ASTM D7094-2012e1 5010 Standard Test Method for Flash Point by Modified Continuously Closed Cup &40 MCCCFP&41 Tester《采用经改良连续闭杯试验仪测定闪点40 MCCCFP41的标准试验方法》.pdf
《ASTM D7094-2012e1 5010 Standard Test Method for Flash Point by Modified Continuously Closed Cup &40 MCCCFP&41 Tester《采用经改良连续闭杯试验仪测定闪点40 MCCCFP41的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7094-2012e1 5010 Standard Test Method for Flash Point by Modified Continuously Closed Cup &40 MCCCFP&41 Tester《采用经改良连续闭杯试验仪测定闪点40 MCCCFP41的标准试验方法》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7094 121Standard Test Method forFlash Point by Modified Continuously Closed Cup(MCCCFP) Tester1This standard is issued under the fixed designation D7094; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of las
2、t revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEEditorially removed sole source of supply footnote from 6.1 in February 2014.1. Scope*1.1 This test method covers the determinat
3、ion of the flashpoint of fuels, lube oils, solvents and other liquids by acontinuously closed cup tester utilizing a specimen size of 2mL, cup size of 7 mL, with a heating rate of 2.5C per minute.1.1.1 Apparatus requiring a specimen size of 1 mL, cup sizeof 4 mL, and a heating rate of 5.5C per minut
4、e must be runaccording to Test Method D6450.1.2 This flash point test method is a dynamic method anddepends on definite rates of temperature increase. It is one ofthe many flash point test methods available and every flashpoint test method, including this one, is an empirical method.NOTE 1Flash poin
5、t values are not a constant physical chemicalproperty of materials tested. They are a function of the apparatus design,the condition of the apparatus used, and the operational procedure carriedout. Flash point can, therefore, only be defined in terms of a standard testmethod and no general valid cor
6、relation can be guaranteed between resultsobtained by different test methods or where different test apparatus isspecified.1.3 This test method utilizes a closed but unsealed cup withair injected into the test chamber.1.4 This test method is suitable for testing samples with aflash point from 35 to
7、225C.NOTE 2Flash point determinations below 35C and above 225C maybe performed; however, the precision has not been determined below andabove these temperatures.1.5 If the users specification requires a defined flash pointmethod other than this method, neither this method nor anyother test method sh
8、ould be substituted for the prescribed testmethod without obtaining comparative data and an agreementfrom the specifier.1.6 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard. Temperatures are in degrees Celsius, pressure inkilo-Pa
9、scals.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For spec
10、ificwarning statements, see 7.2, 8.5, and 10.1.2.2. Referenced Documents2.1 ASTM Standards:2D93 Test Methods for Flash Point by Pensky-MartensClosed Cup TesterD4057 Practice for Manual Sampling of Petroleum andPetroleum ProductsD4177 Practice for Automatic Sampling of Petroleum andPetroleum Products
11、D6450 Test Method for Flash Point by Continuously ClosedCup (CCCFP) TesterD6708 Practice for Statistical Assessment and Improvementof Expected Agreement Between Two Test Methods thatPurport to Measure the Same Property of a MaterialE300 Practice for Sampling Industrial Chemicals2.2 ISO Standards:3IS
12、O Guide 34 Quality Systems Guidelines for the Produc-tion of Reference MaterialsISO Guide 35 Certification of Reference MaterialsGeneraland Statistical Principles3. Terminology3.1 Definitions:3.1.1 flash point, nthe lowest temperature corrected to apressure of 101.3 kPa at which application of an ig
13、nitionsource causes the vapors of a specimen of the sample to ignitemomentarily under specified conditions of the test.1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility ofSubcommittee D02.08 on Volati
14、lity.Current edition approved Dec. 1, 2012. Published March 2013. Originallyapproved in 2004. Last previous edition approved in 2009 as D709404(2009).DOI: 10.1520/D7094-12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For An
15、nual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.*A Summary of Changes section appears at the end of this stand
16、ardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.1.1.1 DiscussionFor the purpose of this test method, thetest specimen is deemed to have flashed when the hot flame ofthe ignited vapor causes an instantaneous pressure increase ofat
17、 least 20 kPa inside the closed measuring chamber.3.1.2 dynamic, adjthe condition where the vapor abovethe test specimen and the test specimen are not in temperatureequilibrium at the time at which the ignition source is applied.4. Summary of Test Method4.1 The lid of the test chamber is regulated t
18、o a temperatureat least 18C below the expected flash point.A2 6 0.2 mL testspecimen of a sample is introduced into the sample cup. Bothspecimen and cup are at a temperature at least 18C below theexpected flash point; cooled if necessary. The cup is then raisedand pressed onto the lid of specified di
19、mensions to form thecontinuously closed but unsealed test chamber with an overallvolume of 7.0 6 0.3 mL.4.2 After closing the test chamber, the temperatures of thetest specimen and the regulated lid are allowed to equilibrate towithin 1C. Then the lid is heated at a prescribed, constant rate.For the
20、 flash tests, a high voltage arc of defined energy isdischarged inside the test chamber at regular intervals. Aftereach ignition, a variable amount of air (see Table 1)isintroduced into the test chamber to provide the necessaryoxygen for the next flash test. The pressure inside the continu-ously clo
21、sed but unsealed test chamber remains at ambientbarometric pressure except for the short time during the airintroduction and at a flash point.4.3 After each arc application, the instantaneous pressureincrease above the ambient barometric pressure inside the testchamber is monitored. When the pressur
22、e increase exceeds 20kPa, the temperature at that point is recorded as the uncorrectedflash point, which is then corrected to barometric pressure.5. Significance and Use5.1 The flash point temperature is one measure of thetendency of the test specimen to form a flammable mixturewith air under contro
23、lled laboratory conditions. It is only oneof a number of properties which must be considered inassessing the overall flammability hazard of a material.5.2 Flash point is used in shipping and safety regulations todefine flammable and combustible materials and for classifica-tion purposes. This defini
24、tion may vary from regulation toregulation. Consult the particular regulation involved for pre-cise definitions of these classifications.5.3 This test method can be used to measure and describethe properties of materials in response to heat and an ignitionsource under controlled laboratory condition
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD70942012E15010STANDARDTESTMETHODFORFLASHPOINTBYMODIFIEDCONTINUOUSLYCLOSEDCUP40MCCCFP41TESTER 采用

链接地址:http://www.mydoc123.com/p-524473.html