《ASTM E1140-1995(2010)e1 Standard Practice for Testing Nitrogen Phosphorus Thermionic Ionization Detectors for Use In Gas Chromatography《测试气相色谱法用氮 磷热离子电离探测器的标准操作规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1140-1995(2010)e1 Standard Practice for Testing Nitrogen Phosphorus Thermionic Ionization Detectors for Use In Gas Chromatography《测试气相色谱法用氮 磷热离子电离探测器的标准操作规程》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1140 95 (Reapproved 2010)1Standard Practice forTesting Nitrogen/Phosphorus Thermionic IonizationDetectors for Use In Gas Chromatography1This standard is issued under the fixed designation E1140; the number immediately following the designation indicates the year oforiginal adoption or,
2、 in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEEditorial corrections were made to Footnote 3 in November 2010.1. Scope1.1 This practic
3、e covers testing the performance of anitrogen/phosphorus thermionic ionization detector (NPD)used as the detection component of a gas chromatographicsystem.1.2 This practice applies to an NPD that employs a heatedalkali metal compound and emits an electrical charge from thatsolid surface.1.3 This pr
4、actice addresses the operation and performanceof the NPD independently of the chromatographic column.However, the performance is specified in terms that the analystcan use to predict overall system performance when thedetector is coupled to the column and other chromatographiccomponents.1.4 For gene
5、ral chromatographic procedures, Practice E260should be followed except where specific changes are recom-mended in this practice for the use of a nitrogen/phosphorus(N/P) thermionic detector.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It i
6、s theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific safetyinformation, see Section 5, Hazards.2. Referenced Documents2.1 ASTM Standards:2E260 Practice for Packed C
7、olumn Gas ChromatographyE355 Practice for Gas Chromatography Terms and Rela-tionships2.2 CGA Standards:3CGA P-1 Safe Handling of Compressed Gases in Contain-ersCGA G-5.4 Standard for Hydrogen Piping Systems at Con-sumer LocationsCGA P-9 The Inert Gases: Argon, Nitrogen and HeliumCGA V-7 Standard Met
8、hod of Determining Cylinder ValveOutlet Connections for Industrial Gas MixturesCGA P-12 Safe Handling of Cryogenic LiquidsHB-3 Handbook of Compressed Gases3. Terminology3.1 Definitions:3.1.1 For definitions of gas chromatography and its variousterms, see Practice E355.3.2 Definitions of Terms Specif
9、ic to This Standard:3.2.1 driftthe average slope of the noise envelope ex-pressed in amps/h as measured over12 h.3.2.2 linear rangerange of mass flow rates of nitrogen orphosphorus in the carrier gas, over which the sensitivity of thedetector is constant to within 5 % as determined from thelinearity
10、 plot.3.2.3 minimum detectabilitythe mass flow rate of nitrogenor phosphorus in the carrier gas that gives a detector signalequal to twice the noise level.3.2.4 noise (short term)the amplitude, expressed in am-peres, of the baseline envelope that includes all randomvariations of the detector signal
11、of a frequency greater than onecycle per minute.3.2.5 selectivitythe ratio of the response per gram ofnitrogen or phosphorus in the test substance to the response pergram of carbon in octadecane.4. Significance and Use4.1 Although it is possible to observe and measure each ofthe several characterist
12、ics of a detector under different andunique conditions, it is the intent of this practice that a1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.19 on Separation Science.Current editio
13、n approved Nov. 1, 2010. Published December 2010. Originallyapproved in 1986. Last previous edition approved in 2005 as E1140 95 (2005).DOI: 10.1520/E1140-95R10E01.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Boo
14、k of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from Compressed Gas Association (CGA), 4221 Walney Rd., 5thFloor, Chantilly, VA 20151-2923, http:/.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, P
15、A 19428-2959, United Splete set of detector specifications be obtained at the sameoperating conditions, including geometry, flow rates, andtemperatures. To specify a detectors capability completely, itsperformance should be measured at several sets of conditionswithin the useful range of the detecto
16、r. The terms and testsdescribed in this practice are sufficiently general so that theymay be used under any chosen conditions.4.2 Linearity and speed of response of the recorder shouldbe such that it does not distort or otherwise interfere with theperformance of the detector. Effective recorder resp
17、onse shouldbe sufficiently fast so that its effect on the sensitivity ofmeasurement is negligible. If additional amplifiers are usedbetween the detector and the final readout device, theircharacteristics should first be established.5. Hazards5.1 Gas Handling SafetyThe safe handling of compressedgase
18、s and cryogenic liquids for use in chromatography is theresponsibility of every laboratory. The Compressed Gas Asso-ciation (CGA), a member group of specialty and bulk gassuppliers, publishes the following guidelines to assist thelaboratory chemist to establish a safe work environment.Applicable CGA
19、 publications include: CGA P-1, CGA G-5.4,CGA P-9, CGA V-7, CGA P-12, and HB-3.6. Application6.1 The N/P thermionic detector is an element-specificionization detector that is essentially a major modification ofthe flame ionization detector (FID). As in the normal FID, itmeasures increase in ionizati
20、on current passing between twoelectrodes, one of which is polarized relative to the other.Usually these are the inorganic salt source and the collector,with one often being at ground potential.6.2 The mechanism of the detector will only be discussedbriefly in this practice partly because full unders
21、tanding of thedetector is not presently available and partly because thesubstantial differences in bead chemistry, detector geometry,and bead heating mechanism prevent a singular view beinggiven.6.3 The addition of a heated alkali metal compound in thedetector area causes enhancement of the response
22、 for carbon-nitrogen and carbon-phosphorus bonds. In addition, the selec-tivity of response can be further enhanced when the bead iselectrically heated. Lower hydrogen and air flow rates thatdiminish the normal flame ionization response for hydrocarboncompounds can be used. This selective enhancemen
23、t allows theNPD to be used for the detection of very small quantities ofnitrogen- and phosphorus-containing compounds without in-terference from the signal of other molecular species.6.4 The selective response to C-N and C-Pbonds means thatthe detector is not suitable for permanent gas or elementaln
24、itrogen or phosphorus analysis in the true definition of theterm. It should be noted, however, that some volatile inorganicphosphorous compounds do give a strong response with thisdetector, comparable to that of organophosphorus compounds.7. Detector Construction7.1 There is a wide variation in the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME114019952010E1STANDARDPRACTICEFORTESTINGNITROGENPHOSPHORUSTHERMIONICIONIZATIONDETECTORSFORUSEINGASCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-528002.html