ASTM E1303-1995(2005) Practice for Refractive Index Detectors Used in Liquid Chromatography《液相色谱法用折射指数检测器规程》.pdf
《ASTM E1303-1995(2005) Practice for Refractive Index Detectors Used in Liquid Chromatography《液相色谱法用折射指数检测器规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1303-1995(2005) Practice for Refractive Index Detectors Used in Liquid Chromatography《液相色谱法用折射指数检测器规程》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1303 95 (Reapproved 2005)Standard Practice forRefractive Index Detectors Used in Liquid Chromatography1This standard is issued under the fixed designation E 1303; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the y
2、ear of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers tests used to evaluate the perfor-mance and to list certain descriptive specifications of
3、 arefractive index (RI) detector used as the detection componentof a liquid chromatographic (LC) system.1.2 This practice is intended to describe the performance ofthe detector both independent of the chromatographic system(static conditions, without flowing solvent) and with flowingsolvent (dynamic
4、 conditions).1.3 The values stated in SI units are to be regarded as thestandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and de
5、termine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 386 Practice for Data Presentation Relating to High-Resolution Nuclear Magnetic Resonance (NMR) Spectros-copy3. Significance and Use3.1 Although it is possible to observe and measure each o
6、fseveral characteristics of a detector under different and uniqueconditions, it is the intent of this practice that a complete set ofdetector test results should be obtained under the same oper-ating conditions. It should also be noted that to specifycompletely a detectors capability, its performanc
7、e should bemeasured at several sets of conditions within the useful rangeof the detector.3.2 The objective of this practice is to test the detector underspecified conditions and in a configuration without an LCcolumn. This is a separation independent test. In certaincircumstances it might also be ne
8、cessary to test the detector inthe separation mode with an LC column in the system, and theappropriate concerns are also mentioned. The terms and testsdescribed in this practice are sufficiently general so that theymay be adapted for use at whatever conditions may be chosenfor other reasons.4. Noise
9、, Drift, and Flow Sensitivity4.1 Descriptions of Terms Specific to This Standard:4.1.1 short term noisethis noise is the mean amplitude inrefractive index units (RIU) for random variations of thedetector signal having a frequency of one or more cycles perminute. Short term noise limits the smallest
10、signal detectableby an RI detector, limits the precision attainable, and sets thelower limit on the dynamic range. This noise corresponds toobserved noise of the RI detector only. (The actual noise of theLC system may be larger or smaller than the observed value,depending upon the method of data col
11、lection, or signalmonitoring of the detector, since observed noise is a function ofthe frequency, speed of response and the band width of therecorder or other electronic circuit measuring the detectorsignal.)4.1.2 long term noisethis noise is the maximum ampli-tude in RIU for random variations of th
12、e detector signal withfrequencies between 6 and 60 cycles per h (0.1 and 1.0 cyclesper min). It represents noise that may be mistaken for alate-eluting peak. This noise corresponds to the observed noiseonly and may not always be present.4.1.3 driftthe average slope of the long term noise enve-lope e
13、xpressed in RIU per hour as measured over a period of1h.4.1.4 staticrefers to the noise and drift measured underconditions of no flow.4.1.5 dynamicrefers to the noise and drift measured at aflow rate of 1.0 mL/min.4.1.6 flow sensitivitythe rate of change of signal displace-ment (in RIU) vs flow rate
14、 (in mL/min) resulting from stepchanges in flow rate calculated at 1 mL/min as described in4.3.12.4.2 Test Conditions:4.2.1 The same test solvent must be used in both sample andreference cells. The test solvent used and its purity should bespecified. Water equilibrated with the laboratory atmosphere
15、1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Chromatography and is the direct responsibility of SubcommitteeE13.19 on Chromatography.Current edition approved Feb. 1, 2005. Published March 2005. Originallyapproved in 1989. Last previous edition approved
16、in 2000 as E 1303 95 (2000).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 10
17、0 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.containing minimum impurities is the preferred test solvent formeasuring noise and drift. Water for this purpose (preferablypurified by distillation, deionization, or reverse osmosis)should be drawn, filtered through a
18、 0.45-m filter, and allowedto stand in a loosely covered container for several hours atambient temperature in the laboratory in which testing is to becarried out. This will ensure complete equilibration of thewater with the gases in the laboratory atmosphere.NOTE 1It is essentially impossible to mai
19、ntain a constant RI value ofde-gassed water and of very dilute samples in de-gassed water. This is dueto the fact that the difference in refractive index between completelyde-gassed water and atmosphere-equilibrated water is 1.5 3 106RIU.3Thus, small differences in the concentration of dissolved gas
20、es betweenthe sample and the trapped reference can lead to significant errors inmeasurement of solutions where the expected difference in RI due tosolute is of the order of 106RIU or less. Therefore, in order to minimizeerror in determining samples with small RIU differences between them,atmosphere-
21、equilibrated water (5.2.1) is recommended as the solvent fordetermining linearity and minimum detectability (Section 5).4.2.2 The detector should be located at the test site andswitched on at least 24 h prior to the start of testing. Somedetectors provide an oven to thermostat the optics assembly.Th
22、e oven should be set at a suitable temperature, following themanufacturers recommendations, and this temperature shouldbe noted and maintained throughout the test procedures.4.2.3 Linearity and speed of response of the recorder orother data acquisition device used should be such that it doesnot dist
23、ort or otherwise interfere with the performance of thedetector.4If additional amplifiers are used between the detectorand the final readout device, their characteristics should alsofirst be established.4.3 Methods of Measurement:4.3.1 Connecta1m(39.37 in.) length of clean, dry,stainless steel tubing
24、 of 0.25 mm (0.009 to 0.01 in.) insidediameter in place of the analytical column. The tubing can bestraight or coiled to minimize the space requirement. Thetubing should terminate in standard low dead volume fittings toconnect with the detector and to the pump. Commercialchromatographs may already c
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME130319952005PRACTICEFORREFRACTIVEINDEXDETECTORSUSEDINLIQUIDCHROMATOGRAPHY 色谱 折射 指数 检测器 规程 PDF

链接地址:http://www.mydoc123.com/p-528365.html