ASTM E1270-1988(2014) Standard Test Method for Equal Arm Balances《等臂天平的标准试验方法》.pdf
《ASTM E1270-1988(2014) Standard Test Method for Equal Arm Balances《等臂天平的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1270-1988(2014) Standard Test Method for Equal Arm Balances《等臂天平的标准试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1270 88 (Reapproved 2014)Standard Test Method forEqual Arm Balances1This standard is issued under the fixed designation E1270; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in par
2、entheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONThis test method is designed to test balances whose lever-arm ratio is substantially equal to unity.Although largely superseded by new technologies,
3、 equal-arm balances retain a special niche for veryhigh precision weighing of larger samples (usually greater than 1 kg) as well as objects with largebuoyancy (such as gas bottles). Balances of this type can range from simple instruments of moderateprecision (1:10 000) to extremely high precision de
4、vices with precision of 1:10 000 000 or better. Anumber of accessory devices may be included for assisting in the weighing process. These devicesmay contribute to errors as well as can the basic lever mechanism. This method is designed to test theentire instrument including the accessories.1. Scope1
5、.1 This test method can be used for testing equal-armbalances of any capacity and sensitivity. The testing procedureshould enable the user to characterize his instrument suffi-ciently to determine whether or not it is suitable for the purposefor which it is to be used.1.2 The characteristics to be e
6、xamined include:1.2.1 Sensitivity at all loads,1.2.2 Lever arm ratio,1.2.3 Damping ratio (for instruments without accessorydampers),1.2.4 Period of oscillation,1.2.5 Precision, and1.2.6 Linearity and calibration of accessory devices thatprovide on-scale indication of weight.1.3 This standard does no
7、t purport to address all of thesafety concerns associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety andhealth practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E617 Speci
8、fication for Laboratory Weights and PrecisionMass Standards3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 capacitymaximum load recommended by the manu-facturer. Usually, the capacity refers to the maximum load oneach pan simultaneously.3.1.2 readabilityvalue of the smallest u
9、nit of weightwhich can be read. This may include the estimation of somefraction of a scale division or, in the case of a digital display,will represent the minimum value of the least significant digit.3.1.3 sensitivitysmallest value of weight which will causea change of indication which can be deter
10、mined by the user.This may be independent of the readability because of thechoice of the reading device used. For example, a magnifyingglass may be used in conjunction with a reading scale toobserve a sensitivity not readily determined without the mag-nifying glass.3.1.4 precisionrepeatability of th
11、e balance indication withthe same load under essentially the same conditions. The moreclosely the measurements are grouped, the smaller the index ofprecision will be. The precision should be measured underenvironmental conditions that represent the conditions underwhich the balance is normally used.
12、3.1.5 accuracydegree of agreement of the measurementwith the true value of the magnitude of the quantity measured.3.1.6 linearitycharacteristic of a direct reading device. If adevice is linear, calibration at 2 points (for example, 0 andfull-scale) calibrates the device (for example, 2 points deter-
13、mine a straight line); if a device is nonlinear, additional pointsare needed (perhaps a great many).1This test method is under the jurisdiction of ASTM Committee E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E41.06 onWeighing Devices.Current edition approved Nov. 1, 201
14、4. Published November 2014. Originallyapproved in 1988. Last previous edition approved in 2008 as E127088 (2008).DOI: 10.1520/E1270-88R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume
15、information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.1.7 standard weightany weight whose mass is given.Since weights are not always available with documentedcor
16、rections, weights defined by class (see Specification E617)may be used if the class has sufficiently small tolerance limitsand there is an understanding that errors perceived as beinginstrumental could be attributed to incorrectly adjustedweights.3.1.8 off-center errorsdifferences in indicated weigh
17、twhen a sample is shifted to various positions on the weighingarea of the weighing pan. No separate test is described.3.1.9 full-scale calibration of an accessory deviceindicated reading at equilibrium of an accessory device when astandard weight equal to the full-scale range of the deviceisplaced o
18、n the sample pan. Usually, some means is providedby the manufacturer to adjust the full-scale to match the weightof the standard.4. Summary of Test Method4.1 Throughout this test method, the instrument is to beused in the manner for which it is intended by the manufac-turer. All measurements are mad
19、e with weights whose valuesare sufficiently well known for the purpose of the user. Thenominal value of the weights used will be determined by thecapacity and rated sensitivity of the balance as well as by theresolution and range of the accessory reading devices.5. Significance and Use5.1 This test
20、method should enable the user of the balance tointerpret data determined thereon in terms of accuracy andprecision. It should be helpful in using a particular instrumentto best advantage. Weaknesses as well as strengths shouldbecome apparent. It is not the intention of this test method tocompare sim
21、ilar instruments of different manufacture butrather to assist in choosing an instrument which will meet theneeds of the user.6. Apparatus6.1 Standard WeightsIndividual or summations of weightsequal to approximately14,12,34 and the total capacity.6.2 Tare WeightsWeights of the same denominations asth
22、e standard weights but not necessarily calibrated.6.3 Calibrating WeightsBalances equipped with acces-sory devices such as sliding beam weights, chainweights,optical scales or electrical transducers require small standardweights equal to the full-scale reading as well as smallerweights suitable for
23、calibrating intermediate points between thezero and full-scale points of the devices. Summations of smallstandards can be used for this purpose.6.4 Stop Watch:6.5 Aroom-temperature thermometer with a resolution of atleast 1C.7. Preparation of Apparatus7.1 Place the instrument in the location at whic
24、h it is to betested. If electrically operated, plug in the line cord to the typeof socket recommended by the manufacturer.7.2 Place the standard weights near (or within) the instru-ment.7.3 Place the thermometer on the bench in position so that itmay be read without being touched.7.4 Make sure that
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME127019882014STANDARDTESTMETHODFOREQUALARMBALANCES 天平 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-528312.html