1、Designation: E 70 07Standard Test Method forpH of Aqueous Solutions With the Glass Electrode1This standard is issued under the fixed designation E 70; the number immediately following the designation indicates the year of originaladoption or, in the case of revision, the year of last revision.Anumbe
2、r in parentheses indicates the year of last reapproval.Asuperscriptepsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope*1.1 This test method specifies the apparatus and proceduresfor
3、the electrometric measurement of pH values of aqueoussolutions with the glass electrode. It does not deal with themanner in which the solutions are prepared. pH measurementsof good precision can be made in aqueous solutions containinghigh concentrations of electrolytes or water-soluble organiccompou
4、nds, or both. It should be understood, however, that pHmeasurements in such solutions are only a semiquantitativeindication of hydrogen ion concentration or activity. Themeasured pH will yield an accurate result for these quantitiesonly when the composition of the medium matches approxi-mately that
5、of the standard reference solutions. In general, thistest method will not give an accurate measure of hydrogen ionactivity unless the pH lies between 2 and 12 and the concen-tration of neither electrolytes nor nonelectrolytes exceeds 0.1mol/L (M).1.2 The values stated in SI units are to be regarded
6、asstandard. The values in parentheses are for information only.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applic
7、a-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1193 Specification for Reagent WaterE 180 Practice for Determining the Precision of ASTMMethods for Analysis and Testing of Industrial and Spe-cialty ChemicalsE 691 Practice for Conducting an Interlaboratory
8、 Study toDetermine the Precision of a Test Method3. Terminology3.1 Definition:3.1.1 pHdefined formally as the negative logarithm to thebase 10 of the conventional hydrogen ion activity. See Appen-dix X1.3.2 Definitions of Terms Specific to This Standard:3.2.1 For the purpose of this test method, the
9、 term “meter”shall apply to the instrument used for the measurement ofpotential (either in millivolts or in terms of pH units), the term“electrodes” to the glass electrode and the reference electrode,and the term “assembly” to the combination of the meter andthe electrodes. The performance of the me
10、ter shall be differ-entiated from that of the electrodes.4. Significance and Use4.1 pH is, within the limits described in 1.1, an accuratemeasurement of the hydrogen ion concentration and thus iswidely used for the characterization of aqueous solutions.4.2 pH measurement is one of the main process c
11、ontrolvariables in the chemical industry and has a prominent place inpollution control.5. Apparatus5.1 pH metersMany excellent pH meters are availablefrom commercial sources. To some extent, the choice of meterwill depend on the desired precision of measurement. Themeter may operate on a null-detect
12、ion principle or may utilizedigital readout or a direct deflection meter with a large scale.Power may be supplied by batteries or a-c operation may beprovided. The maximum grid current drawn from the glasselectrode during measurement shall not exceed 2 3 1012A.Automatic or manual adjustment shall al
13、low for changes in1This test method is under the jurisdiction of ASTM Committee E15 onIndustrial and Specialty Chemicals and is the direct responsibility of SubcommitteeE15.01 on General Standards.Current edition approved April 1, 2007. Published May 2007. Originallyapproved in 1952. Last previous e
14、dition approved in 2002 as E 70 97 (2002).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1*A Summary of Chan
15、ges section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.F/(RT ln 10) when the temperature of the assembly is altered.For referee work, or in case of dispute, meters capable ofdiscriminating chan
16、ges of pH to 0.01 unit (0.6 mV) or less shallbe used.5.2 Reference Electrodes and Glass Electrodes:5.2.1 The saturated calomel electrode and the 3.5 mol/L(M)calomel electrode are suitable as reference electrodes in pHassemblies (Note 1). If the saturated electrode is used, a fewcrystals of solid pot
17、assium chloride shall be present in thechamber surrounding the electrode element at each tempera-ture. The design of the electrode shall permit a fresh liquidjunction between the solution of potassium chloride and thebuffer or test solution to be formed for each test and shall allowtraces of solutio
18、n to be readily removed by washing.NOTE 1Other reference electrodes of constant potential may be used,provided no difficulty is experienced in standardizing the assembly asdescribed in Section 8.5.2.2 The silver-silver chloride electrode also is used widelyas a reference electrode.5.2.3 Commercial g
19、lass electrodes are designed for certainspecific ranges of pH and temperature; consequently, the pHand temperature of the test solutions shall be considered inselecting the glass electrode for use. The pH response shallconform with the requirements set forth in Section 7. The leadsshall be shielded
20、from the effects of body capacitance.5.2.4 If the assembly is in intermittent use, the ends of theelectrodes shall be immersed in distilled water between mea-surements. The high-alkalinity type of glass electrode shall bestored in the borax buffer solution. For prolonged storage, glasselectrodes may
21、 be allowed to become dry, and referenceelectrodes shall be capped to prevent undue evaporation.NOTE 2New glass electrodes and those that have been stored dry shallbe conditioned as recommended by the manufacturer. Requirements forthe physical dimensions and shape of the electrodes and the compositi
22、onof the internal reference solution are not considered part of this testmethod.6. Reagents and Materials6.1 The pH(S) of six recommended standard solutions atseveral temperatures is listed in Table 1. The buffer solutionsshall be prepared from highly purified materials sold specifi-cally as pH stan
23、dards (Note 3). Potassium hydrogen phthalateand the two phosphate salts shall be dried at 110C for 1 hbefore use, but borax and sodium bicarbonate shall not beheated above room temperature. Potassium dihydrogen citrateshall be dried for1hat80C, and sodium carbonate shall beignited for1hat270C before
24、 use. The standard solutionsshall be prepared as described in 6.4-6.9. They shall bepreserved in bottles of chemically resistant glass or polyethyl-ene and shall be replaced at an age of 6 weeks, or earlier if avisible change should occur in the solution.NOTE 3Six of the buffer salts can be obtained
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME702007STANDARDTESTMETHODFORPHOFAQUEOUSSOLUTIONSWITHTHEGLASSELECTRODE 玻璃 电极 水溶液 PH 测量 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-527609.html