ASTM D7419-2013 red 0000 Standard Test Method for Determination of Total Aromatics and Total Saturates in Lube Basestocks by High Performance Liquid Chromatography (HPLC) with Refr.pdf
《ASTM D7419-2013 red 0000 Standard Test Method for Determination of Total Aromatics and Total Saturates in Lube Basestocks by High Performance Liquid Chromatography (HPLC) with Refr.pdf》由会员分享,可在线阅读,更多相关《ASTM D7419-2013 red 0000 Standard Test Method for Determination of Total Aromatics and Total Saturates in Lube Basestocks by High Performance Liquid Chromatography (HPLC) with Refr.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7419 07D7419 13Standard Test Method forDetermination of Total Aromatics and Total Saturates inLube Basestocks by High Performance LiquidChromatography (HPLC) with Refractive Index Detection1This standard is issued under the fixed designation D7419; the number immediately following the
2、designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope Scope*1.1 This test method co
3、vers the determination of total aromatics and total saturates in additive-free lube basestocks using highperformance liquid chromatography (HPLC) with refractive index (RI) detection. This test method is applicable to samplescontaining total aromatics in the concentration range of 0.2 to 46 mass %.1
4、.1.1 Polar compounds, if present, are combined with the total aromatics. Precision was determined for basestocks with polarscontent 1.5 mL have been used successfully.6.9 Analytical Balanceaccurate to 60.0001 g.7. Reagents and Materials7.1 Heptane, HPLC grade. If necessary, dry solvent with molecula
5、r sieves and then filter before use.7.2 Dichloromethane, HPLC or UV grade. If necessary, dry solvent with molecular sieves and then filter before use.7.3 Octadecylbenzene, 97 % pure.7.4 Hexadecane, 98 % pure.8. Sampling8.1 Follow Practice D4057 or D4177, or a similar standard to obtain a representat
6、ive laboratory sample of the basestock. Mixwell before sampling.9. Preparation of Apparatus9.1 Set up the liquid chromatograph, injection system, columns, backflush valve, optional column oven, optional UV detector,refractive index detector and computing integrator in accordance with the manufacture
7、rs instructions and as depicted in Fig. 1.Insert the backflush valve so that the detector is always connected independently of the direction of flow through the column (seeFig. 1). Maintain the sample injection valve at the same temperature as the sample solution; in most cases this will be at roomt
8、emperature. To minimize drifts in signal, ensure that the ambient temperature is relatively constant during analysis and calibration.FIG. 1 Diagrammatic Representation of Liquid ChromatographD7419 1339.2 New commercial columns may be packed in water/methanol or other polar solvents. Before these col
9、umns can be used flushthem with dichloromethane followed with heptane before proceeding. Other suitable solvents that restore the required resolutionmay be used. If the resolution requirement is not met, the column may be reactivated by flushing it with additionaldichloromethane. If the resolution s
10、till cannot be attained it may be necessary to replace the column or purchase an appropriatecolumn from other vendors. Si60 silica gel was found effective in yielding acceptable resolution and performance when properlyconditioned. When not analyzing samples, column may be flushed with a low flow of
11、heptane such as 0.1 mL/min.9.2.1 Adjust the flow rate of the mobile phase to a constant 3.0 to 3.5 mL/min, and ensure the reference cell of the refractiveindex detector is full of mobile phase. Fill the reference cell as instructed by the manufacturer.9.2.2 To minimize drift, it is essential to make
12、 sure the reference cell of the RI detector is full of solvent. The best way toaccomplish this is either (1) to flush the mobile phase through the reference cell (then isolate the reference cell to preventevaporation of the solvent) immediately prior to analysis, or (2) to continuously make up for s
13、olvent evaporation by supplying asteady independent flow through the reference cell. The make-up flow is optimized so that reference and analytical cell mismatchdue to drying-out, temperature, or pressure gradients is minimized. Typically, this can be accomplished with a make-up flow setat one tenth
14、 of the analytical flow.9.3 Column Resolution and Capacity Factor:9.3.1 Prepare a system performance standard (SPS) by weighing hexadecane (1.0 6 0.1 g) and octadecylbenzene (1.0 6 0.1 g)into a 10 mL volumetric flask and filling to the mark with heptane. For the preparation of standards, use the sam
15、e source for theheptane as that used for the mobile phase. Ensure that the octadecylbenzene is completely dissolved in the mixture, for example,by using an ultrasonic bath.9.3.2 When operating conditions are steady, as indicated by a stable horizontal baseline of the RI detector, inject 10 L of theS
16、PS in the foreflush mode (backflush valve = OFF) and record the chromatogram using the data system. Fig. 2 gives an examplechromatogram of the SPS mixture.9.3.3 Ensure that the resolution between hexadecane and octadecylbenzene is five or greater as defined below. Calculate theresolution between hex
17、adecane and octadecylbenzene as follows:Resolution523t22t1!33y11y2!(1)where:t1 = retention time of the hexadecane peak in minutes,t2 = retention time of the octadecylbenzene peak in minutes,y1 = half-height width of the hexadecane peak in minutes, andy2 = half-height width of the octadecylbenzene pe
18、ak in minutes.If the resolution is less than five, verify that all system components are functioning correctly and that the chromatographic deadvolume has been minimized by using low dead volume connectors, tubing etcetera. Ensure that the mobile phase is of sufficientlyhigh quality. Finally, regene
19、rate or replace the column if necessary. The column may be regenerated by flushing withdichloromethane followed by heptane until the signal is relatively constant on the RI detector. If after regenerating the silicaFIG. 2 Chromatogram of System Performance Standard in Foreflush Mode for Determinatio
20、n of Resolution, Capacity Factor and Back-flush TimeD7419 134columns, the resolution is still less than 5 then replace the silica columns. Si60 was found to be an effective silica gel with properconditioning. For a proper analysis, a resolution of at least five is required.NOTE 1Resolution loss over
21、 time may occur if a heptane mobile phase of low water content is not used. Use heptane as specified in this method.If necessary, dry the heptane with the addition of activated molecular sieves, such as MS 5A and then filter with at least 0.45 micron HPLC filter beforeuse.9.3.4 Calculate the capacit
22、y factor, k, for octadecylbenzene from 9.3.2 as follows:Capacity Factor5k 5t22t1!t1!(2)where:t1 = retention time of the hexadecane peak in minutes,t2 = retention time of the octadecylbenzene peak in minutesEnsure that the capacity factor is 0.4.9.3.5 Using the determined retention times of the hexad
23、ecane and octadecylbenzene peaks in 9.3.2 calculate an approximateswitching valve backflush time, B, in seconds, using the following equation:B 5t110.13t22t1! (3)where:t1 = retention time of hexadecane in minutes, andt2 = retention time of octadecylbenzene in minutes.9.4 Once the backflush time is d
24、etermined, re-inject the SPS mixture with backflush in place and ensure that the backflush timeas observed as a signal marker on the chromatogram occurs at the base of the eluted saturate peak. The return to baseline shalldisplay as shown in Fig. 3, point B. This observation shall be made also for a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD74192013RED0000STANDARDTESTMETHODFORDETERMINATIONOFTOTALAROMATICSANDTOTALSATURATESINLUBEBASESTOCKSBYHIGHPERFORMANCELIQUIDCHROMATOGRAPHYHPLCWITHREFRPDF

链接地址:http://www.mydoc123.com/p-525408.html