《ASTM D7962-2015a red 2262 Standard Practice for Determination of Minimum Immersion Depth and Assessment of Temperature Sensor Measurement Drift《测定温度传感器最小浸入深度以及评估测得漂移的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7962-2015a red 2262 Standard Practice for Determination of Minimum Immersion Depth and Assessment of Temperature Sensor Measurement Drift《测定温度传感器最小浸入深度以及评估测得漂移的标准实施规程》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7962 15D7962 15aStandard Practice forDetermination of Minimum Immersion Depth andAssessment of Temperature Sensor Measurement Drift1This standard is issued under the fixed designation D7962; the number immediately following the designation indicates the year oforiginal adoption or, in
2、the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope Scope*1.1 This practice describes a two procedures for use with temperature measurement
3、 devices. Methodology is described fordetermining minimum immersion depth for thermal sensors, in particular RTDs or similar temperature sensors. Included is aprocedure for consistently preparing a reference bath for the purpose of monitoring measurement drift of thermal sensors such asliquid-in-gla
4、ss or digital contact thermometers.1.2 This practice focuses on temperature measurement drift in a laboratory. If the user requires greater measurement accuracy,then they should follow the instructions in Practice E563.1.3 The values stated in SI units are to be regarded as standard. No other units
5、of measurement are included in this standard.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulat
6、orylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E563 Practice for Preparation and Use of an Ice-Point Bath as a Reference Temperature3. Terminology3.1 Definitions:3.1.1 digital contact thermometer (DCT), nan electronic device consisting of a digital display and associated temp
7、eraturesensing probe.3.1.1.1 DiscussionThis device consists of a temperature sensor connected to a measuring instrument; this instrument measures the temperature-dependent quantity of the sensor, computes the temperature from the measured quantity, and provides a digital output. This digitaloutput g
8、oes to a digital display and/or recording device that may be internal or external to the device. These devices are sometimesreferred to as a “digital thermometer”.3.1.1.2 DiscussionPortable electronic thermometers (PET) is an acronym sometimes used to refer to a subset of the devices covered by this
9、 definition.3.1.2 ice-point bath, nphysical system containing ice and water assembled to realize the ice point as a reference temperature,or to establish a constant temperature near 0 C.3.1.3 minimum immersion length, ndepth that a thermometer should be immersed, in a uniform temperature environment
10、,such that further immersion does not produce a change in indicated temperature greater than the specified tolerance.1 This practice is under the jurisdiction of ASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of SubcommitteeD02.91 on Coordinat
11、ing Subcommittee on Thermometry.Current edition approved April 1, 2015Dec. 1, 2015. Published April 2015February 2016. Originally approved in 2015. Last previous edition approved in 2015 asD7962 15. DOI: 10.1520/D7962-15.10.1520/D7962-15A.2 For referencedASTM standards, visit theASTM website, www.as
12、tm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what c
13、hanges have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the of
14、ficial document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14. Summary of Practice4.1 This practice describes a procedure for consistently preparing an ice bath that
15、 is an intimate mixture of crushed ice or iceparticles and water in a thermally insulating vessel open to the atmosphere.4.1.1 Cautionwhen the ice bath is not made from distilled water, its temperature will differ from the natural fixed-pointtemperature by a consistent amount, typically less than 0.
16、02 C. If the user needs a more accurate ice point, then they should usePractice E563 to prepare the ice bath.4.2 This practice includes a procedure for determining the minimum immersion depth of the temperature sensor using an icebath.4.2.1 This procedure determines minimum immersion depth with a 25
17、 C differential between sensed temperature and ambienttemperature. If the probe is subjected to a greater differential temperature, a larger immersion depth may be required to correctlymeasure the temperature.5. Significance and Use5.1 This practice provides a means for the users of ASTM Committee D
18、02 standards to monitor the drift in sensed temperatureof liquid-in-glass thermometer (LiG), and digital contact thermometers (DCT). Digital contact thermometers are sometimesreferred to as portable electronic thermometers (PET) or simply digital thermometers.5.2 This practice is not suitable for de
19、termining the accuracy or calibration of a temperature measuring device as the error inthe ice bath temperature can be greater than 0.02 C. For greater accuracy, the user should use Practice E563 to prepare the icebath.5.3 The ice point is a common practical industrial reference point of thermometry
20、. The ice point is relatively simple to realizeand provides a readily available natural fixed-point reference temperature.5.4 This practice only checks the measurement drift at a single temperature. It will not detect a change in measurement responsewith change in temperature. Temperature measuring
21、devices should be recalibrated at set intervals. See device supplier forrecommendations.5.5 This practice provides a technique to determine minimum immersion depth of the sensing probe of the thermometer usingan ice bath. The minimum immersion depth determined by this practice may change when the di
22、fferential temperature differssignificantly from the conditions described. A greater differential will likely increase the minimum immersion depth.6. Reagents and Equipment6.1 WaterGood quality drinking water from a consistent source.6.2 IceCrushed ice prepared from the water in 6.1.6.3 Insulated co
23、ntaineran insulated container large enough to hold the thermometer and provide several centimeters ofwater-ice slush around the measuring portion of the thermometer.7. ProcedureDrift in Temperature Measurement7.1 Preparation of the Ice-Point Bath:7.1.1 Determine the water source to be used each time
24、 this practice is followed in your laboratory to determine measurementdrift and note the water source with the measurement. Altering the water source between uses of the practice when monitoring atemperature sensor may result in observing a false drift.7.1.2 Ensure that all equipment is clean and fr
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD79622015ARED2262STANDARDPRACTICEFORDETERMINATIONOFMINIMUMIMMERSIONDEPTHANDASSESSMENTOFTEMPERATURESENSORMEASUREMENTDRIFT

链接地址:http://www.mydoc123.com/p-526573.html