ASTM D7979-2015e1 7947 Standard Test Method for Determination of Perfluorinated Compounds in Water Sludge Influent Effluent and Wastewater by Liquid Chromatography Tandem Mass Spec.pdf
《ASTM D7979-2015e1 7947 Standard Test Method for Determination of Perfluorinated Compounds in Water Sludge Influent Effluent and Wastewater by Liquid Chromatography Tandem Mass Spec.pdf》由会员分享,可在线阅读,更多相关《ASTM D7979-2015e1 7947 Standard Test Method for Determination of Perfluorinated Compounds in Water Sludge Influent Effluent and Wastewater by Liquid Chromatography Tandem Mass Spec.pdf(18页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7979 151Standard Test Method forDetermination of Perfluorinated Compounds in Water,Sludge, Influent, Effluent and Wastewater by LiquidChromatography Tandem Mass Spectrometry (LC/MS/MS)1This standard is issued under the fixed designation D7979; the number immediately following the desig
2、nation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEEditorial corrections were made throug
3、hout in November 2015.1. Scope1.1 This procedure covers the determination of selectedperfluorinated compounds (PFCs) in a water matrix usingliquid chromatography (LC) and detection with tandem massspectrometry (MS/MS). These analytes are qualitatively andquantitatively determined by this method. Thi
4、s method adheresto multiple reaction monitoring (MRM) mass spectrometry.1.2 The Method Detection Limit (MDL)2and ReportingRange3for the target analytes are listed in Table 1.1.2.1 The reporting limit in this test method is the minimumvalue below which data are documented as non-detects. Ana-lyte det
5、ections between the method detection limit and thereporting limit are estimated concentrations and are not re-ported following this test method. In most cases, the reportinglimit is the concentration of the Level 1 calibration standard asshown in Table 4 for the perfluorinated compounds after taking
6、into account the 50 % dilution with methanol. It is above theLevel 1 calibration concentration for PFOS, PFBS, FHEA andFOEA, these compounds can be identified at the Level 1concentration but the standard deviation among replicates atthis lower spike level resulted in a higher reporting limit.1.3 Uni
7、tsThe values stated in SI units are to be regardedas standard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-
8、priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:4D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Metho
9、ds of Committee D19 on WaterD3856 Guide for Management Systems in LaboratoriesEngaged in Analysis of WaterD3694 Practices for Preparation of Sample Containers andfor Preservation of Organic ConstituentsD4841 Practice for Estimation of Holding Time for WaterSamples Containing Organic and Inorganic Co
10、nstituentsD5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water AnalysisE2554 Practice for Estimating and Monitoring the Uncer-tainty of Test Results of a Test Method Using ControlChart Techniques2.2 Other Standards:5EPAPublication SW-846, Test Methods for Eval
11、uating SolidWaste, Physical/Chemical MethodsThe Code of Federal Regulations 40 CFR Part 136, Appen-dix B3. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this standard, refer toTerminology D1129.1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the d
12、irect responsibility of Subcommittee D19.06 on Methods forAnalysis forOrganic Substances in Water.Current edition approved Jan. 1, 2015. Published February 2015. DOI: 10.1520/D7979-15E01.2The MDL is determined following the Code of Federal Regulations, 40 CFRPart 136, Appendix B utilizing dilution a
13、nd filtration. 5 mL sample of water wasutilized. A detailed process determining the MDL is explained in the reference andis beyond the scope of this test method to be explained here.3Reporting range concentration is calculated from Table 4 concentrationsassuming a 30 L injection of the Level 1 calib
14、ration standard for PFCs, and thehighest level calibration standard with a 10 mLfinal extract volume ofa5mLwatersample. Volume variations will change the reporting limit and ranges.4For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.or
15、g. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.5Available from National Technical Information Service (NTIS), U.S. Depart-ment of Commerce, 5285 Port Royal Road, Springfield, VA, 22161 or at http:/www.epa.gov/epawaste/hazard/te
16、stmethods/index.htmCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2 Definitions of Terms Specific to This Standard:3.2.1 perfluorinated compounds, nin this test method, 11perfluoroalkyl carboxylic acids, 3 perfluoroalkylsulfonates,
17、Decafluoro-4-(pentafluoroethyl)cyclohexanesulfonate and 6fluorotelomer acids listed in Table 1 collectively (not includingmass labeled surrogates).3.2.2 reporting limit, nthe minimum concentration belowwhich data are documented as non-detects.3.3 Acronyms:3.3.1 CCC, nContinuing Calibration Check3.3.
18、2 FTAs and FTUAs, nFluorotelomer and UnsaturatedFluorotelomer Acids3.3.2.1 FDEA, n2-perfluorodecyl ethanoic acid3.3.2.2 FHEA, n2-perfluorohexyl ethanoic acid3.3.2.3 FHpPA, n3-perfluoroheptyl propanoic acid3.3.2.4 FHUEA, n2H-perfluoro-2-octenoic acid3.3.2.5 FOEA, n2-perfluorooctyl ethanoic acid3.3.2.
19、6 FOUEA, n2H-perfluoro-2-decenoic acid3.3.3 IC, nInitial Calibration3.3.4 LC, nLiquid Chromatography3.3.5 LCS/LCSD, nLaboratory Control Sample/Laboratory Control Sample Duplicate3.3.6 MDL, nMethod Detection Limit3.3.7 MeOH, nMethanol3.3.8 mM, nmillimolar,110-3moles/L3.3.9 MRM, nMultiple Reaction Mon
20、itoring3.3.10 MPFAS, nIsotopically labeled Perfluoroalkylsul-fonates3.3.10.1 MPFHxS, n18O2-Perfluorohexylsulfonate3.3.10.2 MPFOS, n13C4-Perfluorooctylsulfonate3.3.11 MPFCA, nIsotopically labeled Perfluoroalkylcar-boxylates3.3.11.1 MPFBA, n13C4-Perfluorobutanoate3.3.11.2 MPFDA, n13C2-Perfluorodecanoa
21、te3.3.11.3 MPFDoA, n13C2-Perfluorododecanoate3.3.11.4 MPFHxA, n13C2-Perfluorohexanoate3.3.11.5 MPFNA, n13C5-Perfluorononanoate3.3.11.6 MPFOA, n13C4-Perfluorooctanoate3.3.11.7 MPFUnA, n13C2-Perfluoroundecanoate3.3.12 MS/MSD, nMatrix Spike/Matrix Spike Duplicate3.3.13 NA, adjNot Available3.3.14 ND, nn
22、on-detect3.3.15 P however, this test method isintended to be performance based and alternative operatingconditions can be used to perform this method provided dataquality objectives are attained.TABLE 1 Method Detection Limit and Reporting RangeAnalyteAMDL(ng/L)Reporting Ranges(ng/L)PFTreA 1.74 10 4
23、00PFTriA 2.65 10 400PFDoA 2.42 10 400PFUnA 1.08 10 400PFDA 3.03 10 400PFOS 4.19 15 400PFNA 1.76 10 400PFecHS 1.93 10 400PFOA 3.04 10 400PFHxS 2.51 10 400PFHpA 2.32 10 400PFHxA 1.31 10 400PFBS 7.60 30 400PFPeA 11.59 50 2000PFBA 13.85 50 2000FHEA 92.93 300 8000FOEA 106.75 300 8000FDEA 47.17 200 8000FO
24、UEA 2.31 10 400FHpPA 3.25 10 400FHUEA 1.53 10 400AAcronyms are defined in 3.3.D7979 15124.2 For PFC analysis, samples are shipped to the lab at atemperature between 0C and 6C and analyzed within 28 daysof collection. A sample (5 mL) is transferred to a polypropyl-ene tube (ora5mLsample is collected
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD79792015E17947STANDARDTESTMETHODFORDETERMINATIONOFPERFLUORINATEDCOMPOUNDSINWATERSLUDGEINFLUENTEFFLUENTANDWASTEWATERBYLIQUIDCHROMATOGRAPHYTANDEMMASSSPECPDF

链接地址:http://www.mydoc123.com/p-526595.html