《ASTM E1001-2006 Standard Practice for Detection and Evaluation of Discontinuities by the Immersed Pulse-Echo Ultrasonic Method Using Longitudinal Waves《用纵波浸入脉冲回波超声法检测与评估间断性的标准实施规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1001-2006 Standard Practice for Detection and Evaluation of Discontinuities by the Immersed Pulse-Echo Ultrasonic Method Using Longitudinal Waves《用纵波浸入脉冲回波超声法检测与评估间断性的标准实施规范》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1001 06Standard Practice forDetection and Evaluation of Discontinuities by theImmersed Pulse-Echo Ultrasonic Method Using LongitudinalWaves1This standard is issued under the fixed designation E 1001; the number immediately following the designation indicates the year oforiginal adopti
2、on or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. S
3、cope1.1 This practice describes procedures for the ultrasonicexamination of bulk materials or parts by transmitting pulsed,longitudinal waves through a liquid couplant into the materialand observing the indications of reflected waves (see Fig. 1). Itcovers only examinations in which one search unit
4、is used asboth transmitter and receiver (pulse-echo) and in which the partor material being examined is coupled to the test part by aliquid column or is totally submerged in the couplant (eithermethod is considered to be immersion testing). This practiceincludes general requirements and procedures w
5、hich may beused for detecting discontinuities and for making a relative orapproximate evaluation of the size of discontinuities.1.2 This practice replaces Practice E 214 and provides moredetailed procedures for the selection, calibration, and operationof an inspection system and for evaluation of th
6、e indicationsobtained.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior
7、to use.2. Referenced Documents2.1 ASTM Standards:2C 1212 Practice for Fabricating Ceramic Reference Speci-mens Containing Seeded VoidsC 1336 Practice for Fabricating Non-Oxide Ceramic Refer-ence Specimens Containing Seeded InclusionsE 127 Practice for Fabricating and Checking AluminumAlloy Ultrasoni
8、c Standard Reference BlocksE 317 Practice for Evaluating Performance Characteristicsof Ultrasonic Pulse-Echo Testing Instruments and Systemswithout the Use of Electronic Measurement InstrumentsE 428 Practice for Fabrication and Control of Metal, Otherthan Aluminum Reference, Blocks Used in Ultrasoni
9、cExaminationE 543 Specification for Agencies Performing Nondestruc-tive TestingE 1316 Terminology for Nondestructive Examinations2.2 ASNT Documents:3SNT-TC-1A Recommended Practice for Personnel Qualifi-cation and Certification in Nondestructive TestingANSI/ASNT-CP-189 Standard for Qualification and
10、Certi-fication of Nondestructive Testing Personnel2.3 Aerospace Industries Association Document:NAS-410 Certification and Qualification of NondestructiveTesting Personnel4NOTE 1For DoD contracts, unless otherwise specified the issues ofthe documents, which are DoD adopted, are those listed in the is
11、sue of theDoDISS (Department of Defense Index of Specifications Standards) citedin the solicitation.3. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this practice, seeTerminology E 1316.3.2 Definitions of Terms Specific to This Standard:3.2.1 effective beam diameterthat distance
12、through whicha search unit can be traversed across a calibration reflector sothat the corresponding echo amplitude is at least one half (-6dB) of the maximum amplitude. The effective beam diameter isnot a characteristic of the search unit alone, but is dependent onpropagating medium, distance to the
13、 discontinuity, reflectorgeometry, etc.3.2.2 scan indexthe length of the step created by indexingthe scan of the search unit over the part, that is continuously1This practice is under the jurisdiction of ASTM Committee E07 on Nonde-structive Testing and is the direct responsibility of Subcommittee E
14、07.06 onUltrasonic Method.Current edition approved Dec. 1, 2006. Published January 2007. Originallyapproved in 1984. Last previous edition approved in 2004 as E 1001 - 04.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Ann
15、ual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from TheAmerican Society for Nondestructive Testing (ASNT), P.O.Box 28518, 1711 Arlingate Ln., Columbus, OH 43228-0518.4Available from Aerospace Industries Association of America,
16、 Inc. (AIA), 1000Wilson Blvd., Suite 1700,Arlington, VA22209-3928, http:/www.aia-aerospace.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.scanning in one direction, then stepping in the directionperpendicular to the scan or maki
17、ng a linear advance perrotation (pitch) for rotary scan of cylindrical parts. Theallowable scan index should be correlated with the search uniteffective beam diameter to ensure full coverage of the part asdescribed in 8.2 below.3.2.3 transfera change in scanning gain to compensate fordifferences in
18、attenuation of the reference standard and the partor material being inspected.4. Summary of Practice4.1 This practice describes a means for obtaining an evalu-ation of discontinuities in materials by immersed examinationwith longitudinal ultrasonic waves. Equipment, reference stan-dards, examination
19、 and evaluation procedures, and documen-tation are described in detail.5. Basis of Application5.1 Personnel QualificationIf specified in the contractualagreement, personnel performing examinations to this standardshall be qualified in accordance with a nationally recognizedNDT personnel qualificatio
20、n practice or standard such asANSI/ASNT-CP-189, SNT-TC-1A, NAS-410, or a similardocument certified by the employer or certifying agency, asapplicable. The practice or standard used and its applicablerevision shall be identified in the contractual agreement be-tween the using parties.5.2 Written Proc
21、edureA detailed written procedure shallbe used for the performance of ultrasonic examinations inaccordance with this practice. The written procedure shouldaddress all applicable portions of this practice, for the purposesof consistency in application and the repeatability of theexamination results.
22、Specific requirements regarding prepara-tion and approval of the written procedure should be deter-mined by purchaser and supplier agreement.6. Significance and Use6.1 This practice provides guidelines for the application ofimmersed longitudinal wave examination to the detection andquantitative eval
23、uation of discontinuities in materials.6.2 Although not all requirements of this practice can beapplied universally to all inspection situations and materials, itdoes provide a basis for establishing contractual criteriabetween suppliers and purchasers of materials for performingimmersed pulse-echo
24、examination, and may be used as ageneral guide for writing detailed specifications for particularapplications.6.3 This practice is directed towards the evaluation ofdiscontinuities detectable at normal beam incidence. If discon-tinuities at other orientations are of concern, alternate scanningtechni
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME10012006STANDARDPRACTICEFORDETECTIONANDEVALUATIONOFDISCONTINUITIESBYTHEIMMERSEDPULSEECHOULTRASONICMETHODUSINGLONGITUDINALWAVES

链接地址:http://www.mydoc123.com/p-527703.html