ASTM D7772-2015 red 4435 Standard Test Method for Carbon Black Extractables &x2013 Absorbance of Cyclohexane Extract《环己烷萃取的可萃取炭黑吸收率的标准试验方法》.pdf
《ASTM D7772-2015 red 4435 Standard Test Method for Carbon Black Extractables &x2013 Absorbance of Cyclohexane Extract《环己烷萃取的可萃取炭黑吸收率的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7772-2015 red 4435 Standard Test Method for Carbon Black Extractables &x2013 Absorbance of Cyclohexane Extract《环己烷萃取的可萃取炭黑吸收率的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7772 11D7772 15Standard Test Method forCarbon Black Extractables Absorbance of CyclohexaneExtract1This standard is issued under the fixed designation D7772; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of
2、last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This method covers the determination of extinction (absorbance) of carbon black cyclohexane extract at 386 nm using a
3、UVspectrophotometer.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this sta
4、ndard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D4483 Practice for Evaluating Precision for Test Method Standards in the Rubber and Carbon Black Manufacturing Industries3. Summ
5、ary of Test Method3.1 A carbon black sample is gently agitated in cyclohexane at room temperature to remove any extractable material. Themixture is left standing in the dark for 24 h at room temperature. The carbon black is separated from the cyclohexane by filtration,with subsequent determination o
6、f the extinction of the filtrate at 386 nm using a 50-mm cuvette.4. Significance and Use4.1 This procedure serves as a screening process for carbon blacks used in the manufacturing of products that are to come intocontact with food. The cyclohexane extract absorbance at 386 nm must be less than 0.10
7、 for a 50 mm cuvette in order to fulfillvarious regulatory requirements.5. Apparatus5.1 Spectrophotometer, 20 nm maximum spectral bandpass, capable of measuring absorbance at 386 nm.5.2 Cuvettes, with an optical light path of 50 mm.5.3 Balance, analytical, sensitivity 0.01 g.5.4 Filter Paper, with a
8、 particle retention in liquid less than 2 m.5.5 Glass Filtering Funnels, 75 mm diameter at top.5.6 Volumetric Flask, 100 cm3.5.7 Lens Cleaning Tissue, lint free.5.8 Timer, with 1 s divisions.5.9 Oven, gravity-convection type, capable of temperature regulation of 125 6 5C.5.10 Desiccator.1 This test
9、method is under the jurisdiction of ASTM Committee D24 on Carbon Black and is the direct responsibility of Subcommittee D24.66 on Environment, Health,and Safety.Current edition approved Nov. 1, 2011June 1, 2015. Published January 2012September 2015. Originally approved in 2011. Last previous edition
10、 approved in 2011 asD7772 11. DOI: 10.1520/D7772-11.10.1520/D7772-15.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM
11、website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult p
12、rior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.11 As a good laboratory practice, it c
13、ould be suggested the use of a Neutral Density Filter with a spectral absorbance calibratedat 386 nm to ensure the spectrophotometer is correctly calibrated.6. Reagents and Materials6.1 Cyclohexane, analytical reagent grade.7. Hazards7.1 This test involves hazardous materials, operations and equipme
14、nt. This procedure does not attempt to address the safetyproblems associated with this test.Ahazards review must be conducted by all personnel performing the test. It is the responsibilityof the user to review all MSDS, manuals and hazards procedures and establish the appropriate safety measures.8.
15、Calibration, Standardization, and Environmental Conditions8.1 Standard Laboratory Test Conditions:Conditions:8.1.1 Unless stated elsewhere, all testing, calibration and standardization should be carried out in the ambient temperature rangeof 23 6 3C.8.1.2 The date of calibration or testing should al
16、so be recorded with any relevant data.8.1.3 All calibration data should be recorded on a dedicated file.8.2 Instrument Set-Up Frequency Day of Test:8.2.1 Turn on the spectrophotometer and allow it to warm up for 30 min.8.2.2 Set the wavelength to 386 nm and zero the instrument.8.2.3 Place the Neutra
17、l Density Filter inside the sample compartment of the instrument and verify the absorbance value is withinthe absorbance tolerance. If the filter is found to be outside the limits, verify the filters absorbance using a more sophisticatedUV/Vis Spectrometer. If in this case nothing abnormal is found
18、for the filter, service and calibrate the spectrophotometer.8.2.4 It is imperative that steps 7.2.58.2.5 7.2.88.2.8 are carried out in a fume hood using disposable nitrile gloves. This kindof gloves are recommended for cyclohexane to prevent skin exposure (cyclohexane can cause skin irritation).8.2.
19、5 Filter approximately 50 cm3 cyclohexane through the filter paper into a stoppered graduated cylinder.8.2.6 Transfer a portion of the filtered cyclohexane to a clean cuvette. The cuvette should not be handled on the clear workingsurface by fingers.8.2.7 Rinse the cuvette with filtered cyclohexane t
20、hree times, filling approximately 13 full each time. Discard the rinsingcyclohexane into an approved safety can.8.2.8 Fill the cuvette and wipe the outside surfaces with lint free wiping paper while holding the cuvette in front of a suitablelight source to check for smudges. The cyclohexane must be
21、free of any contaminants such as lint particles, which might causelight scattering and influence test results.8.2.9 Insert the cuvette into the spectrophotometer, adjust it to read the 386 nm wavelength and zero the instrument to achieve0.000 absorbance at 386 nm.9. Procedure9.1 Dry the sample(s) of
22、 carbon black for 1 h at 125 6 5C prior to testing.9.2 Allow the sample to cool to room temperature in a desiccator.9.3 Weigh 1.00 6 0.01 g sample and transfer to a 100 cm3 volumetric flask.9.4 It is imperative that steps 8.59.5 8.99.9 are carried out in a fume hood using disposable nitrile gloves.
23、This kind of glovesare recommended for cyclohexane to prevent skin exposure (cyclohexane can cause skin irritation).9.5 Add 25 to 30 cm3 of cyclohexane to the sample flask and stopper the flask.9.6 Shake the mixture gently for a few seconds.9.7 Fill up to the mark with cyclohexane.TABLE 1 Precision
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD77722015RED4435STANDARDTESTMETHODFORCARBONBLACKEXTRACTABLESX2013ABSORBANCEOFCYCLOHEXANEEXTRACT 环己烷

链接地址:http://www.mydoc123.com/p-526293.html