《ASTM D7511-2009e1 7500 Standard Test Method for Total Cyanide by Segmented Flow Injection Analysis In-Line Ultraviolet Digestion and Amperometric Detection.pdf》由会员分享,可在线阅读,更多相关《ASTM D7511-2009e1 7500 Standard Test Method for Total Cyanide by Segmented Flow Injection Analysis In-Line Ultraviolet Digestion and Amperometric Detection.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7511 091Standard Test Method forTotal Cyanide by Segmented Flow Injection Analysis, In-LineUltraviolet Digestion and Amperometric Detection1This standard is issued under the fixed designation D7511; the number immediately following the designation indicates the year oforiginal adoption
2、 or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEA typographical error was editorially corrected in Section 8.13 in March 2010.1. Sc
3、ope1.1 This method is used for determining total cyanide indrinking and surface waters, as well as domestic and industrialwastes. Cyanide ion (CN-), hydrogen cyanide in water (HC-N(aq), and the cyano-complexes of zinc, copper, cadmium,mercury, nickel, silver, and iron may be determined by thismethod
4、. Cyanide ions from Au(I), Co(III), Pd(II), and Ru(II)complexes are only partially determined.1.2 The method detection limit (MDL) is 1.0 g/L cyanideand the minimum level (ML) is 3 g/L. The applicable range ofthe method is 3 to 500 g/L cyanide using a 200-L sampleloop. Extend the range to analyze hi
5、gher concentrations bysample dilution or changing the sample loop volume.1.3 This method can be used by analysts experienced withequipment using segmented flow analysis (SFA) and flowinjection analysis (FIA) or working under the close supervisionof such qualified persons.1.4 The values stated in SI
6、units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health p
7、ractices and determine the applica-bility of regulatory limitations prior to use. Specific hazardstatements are given in Note 2 and Section 9.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD2036 Test Methods for Cyanides in WaterD27
8、77 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD3370 Practices for Sampling Water from Closed ConduitsD3856 Guide for Good Laboratory Practices in Laborato-ries Engaged in Sampling and Analysis of WaterD4210 Practice for Intralaboratory Quality
9、 Control Proce-dures and a Discussion on Reporting Low-Level Data3D5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water AnalysisD6696 Guide for Understanding Cyanide SpeciesD7365 Practice for Sampling, Preservation and MitigatingInterferences in Water Samples f
10、or Analysis of Cyanide3. Terminology3.1 DefinitionsFor definitions of terms used in this testmethod, refer to Terminology D1129 and Guide D6696.3.1.1 total cyaniderefers to all cyanide-containing com-pounds in a sample, including free cyanide, WAD cyanidecompounds, and strong metal cyanide complexes
11、.4. Summary of Test Method4.1 Prior to analysis, treat the sample to remove potentialinterferences (Sections 4 and 8). Ultraviolet (UV) digestionreleases cyanide from cyanide complexes. Acid addition con-verts cyanide ion to hydrogen cyanide gas (HCN), whichpasses under a gas diffusion membrane. The
12、 hydrogen cyanidegas diffuses through the membrane into an alkaline receivingsolution, where it converts back to cyanide ion. A silverworking electrode, silver/silver chloride reference electrode,and platinum/stainless steel counter electrode at an appliedpotential of zero volt amperometrically moni
13、tor the cyanideion. The current generated is proportional to the cyanideconcentration present in the original sample.1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.06 on Methods forAnalysis forOrganic Substances in Water
14、.Current edition approved Feb. 15, 2009. Published March 2009. DOI: 10.1520/D7511-09.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary
15、page onthe ASTM website.3Withdrawn. The last approved version of this historical standard is referencedon www.astm.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.4.2 Calibrations and data are processed with the instru-ments data
16、 acquisition software.5. Significance and Use5.1 Cyanide and hydrogen cyanide are highly toxic. Regu-lations have been established to require the monitoring ofcyanide in industrial and domestic wastewaters and surfacewaters.45.2 This test method is applicable for natural water, salinewaters, and was
17、tewater effluent.5.3 The method may be used for process control in waste-water treatment facilities.5.4 The spot test outlined in Test Methods D2036, AnnexA1 can be used to detect cyanide and thiocyanate in water orwastewater, and to approximate its concentration.6. Interferences6.1 Method interfere
18、nces can be caused by contaminants inthe reagents, reagent water, glassware, etc., which may bias theresults. Take care to keep all such items free of contaminants.6.2 Sulfide and sulfide-containing compounds are positiveinterferents in this method. When acidified, sulfide formshydrogen sulfide, whi
19、ch passes through the gas diffusionmembrane and produces a signal at the silver electrode. Inaddition, sulfide ion reacts with cyanide ion in solution toreduce its concentration over time. Treat samples containingsulfide according to Section 11.4. During UV digestion, somesulfur compounds may produc
20、e sulfide. TA2 reagent contains asulfide scrubber that can remove up to 50 mg/L S2-from thesystem prior to amperometric detection.6.3 Treat sample containing water-soluble aldehydes, suchas formaldehyde or acetaldehyde, by adding ethylenediaminesolution.6.4 Remove oxidizing agents that decompose cya
21、nides byadding ascorbic acid, or sodium arsenite.6.5 Thiocyanate can produce positive interference whenthey decompose to cyanide by UV irradiation or oxidation.This method uses 312 nm as the irradiation wavelength, whichkeeps thiocyanate interference from UV irradiation minimal.Use of Total Acid Rea
22、gentModified, TA1M (see 8.21) mini-mizes interference from thiocyanate.6.6 High concentrations of carbonate can result in a nega-tive response in the amperometric detector when carbondioxide diffuses across the gas diffusion membrane into thealkaline receiving solution, reducing its pH. Treat efflue
23、ntsfrom high carbonate containing wastes, such as coal gasifica-tion waste and atmospheric emission scrub water, with hy-drated lime to stabilize the sample.6.7 High concentrations of surfactants interfere by changingthe characteristics of the gas diffusion membrane, allowingacid solution to pass th
24、rough the membrane and enter thedetector.6.8 Nitrate and nitrite if treated with sulfamic acid do notinterfere in this method.6.9 Sodium sulfite, sulfur dioxide, or sodium bisulfite do notinterfere at up to 2000 ppm SO3-.6.10 Sodium Thiosulfate exhibits a slight positive bias atconcentrations above
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD75112009E17500STANDARDTESTMETHODFORTOTALCYANIDEBYSEGMENTEDFLOWINJECTIONANALYSISINLINEULTRAVIOLETDIGESTIONANDAMPEROMETRICDETECTIONPDF

链接地址:http://www.mydoc123.com/p-525702.html