ASTM D7419-2007 695 Standard Test Method for Determination of Total Aromatics and Total Saturates in Lube Basestocks by High Performance Liquid Chromatography (HPLC) with Refractiv.pdf
《ASTM D7419-2007 695 Standard Test Method for Determination of Total Aromatics and Total Saturates in Lube Basestocks by High Performance Liquid Chromatography (HPLC) with Refractiv.pdf》由会员分享,可在线阅读,更多相关《ASTM D7419-2007 695 Standard Test Method for Determination of Total Aromatics and Total Saturates in Lube Basestocks by High Performance Liquid Chromatography (HPLC) with Refractiv.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 7419 07An American National StandardStandard Test Method forDetermination of Total Aromatics and Total Saturates inLube Basestocks by High Performance LiquidChromatography (HPLC) with Refractive Index Detection1This standard is issued under the fixed designation D 7419; the number imm
2、ediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 Th
3、is test method covers the determination of totalaromatics and total saturates in additive-free lube basestocksusing high performance liquid chromatography (HPLC) withrefractive index (RI) detection. This test method is applicableto samples containing total aromatics in the concentrationrange of 0.2
4、to 46 mass %.1.1.1 Polar compounds, if present, are combined with thetotal aromatics. Precision was determined for basestocks withpolars content 1.5 mL have been used successfully.6.9 Analytical Balanceaccurate to 60.0001 g.7. Reagents and Materials7.1 Heptane, HPLC grade. If necessary, dry solvent
5、withmolecular sieves and then filter before use.7.2 Dichloromethane, HPLC or UV grade. If necessary, drysolvent with molecular sieves and then filter before use.7.3 Octadecylbenzene, $ 97 % pure.7.4 Hexadecane, $ 98 % pure.8. Sampling8.1 Follow Practice D 4057 or D 4177, or a similar standardto obta
6、in a representative laboratory sample of the basestock.Mix well before sampling.9. Preparation of Apparatus9.1 Set up the liquid chromatograph, injection system,columns, backflush valve, optional column oven, optional UVdetector, refractive index detector and computing integrator inaccordance with t
7、he manufacturers instructions and as de-picted in Fig. 1. Insert the backflush valve so that the detectoris always connected independently of the direction of flowthrough the column (see Fig. 1). Maintain the sample injectionvalve at the same temperature as the sample solution; in mostcases this wil
8、l be at room temperature. To minimize drifts insignal, ensure that the ambient temperature is relatively con-stant during analysis and calibration.9.2 New commercial columns may be packed in water/methanol or other polar solvents. Before these columns can beused flush them with dichloromethane follo
9、wed with heptanebefore proceeding. Other suitable solvents that restore therequired resolution may be used. If the resolution requirementis not met, the column may be reactivated by flushing it withadditional dichloromethane. If the resolution still cannot beattained it may be necessary to replace t
10、he column or purchasean appropriate column from other vendors. Si60 silica gel wasTABLE 1 Examples of Operating Conditions Used in Cooperative StudiesLab A Lab B Lab CSilica Column Varian, 50 cm length by 7.7 mm i.d. 5 m Si60 Varian, 50 cm by 7.7 mm Si60 (CP28526) Phenomenex, 2 x Si60 (10 by 250 mm,
11、 5 mCyano Column Alltech/YMC, 100 by 10 mm 10 m Waters/YMC, 100 by 12 mm 5 m YMC, 10 by 100 mm 5 mRI Detector Agilent 1200 Hewlett Packard RI, model HP1047A Shimadzu RID-10AHeptane Flow (mL/min) 3.5 mL/min 3.0 3.0Resolution 5 5-6 10.3Injected Volume (microlitres) 10 10 10D7419072found effective in y
12、ielding acceptable resolution and perfor-mance when properly conditioned. When not analyzingsamples, column may be flushed with a low flow of heptanesuch as 0.1 mL/min.9.2.1 Adjust the flow rate of the mobile phase to a constant3.0 to 3.5 mL/min, and ensure the reference cell of therefractive index
13、detector is full of mobile phase. Fill thereference cell as instructed by the manufacturer.9.2.2 To minimize drift, it is essential to make sure thereference cell of the RI detector is full of solvent. The best wayto accomplish this is either (1) to flush the mobile phasethrough the reference cell (
14、then isolate the reference cell toprevent evaporation of the solvent) immediately prior toanalysis, or (2) to continuously make up for solvent evapora-tion by supplying a steady independent flow through thereference cell. The make-up flow is optimized so that referenceand analytical cell mismatch du
15、e to drying-out, temperature, orpressure gradients is minimized. Typically, this can be accom-plished with a make-up flow set at one tenth of the analyticalflow.9.3 Column Resolution and Capacity Factor:9.3.1 Prepare a system performance standard (SPS) byweighing hexadecane (1.0 6 0.1 g) and octadec
16、ylbenzene (1.06 0.1 g) into a 10 mL volumetric flask and filling to the markwith heptane. For the preparation of standards, use the samesource for the heptane as that used for the mobile phase. Ensurethat the octadecylbenzene is completely dissolved in themixture, for example, by using an ultrasonic
17、 bath.9.3.2 When operating conditions are steady, as indicated bya stable horizontal baseline of the RI detector, inject 10 L ofthe SPS in the foreflush mode (backflush valve = OFF) andrecord the chromatogram using the data system. Fig. 2 gives anexample chromatogram of the SPS mixture.9.3.3 Ensure
18、that the resolution between hexadecane andoctadecylbenzene is five or greater as defined below. Calculatethe resolution between hexadecane and octadecylbenzene asfollows:Resolution 52 3 t22 t1!3 3 y11 y2!(1)where:t1= retention time of the hexadecane peak in minutes,t2= retention time of the octadecy
19、lbenzene peak in min-utes,y1= half-height width of the hexadecane peak in minutes,andy2= half-height width of the octadecylbenzene peak inminutes.If the resolution is less than five, verify that all systemcomponents are functioning correctly and that the chromato-graphic dead volume has been minimiz
20、ed by using low deadvolume connectors, tubing etcetera. Ensure that the mobilephase is of sufficiently high quality. Finally, regenerate orreplace the column if necessary. The column may be regener-ated by flushing with dichloromethane followed by heptaneuntil the signal is relatively constant on th
21、e RI detector. If afterregenerating the silica columns, the resolution is still less than5 then replace the silica columns. Si60 was found to be aneffective silica gel with proper conditioning. For a properanalysis, a resolution of at least five is required.NOTE 1Resolution loss over time may occur
22、if a heptane mobilephase of low water content is not used. Use heptane as specified in thismethod. If necessary, dry the heptane with the addition of activatedmolecular sieves, such as MS 5A and then filter with at least 0.45 micronHPLC filter before use.FIG. 1 Diagrammatic Representation of Liquid
23、ChromatographD74190739.3.4 Calculate the capacity factor, k, for octadecylbenzenefrom 9.3.2 as follows:Capacity Factor 5 k 5t2 t1!t1!(2)where:t1= retention time of the hexadecane peak in minutes,t2= retention time of the octadecylbenzene peak in minutesEnsure that the capacity factor is 0.4.9.3.5 Us
24、ing the determined retention times of the hexade-cane and octadecylbenzene peaks in 9.3.2 calculate an approxi-mate switching valve backflush time, B, in seconds, using thefollowing equation:B 5 t11 0.1 3 t2 t1! (3)where:t1= retention time of hexadecane in minutes, andt2= retention time of octadecyl
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD74192007695STANDARDTESTMETHODFORDETERMINATIONOFTOTALAROMATICSANDTOTALSATURATESINLUBEBASESTOCKSBYHIGHPERFORMANCELIQUIDCHROMATOGRAPHYHPLCWITHREFRACTIVPDF

链接地址:http://www.mydoc123.com/p-525407.html