ASTM E1137 E1137M-2008(2014) Standard Specification for Industrial Platinum Resistance Thermometers《工业用铂电阻温度计的标准规格》.pdf
《ASTM E1137 E1137M-2008(2014) Standard Specification for Industrial Platinum Resistance Thermometers《工业用铂电阻温度计的标准规格》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1137 E1137M-2008(2014) Standard Specification for Industrial Platinum Resistance Thermometers《工业用铂电阻温度计的标准规格》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1137/E1137M 08 (Reapproved 2014)Standard Specification forIndustrial Platinum Resistance Thermometers1This standard is issued under the fixed designation E1137/E1137M; the number immediately following the designation indicates the yearof original adoption or, in the case of revision, t
2、he year of last revision. A number in parentheses indicates the year of last reapproval.A superscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers the requirements for metal-sheathed industrial platinum resistance thermometers (
3、PRTs)suitable for direct immersion temperature measurement. Itapplies to PRTs with an average temperature coefficient ofresistance between 0 and 100C of 0.385 %C and nominalresistance at 0C of 100 or other specified value. Thisspecification covers PRTs suitable for all or part of thetemperature rang
4、e 200 to 650C. The resistance-temperaturerelationship and tolerances are specified as well as physical,performance, and testing requirements.1.2 The values of temperature in this specification are basedon the International Temperature Scale of 1990 (ITS-90).21.3 The values stated in inch-pound units
5、 or SI (metric)units may be regarded separately as standard. The values statedin each system are not exact equivalents, and each system shallbe independent of the other.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility o
6、f the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3A269 Specification for Seamless and Welded AusteniticStainless Steel Tubing for General ServiceB167 Spe
7、cification for Nickel-Chromium-Iron Alloys (UNSN06600, N06601, N06603, N06690, N06693, N06025,N06045, and N06696), Nickel-Chromium-Cobalt-Molybdenum Alloy (UNS N06617), and Nickel-Iron-Chromium-Tungsten Alloy (UNS N06674) Seamless Pipeand TubeE344 Terminology Relating to Thermometry and Hydrom-etryE
8、644 Test Methods for Testing Industrial Resistance Ther-mometersE1652 Specification for Magnesium Oxide and AluminumOxide Powder and Crushable Insulators Used in theManufacture of Base Metal Thermocouples, Metal-Sheathed Platinum Resistance Thermometers, and NobleMetal Thermocouples3. Terminology3.1
9、 DefinitionsFor definitions of terms used in this speci-fication see Terminology E344.3.2 Definitions of Terms Specific to This Standard:3.2.1 connecting wire end closure, nmoisture barrier atthe connecting wire end of the sheath.3.2.1.1 DiscussionThe closure is intended to provide aseal sufficient
10、to prevent the sensors insulation resistance fromdropping below the minimum requirements.3.2.2 connecting wires, nwires that run from the elementthrough the connecting wire end closure and external to thesheath.3.2.3 excitation, nelectrical current passing through theelement.3.2.4 g-level, naccelera
11、tion of an object relative to thelocal acceleration of gravity.3.2.4.1 DiscussionFor example, a g-level of 5 is equiva-lent to an acceleration of approximately 5 9.8 m/s2= 49.0m/s2.3.2.5 minimum immersion length, ndepth that a thermom-eter should be immersed, in a uniform temperatureenvironment, suc
12、h that further immersion does not produce achange in indicated temperature greater than the specifiedtolerance.3.2.6 PRT design, ngeneric term used to differentiatebetween different PRT construction details, such as elementand connecting wire construction, insulation methods, sealingtechniques, and
13、mounting methods (for example, spring loadedor direct mounting).1This specification is under the jurisdiction of ASTM Committee E20 onTemperature Measurement and is the direct responsibility of Subcommittee E20.03on Resistance Thermometers.Current edition approved Dec. 1, 2014. Published December 20
14、14. Originallyapproved in 1987. Last previous edition approved in 2008 as E1137 08. DOI:10.1520/E1137_E1137M-08R14.2Preston-Thomas, H., “The International Temperature Scale of 1990 (ITS-90),”Metrologia , Vol 27, No. 1. 1990, pp 3 10, ibid, Vol 27, No. 2, 1990, p1073For referenced ASTM standards, vis
15、it the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-29
16、59. United States13.2.7 self-heating, nchange in temperature of the elementcaused by the heating effect of the excitation.3.2.8 sheath, ncylindrical metal tube with an integralwelded closure at the end in which the element is located.4. Significance and Use4.1 This specification is written to provid
17、e commonterminology, resistance versus temperature characteristics, ac-curacy classification, and inspection requirements for a speci-fied configuration of a typical industrial platinum resistancethermometer (PRT).4.2 This specification may be used as part of the documen-tation to support negotiatio
18、ns for the purchase and discussionof such thermometers.5. Classification of Tolerances5.1 The PRT shall conform to the resistance-temperaturerelation (see 9.2.1) within the following tolerances:Grade A 560.1310.0017? t ? #C (1)Grade B 560.2510.0042 ? t ? #C (2)where:| t | = value of temperature with
19、out regard to sign, C.5.1.1 The tolerances are given in Table 1 for a PRT with anominal resistance of 100 at 0C.6. Ordering Information6.1 The purchase order documents shall specify the follow-ing information to ensure that the PRT is adequately described:6.1.1 The number of this specification,6.1.2
20、 Sheath diameter and overall length (see Fig. 1),6.1.3 Sheath material,6.1.4 Minimum and maximum sensed temperature,6.1.5 Maximum and minimum temperature at connectingwire end closure,6.1.6 Connection configuration; 2-Wire, 3-Wire, 4-Wire(potentiometric), and compensating loop (4-Wire) (see Fig. 2),
21、6.1.7 Tolerance, (Grade A, or Grade B),6.1.8 Nominal resistance at 0C (100 unless otherwisespecified), and6.1.9 Serial Number identification requirement (mandatoryif an individual calibration or test record will be maintained).7. Materials and Manufacture7.1 All materials used shall be in accordance
22、 with thefollowing requirements:7.1.1 Sheath MaterialsFor temperatures not exceeding480C, austenitic stainless steel tubing, conforming to Speci-fication A269. For temperatures not exceeding 650C, high-nickel alloy tubing, conforming to Specification B167.7.1.2 Sensing ElementSensing element shall b
23、e platinum.7.1.3 InsulationThe insulating material within the PRTshall be compatible with the temperature range 200 to 650Cor as specified in 6.1.4. Magnesium oxide (MgO) and alumi-num oxide (Al2O3) powders and crushable insulators conform-ing to Specification E1652 satisfy this requirement.7.1.4 Co
24、nnecting Wire End Closure MaterialsClosurematerials shall provide a barrier against water and other liquidsand generally prevent the penetration of water vapor. Anymaterial used shall be compatible with the ambient tempera-tures specified for the application (see 6.1.5).7.1.4.1 Typically, epoxy mate
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME1137E1137M20082014STANDARDSPECIFICATIONFORINDUSTRIALPLATINUMRESISTANCETHERMOMETERS 工业 铂电阻 温度计 标准

链接地址:http://www.mydoc123.com/p-527996.html