1、Designation: E181 10Standard Test Methods forDetector Calibration and Analysis of Radionuclides1This standard is issued under the fixed designation E181; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A
2、number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These methods cover general procedures for the cali-bration of radiation detectors and the analysis of radionuclides.For each individua
3、l radionuclide, one or more of these methodsmay apply.1.2 These methods are concerned only with specific radio-nuclide measurements. The chemical and physical propertiesof the radionuclides are not within the scope of this standard.1.3 The measurement standards appear in the followingorder:SectionsS
4、pectroscopy Methods:Calibration and Usage of Germa-nium Detectors 3-12Calibration and Usage of ScintillationDetector Systems: 13-20Calibration and Usage of ScintillationDetectors for Simple Spectra 16Calibration and Usage of ScintillationSectionsDetectors for Complex Spectra 17Counting Methods:Beta
5、Particle Counting 25-26Aluminum Absorption Curve 27-31Alpha Particle Counting 32-39Liquid Scintillation Counting 40-481.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesa
6、fety problems, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Document2.1 ASTM Standards:2E170 Terminology Relating to Ra
7、diation Measurements andDosimetrySPECTROSCOPY METHODS3. Terminology3.1 Definitions:3.1.1 certified radioactivity standard sourcea calibratedradioactive source, with stated accuracy, whose calibration iscertified by the source supplier as traceable to the NationalRadioactivity Measurements System (1)
8、.33.1.2 check sourcea radioactivity source, not necessarilycalibrated, that is used to confirm the continuing satisfactoryoperation of an instrument.3.1.3 FWHM(full width at half maximum) the full widthof a gamma-ray peak distribution measured at half the maxi-mum ordinate above the continuum.3.1.4
9、national radioactivity standard sourcea calibratedradioactive source prepared and distributed as a standardreference material by the U.S. National Institute of Standardsand Technology.3.1.5 resolution, gamma raythe measured FWHM, afterbackground subtraction, of a gamma-ray peak distribution,expresse
10、d in units of energy.3.2 Abbreviations:Abbreviations:3.2.1 MCAMultichannel Analyzer.3.2.2 SCASingle Channel Analyzer.3.2.3 ROIRegion-Of-Interest.3.3 For other relevant terms, see Terminology E170.3.4 correlated photon summingthe simultaneous detec-tion of two or more photons originating from a singl
11、e nucleardisintegration.1These methods are under the jurisdiction ofASTM Committee E10 on NuclearTechnology and Applications .Current edition approved Jan. 1, 2010. Published February 2010. Originallyapproved in 1961. Last previous edition approved in 2003 as E181 98(2003). DOI:10.1520/E0181-10.2For
12、 referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The boldface numbers in parentheses refer to the list of referen
13、ces at the end ofthese methods.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.5 dead timethe time after a triggering pulse duringwhich the system is unable to retrigger.NOTE 1The terms “standard source” and “radioactivity standard
14、” aregeneral terms used to refer to the sources and standards of NationalRadioactivity Standard Source and Certified Radioactivity StandardSource.CALIBRATION AND USAGE OF GERMANIUMDETECTORS4. Scope4.1 This standard establishes methods for calibration, usage,and performance testing of germanium detec
15、tors for themeasurement of gamma-ray emission rates of radionuclides. Itcovers the energy and full-energy peak efficiency calibration aswell as the determination of gamma-ray energies in the 0.06 to2-MeV energy region and is designed to yield gamma-rayemission rates with an uncertainty of 63 % (see
16、Note 2). Thismethod applies primarily to measurements that do not involveoverlapping peaks, and in which peak-to-continuum consider-ations are not important.NOTE 2Uncertainty U is given at the 68 % confidence level; that is,U 5 =(si21 1/3(di2where diare the estimated maximum system-atic uncertaintie
17、s, and siare the random uncertainties at the 68 %confidence level (2). Other methods of error analysis are in use (3, 4).5. Apparatus5.1 A typical gamma-ray spectrometry system consists of agermanium detector (with its liquid nitrogen cryostat, pream-plifier, and possibly a high-voltage filter) in c
18、onjunction with adetector bias supply, linear amplifier, multichannel analyzer,and data readout device, for example, a printer, plotter,oscilloscope, or computer. Gamma rays interact with thedetector to produce pulses which are analyzed and counted bythe supportive electronics system.6. Summary of M
19、ethods6.1 The purpose of these methods is to provide a standard-ized basis for the calibration and usage of germanium detectorsfor measurement of gamma-ray emission rates of radionu-clides. The method is intended for use by knowledgeablepersons who are responsible for the development of correctproce
20、dures for the calibration and usage of germanium detec-tors.6.2 A source emission rate for a gamma ray of a selectedenergy is determined from the counting rate in a full-energypeak of a spectrum, together with the measured efficiency ofthe spectrometry system for that energy and source location. Iti
21、s usually not possible to measure the efficiency directly withemission-rate standards at all desired energies. Therefore acurve or function is constructed to permit interpolation be-tween available calibration points.7. Preparation of Apparatus7.1 Follow the manufacturers instructions for setting up
22、and preliminary testing of the equipment. Observe all of themanufacturers limitations and cautions. All tests described inSection 12 should be performed before starting the calibra-tions, and all corrections shall be made when required.Achecksource should be used to check the stability of the system
23、 atleast before and after the calibration.8. Calibration Procedure8.1 Energy CalibrationDetermine the energy calibration(channel number versus gamma-ray energy) of the detectorsystem at a fixed gain by determining the channel numberscorresponding to full energy peak centroids from gamma raysemitted
24、over the full energy range of interest from multipeakedor multinuclide radioactivity sources, or both. Determinenonlinearity correction factors as necessary (5).8.1.1 Using suitable gamma-ray compilations (6-14), plot orfit to an appropriate mathematical function the values for peakcentroid (in chan
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME18120106875STANDARDTESTMETHODSFORDETECTORCALIBRATIONANDANALYSISOFRADIONUCLIDES 放射性 核素 探测器 校准 分析

链接地址:http://www.mydoc123.com/p-527083.html