratory Fourier Transform Near-Infrared (FT-NIR)Spectrometers: Level Zero and Level One Tests3. Terminology3.1 DefinitionsFor definitions of
9、terms used in this prac-tice, refer to Terminology E 131. All identifications of spectralregions and absorption band positions are given in wavenum-bers (cm1), and spectral energy, transmittance, and absorbanceare signified in equations by the letters E, T, and A respectively.The ratio of two transm
10、ittance or absorbance values, and theratio of energy levels at two different wavenumbers aresignified by the letter R. A subscripted number signifies aspectral position in wavenumbers (for example, A3082, theabsorbance at 3082 cm1).3.1.1 level one (1) test, na simple series of measurementsdesigned t
11、o provide quantitative data on various aspects ofinstrument performance and information on which to base thediagnosis of problems.3.1.2 level zero (0) test, na routine check of instrumentperformance, that can be done in a few minutes, designed tovisually detect significant changes in instrument perf
12、ormanceand provide a database to determine instrument function overtime.4. Significance and Use4.1 This practice permits an analyst to compare the generalperformance of an instrument on any given day with the priorperformance of an instrument. This practice is not necessarilymeant for comparison of
13、different instruments with each othereven if the instruments are of the same type and model. Thispractice is not meant for comparison of the performance of oneinstrument operated under differing conditions.5. Test Conditions5.1 Operating ConditionsA record should be kept todocument the operating con
14、ditions selected so that they can beduplicated. In obtaining spectrophotometric data, the analyst1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.03 on Infrared and Near Infrared Spect
15、roscopy.Current edition approved March 1, 2009. Published March 2009. Originallyapproved in 1991. Last previous edition approved in 2004 as E 1421 99 (2004).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of AS
16、TMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.must select proper instrumental operating conditions such aswarm-up time, purge rate, and
17、 beam splitter alignment in orderto realize satisfactory instrument performance. Operating con-ditions for individual instruments are best obtained from themanufacturers literature because of variations with instrumentdesign. It should be noted that many FT-IR instruments aredesigned to work best wh
18、en left on or in the standby mode.Also note that spectrometers are to be tested only within theirrespective wavenumber ranges.NOTE 1This practice is designed to be used in situations where thedetector is not saturated. In some instruments, with some combinations ofoptics and detectors, the detector
19、electronics are saturated with an emptybeam. These instruments are designed to have the infrared beam attenu-ated in the spectrometer or sample compartment to eliminate detectorsaturation. Consult your instrument manual or discuss appropriate attenu-ation techniques with the instrument vendor.5.2 Th
20、e environment in which a spectrometer is operatedcan affects its performance. Spectrometers should only beoperated in environments consistent with manufacturers rec-ommendations. Changes in the instrument environment includ-ing variations in temperature, vibration or sound levels, elec-trical power
21、or magnetic fields should be recorded.5.3 Instrumental characteristics can influence these mea-surements in several ways.5.3.1 Vignetting of the beam reduces the transmittancevalue measured in nonabsorbing regions, and on most instru-ments can change the apparent wavenumber scale by a smallamount, u
22、sually less than 0.1 cm1. Make sure that the filmholder does not vignet the beam.5.3.2 Focus changes can also change transmittance values,so the sample should be positioned in approximately the samelocation in the sample compartment each time.5.3.3 The angle of acceptance (established by the f numbe
23、r)of the optics between the sample and detector significantlyaffects apparent transmittance. Changes to the optical pathincluding the introduction of samples can alter the acceptanceangle.5.3.4 Heating of the sample by the beam or by the highertemperatures which exist inside most spectrometers chang
24、esabsorbances somewhat, and even changes band ratios andlocations slightly. Allow the sample to come to thermalequilibrium before measurement.5.4 The recommended sample of matte-finish polystyreneused for these tests is approximately 38-m (1.5-mil) thick filmmounted on a card. The sample is mounted
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME142119992009STANDARDPRACTICEFORDESCRIBINGANDMEASURINGPERFORMANCEOFFOURIERTRANSFORMMIDINFRAREDFTMIRSPECTROMETERSLEVELZEROANDLEVELONETESTS

链接地址:http://www.mydoc123.com/p-528666.html