1、Designation: D7535 091Standard Test Method forLead-210 in Water1This standard is issued under the fixed designation D7535; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates
2、 the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEAdded research report footnote to Section 15 editorially in June 2013.1. Scope1.1 This test method covers the determination of radioac-tive210Pb in environmental water sampl
3、es (for example,drinking, non-process and effluent waters) in the range of 37mBq/L (1.0 pCi/L) or greater.1.2 The values stated in SI units are to be regarded asstandard. The values given in parentheses are provided forinformation purposes only.1.3 This method has been used successfully with tap wat
4、er.It is the users responsibility to ensure the validity of this testmethod for samples larger than 500 mL and for waters ofuntested matrices.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard
5、 to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specifichazards statements, see Section 9.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD2777 Pract
6、ice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD3370 Practices for Sampling Water from Closed ConduitsD4448 Guide for Sampling Ground-Water Monitoring WellsD5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water Ana
7、lysisD6001 Guide for Direct-Push Groundwater Sampling forEnvironmental Site CharacterizationD7282 Practice for Set-up, Calibration, and Quality Controlof Instruments Used for Radioactivity MeasurementsE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method3.
8、Terminology3.1 DefinitionsFor definitions of terms used in this testmethod, refer to Terminology D1129.4. Summary of Test Method4.1 This test method is based on the utilization of solidphase extraction of lead from water samples with detection ofthe radioactive lead by beta gas flow proportional cou
9、nting.Analiquot of the sample is measured into a beaker; iron carrier andlead carrier are added. Lead is scavenged by an iron hydroxideprecipitation. Lead is then selectively sorbed on a solid phaseextraction column and eluted with water. Lead is precipitatedas lead sulfate and is collected on a fil
10、ter paper.The lead sulfateprecipitate is covered with aluminum foil and held for 5 daysor longer for210Bi ingrowth. It is then counted for betaradiation on a gas flow proportional counter.5. Significance and Use5.1 This test method was developed to measure the concen-tration of210Pb in nonprocess wa
11、ter samples. This test methodmay be used to determine the concentration of210Pb inenvironmental samples.6. Interferences6.1 Significant amounts of stable lead (0.3 mg/L) presentin the sample will interfere with the chemical yielddetermination, leading to positive bias in the yield. If it isknown or
12、suspected that natural lead is present in the sample,blank sample aliquots to which no lead carrier content is addedshould be analyzed. The amount of natural lead contained inthe sample shall be used to correct the yield.6.2 In most cases measurable amounts of210Pb will bepresent in the Pb carrier u
13、sed in Section 12.1. This additionalcontribution to the measured sample activity must be deter-mined and treated as additional background activity. For eachnew carrier solution, the inherent210Pb in the carrier must bemeasured. A brief description of this additional backgroundcalibration is given in
14、 Section 11.2. Accurate determination of1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.04 on Methods of Radiochemi-cal Analysis.Current edition approved Oct. 1, 2009. Published October 2009. DOI: 10.1520/D7535-09E01.2For
15、 referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700,
16、 West Conshohocken, PA 19428-2959. United States1the combined standard uncertainty and the minimum detectableconcentration should include this additional activity in themethod background count rate.6.3 Previous experimental data suggests that barium con-centrations of approximately 30 ppm can be a s
17、ource ofinterference, leading to chemical yield determinations that maybe non-representative and outside normal acceptance criteria.7. Apparatus7.1 Analytical Balance, 0.0001 g.NOTE 1A thickness of aluminum of approximately 0.003 in. (arealdensity of approximately 0.02 mg/cm2) is needed to effective
18、ly eliminatethe 60 keV beta particles from210Pb.3The thickness of aluminum foilused during calibration of the detector and measurement samples must beheld constant to maintain self-absorption of210Pb and210Bi beta particlesat a constant level.7.2 Aluminum foil.7.3 Centrifuge.7.4 Centrifuge tubes (50
19、 mL plastic).7.5 Filters, 25 mm polypropylene , 0.1 m, with polycar-bonate base and metal screen.7.6 Filter apparatus, polysulfone funnel and 100 mL poly-propylene flask.7.7 Beta Gas Flow Proportional Counting System, (1.0cpm beta), low background.7.8 Glass stir rods.7.9 Glass beakers.7.10 Hot plate
20、.7.11 Petri dish.7.12 Planchets, stainless steel, flat, with diameter largeenough to hold the 25 mm filter.7.13 Tweezers.7.14 Watch glass.8. Reagents and Materials8.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shal
21、l conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are available. Other grades may beused, provided that the reagent is of sufficiently high purity topermit its use without increasing the background of themeasurement. S
22、ome reagents, even those of high purity, maycontain naturally occurring radioactivity, such as isotopes ofuranium, radium, actinium, thorium, rare earths and potassiumcompounds and/or artificially produced radionuclides.Consequently, when such reagents are used in the analysis oflow-radioactivity sa
23、mples, the activity of the reagents shall bedetermined under analytical conditions that are identical tothose used for the sample. The activity contributed by thereagents may be considered to be a component of backgroundand applied as a correction when calculating the test sampleresult. This increas
24、ed background reduces the sensitivity of themeasurement.8.2 Purity of waterUnless otherwise indicated, referenceto water shall be understood to mean reagent water conformingto Specification D1193, Type III.8.3 Ammonium hydroxide15MNH4OH, (concentratedreagent).8.4 Iron carrier (20 mg/mL)Dissolve 9.6
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD75352009E11875STANDARDTESTMETHODFORLEAD210INWATER 水中 20 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-525743.html