1、Designation: E 1079 05Standard Practice forCalibration of Transmission Densitometers1This standard is issued under the fixed designation E 1079; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in
2、 parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice2covers the calibration of transmissiondensitometers used to perform radiographic film density mea-surements (see Note 1).1.2 This
3、 standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.NOTE 1For further infor
4、mation on the design and use of densitom-eters, the following literature is suggested as additional backgroundinformation: ISO 52:2001, ISO 51:2001, and ISO 14807:2001.2. Referenced Documents2.1 ASTM Standards:3E 1316 Terminology for Nondestructive Examinations2.2 ISO Standards:4ISO 52:2001 Photogra
5、phy-Density MeasurementsPart2: Geometric Conditions for Transmission DensityISO 51:2001 Photography-Density MeasurementsPart1: Terms, Symbols, and NotationsISO 14807:2001 Photography-Transmission and ReflectionDensitometersMethod for Determining Performance3. Terminology3.1 DefinitionsFor definition
6、s of terms used in this prac-tice, see Terminology E 1316.4. Significance and Use4.1 This practice provides a means for calibrating transmis-sion densitometers used for the measurement of radiographicfilm density. A transmission densitometer calibrated in accor-dance with this practice provides the
7、assurance that accuratedensity values of radiographs are obtained.5. Apparatus5.1 Apparatus should consist of the following:5.1.1 A calibrated step tablet shall be used. The step tabletmay be a NIST X-ray Step Tablet (Standard Reference MaterialSRM 1001)5or alternately a step tablet from another sup
8、plierwhich is traceable to the NIST SRM 1001 X-ray Step Tablet.The step tablet shall have at least five step densities rangingfrom 0.9 through 4.1. The step tablet may have additional stepdensities less than 0.9 and greater than 4.1. A calibrationcertificate shall be provided with the step tablet in
9、dicating thetablet ID and recorded values for each step density. Forsuppliers of step tablets other than NIST, the certificate shallindicate conformance of traceability to NIST instrumentationused in the calibration process, applicable ANSI standardsused, verification of measurement on a NIST SRM 10
10、01 steptablet, the ID number of the SRM 1001 step tablet, andcalibration date of the step tablet. Precautions should be takenin the storage, handling, and use of the step tablet. In the eventit becomes scratched, blemished, or exhibits other signs ofdeleterious wear, it should be replaced immediatel
11、y. The SRM1001 (or alternate, if used) step tablet shall be replaced fouryears from the date of first use.65.1.2 Transmission Densitometers, with either direct-scalereadout or digital readout displays specifically manufacturedfor the purpose of measuring the range of film densitiesdescribed in 5.1.1
12、 may be used.5.1.3 Manufacturers Operating Instructions for Appropri-ate Transmission Densitometer.6. Calibration6.1 Full-scale linearity calibration should be performed atleast every 90 days during use as follows:6.1.1 Use the manufacturers recommended warm-up timeto stabilize circuitry before star
13、ting the calibration procedure orthe periodic verification checks described in Section 8. Adjustthe “0” reading of the densitometer after the warm-up period.1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommittee E07.
14、01 onRadioglogy (X and Gamma) Method.Current edition approved January 1, 2005. Published January 2005. Originallyapproved in 1985. Last previous edition approved in 2000 as E 1079 - 00.2For ASME Boiler and Pressure Vessel Code applications see related PracticeSE-1079 in Section II of that Code.3For
15、referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4Available from American National Standards Institute (ANSI), 25 W
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME10792005STANDARDPRACTICEFORCALIBRATIONOFTRANSMISSIONDENSITOMETERS 透射 密度计 校准 标准 规程 PDF

链接地址:http://www.mydoc123.com/p-527865.html