ASTM D6977-2004e1 752 Standard Specification for Polychloroprene Examination Gloves for Medical Application.pdf
《ASTM D6977-2004e1 752 Standard Specification for Polychloroprene Examination Gloves for Medical Application.pdf》由会员分享,可在线阅读,更多相关《ASTM D6977-2004e1 752 Standard Specification for Polychloroprene Examination Gloves for Medical Application.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 6977 041Standard Specification forPolychloroprene Examination Gloves for MedicalApplication1This standard is issued under the fixed designation D 6977; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEEditorially corrected 7.5.1 in December 2008.1. Scope1.1 This specification provides certain requirements forpolychloroprene rubb
3、er gloves used in conducting medicalexaminations and diagnostic and therapeutic procedures.1.2 This specification covers polychloroprene rubber exami-nation gloves that fit either hand, paired gloves, and gloves bysize. It also provides for packaged sterile or non-sterile or bulknon-sterile polychlo
4、roprene rubber examination gloves.1.3 This specification is similar to that of D 3578 StandardSpecification for Rubber Examination Gloves, and D 6319Standard Specification for Nitrile Examination Gloves forMedical Applications.2. Referenced Documents2.1 ASTM Standards:2D 412 Test Methods for Vulcani
5、zed Rubber and Thermo-plastic Rubbers and Thermoplastic ElastomersTensionD 573 Test Method for Rubber-Deterioration in anAir OvenD 3578 Specification for Rubber Examination GlovesD 3767 Practice for Rubber-Measurement of DimensionsD 5151 Test Method for Detection of Holes in MedicalGlovesD 6124 Test
6、 Method for Residual Powder on MedicalD 6319 Specification for Nitrile Examination Gloves forMedical Application GlovesD 6355 Test Method for Human Repeat Insult Patch Testingof Medical Gloves2.2 ISO Standard:ISO 2859 Sampling Procedures and Tables for Inspectionby Attributes32.3 Other Document:U.S.
7、 Pharmacopeia43. Significance and Use3.1 This specification is intended as a referee procedure forevaluating the performance and safety of polychloroprenerubber examination gloves. It is not intended for testing prior toroutine lot release. The safe and proper use of polychloroprenerubber examinatio
8、n gloves is beyond the scope of this speci-fication.4. Material4.1 Any polychloroprene rubber compound that producesexamination gloves that meet the requirements of this specifi-cation are allowed.4.2 A lubricant that meets the current requirements of theU.S. Pharmacopoeia for absorbable dusting pow
9、der may beapplied to the glove. Other lubricants may be used if theirsafety and efficacy have been previously established.4.3 The inside and outside surface of the polychloroprenerubber examination gloves shall be free of talc.5. Performance Requirements5.1 Gloves, sampled in accordance with Section
10、 6, shallmeet the following referee performance requirements:5.1.1 Products comply with requirements for sterility whentested in accordance with 7.2 when labeled sterile,5.1.2 Shall comply with freedom from holes when tested inaccordance with 7.3,5.1.3 Have consistent physical dimensions in accordan
11、cewith 7.4,5.1.4 Have acceptable physical property characteristics inaccordance with 7.5,5.1.5 Have a powder residue limit in accordance with 7.6,and5.1.6 Have a recommended maximum powder limit inaccordance with 7.7.1This specification is under the jurisdiction ofASTM Committee D11 on Rubberand is
12、the direct responsibility of Subcommittee D11.40 on Consumer RubberProducts.Current edition approved Jan. 1, 2004. Published February 2004.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume
13、 information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036.4U.S. Pharmacopeia, latest edition, Mack Publishing Co., Easton, PA 19175.1Copyright ASTM International, 100 Barr H
14、arbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. Sampling6.1 For referee purposes, gloves shall be sampled fromfinished product, after sterilization when labeled sterile, andinspected in accordance with ISO 2859. The inspection levelsand acceptable quality levels (AQL) sh
15、all conform to thosespecified in Table 1, or as agreed upon between the purchaserand the seller, if the latter is more comprehensive.7. Referee Test Methods7.1 The following tests shall be conducted to ensure therequirements of Section 8, as prescribed in Table 1.7.2 Sterility TestTesting for steril
16、ity shall be conducted inaccordance with the latest edition of the U.S. Pharmacopeia.7.3 Freedom from HolesTesting for freedom from holesshall be conducted in accordance with Test Method D 5151.7.4 Physical Dimensions Test:7.4.1 The gloves shall comply with the dimension require-ments prescribed in
17、Table 2.7.4.2 The length shall be expressed in millimeters as mea-sured from the outside tip of the middle finger to the outsideedge of the cuff.7.4.3 The width of the palm shall be expressed in millime-ters as measured at a level between the base of the index fingerand the base of the thumb. Values
18、 of width per size other thanlisted shall meet the stated tolerance specified in Table 2.7.4.4 The minimum thickness shall be expressed in milli-meters as specified in Table 2 when using a dial or digitalmicrometer that meets requirements described in Test MethodsD 412 and Practice D 3767, and in th
19、e locations indicated inFig. 1. For referee tests, cutting the glove is necessary to obtainsingle-thickness measurements. (See Practice D 3767 for moreinformation.)7.5 Physical Requirements Test:7.5.1 Before and after accelerated aging, the gloves shallconform to the physical requirements specified
20、in Table 3.Tests shall be conducted in accordance with Test MethodsD 412. Die C is recommended.7.5.2 Accelerated AgingThe gloves shall be aged inaccordance with Test Method D 573. Test the gloves by eitherone of the following methods:7.5.2.1 After being subjected to a temperature of 70 6 2Cfor 166 6
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD69772004E1752STANDARDSPECIFICATIONFORPOLYCHLOROPRENEEXAMINATIONGLOVESFORMEDICALAPPLICATIONPDF

链接地址:http://www.mydoc123.com/p-524137.html