1、Designation: D 6698 07Standard Test Method forOn-Line Measurement of Turbidity Below 5 NTU in Water1This standard is issued under the fixed designation D 6698; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisi
2、on. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method is applicable to the on-line measure-ment of turbidity under 5 nephelometric turbidity units (NTU)in water.1.2
3、It is the users responsibility to ensure the validity of thistest method for waters of untested matrices.1.3 In this test method calibration standards are defined inNTU values, but other assigned turbidity units are assumed tobe equivalent.1.4 This method assigns traceable reporting units to the typ
4、eof respective technology that was used to perform the mea-surement. Units are numerically equivalent with respect to thecalibration standard. For example, a 1 NTU formazin standardis also equal to a 1FNU standard, a 1 FNRU standard and soforth.1.5 This standard does not purport to address all of th
5、esafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1129 Terminology Relating
6、 to WaterD 1193 Specification for Reagent WaterD 1889 Test Method for Turbidity of Water3D 2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD 3370 Practices for Sampling Water from Closed ConduitsD 3864 Guide for Continual On-Line Monitoring Sy
7、stemsfor Water Analysis2.2 Other StandardsEPA 180.1 Methods for Chemical Analysis of Water andWastes, Turbidity4GLI Method 24Hach Method 8195 Determination of Turbidity byNephelometry EMMC Format5ISO 7027 Determination of Turbidity6Standard Method 2130B72.3 Other DocumentsUS Patent 4,283,1438US Pate
8、nt 4,291,9808US Patent 5,777,01153. Terminology3.1 DefinitionsFor definitions of terms used in thismethod refer to Terminology D 1129.3.2 Definitions of Terms Specific to This Standard:3.2.1 calibration turbidity standarda turbidity standardthat is traceable and equivalent to the reference turbidity
9、standard to within statistical errors, including commerciallyprepared 4000 NTU Formazin, stabilized formazin, and sty-renedivinylbenzene (SDVB). These standards may be used tocalibrate the instrument.NOTE 1Calibration standards may be instrument specific.3.2.2 calibration verification standardsdefin
10、ed standardsused to verify the accuracy of a calibration in the measurementrange of interest. These standards may not be used to performcalibrations, only calibration verifications. Included standardsare opto-mechanical light scatter devices, gel-like standards, orany other type of stable liquid sta
11、ndard.1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.03 on Sampling Water andWater-Formed Deposits, Analysis of Water for Power Generation and Process Use,On-Line Water Analysis, and Surveillance of WaterCurrent edition
12、approved June 15, 2007. Published July 2007. Originallyapproved in 2001. Last previous edition approved in 2001 as D 6698 01.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information,
13、refer to the standards Document Summary page onthe ASTM website.3Withdrawn.4Available from United States Environmental Protection Association (EPA),Ariel Rios Bldg., 1200 Pennsylvania Ave., NW, Washington, DC 20460, http:/www.epa.gov.5Hach Company, P.O. Box 389, Loveland, CO, 80539-0389, .6Available
14、 from International Organization for Standardization (ISO), 1 rue deVaremb, Case postale 56, CH-1211, Geneva 20, Switzerland, http:/www.iso.ch.7Standard Methods for the Examination of Water and Wastewater, 21st Edition,American Public HealthAssociation, Washington, DC, 2005, www.standardmethod-s.org
15、.8Available from AMCO Clear, P.O. Box 245, Powell, OH, 43065, www.amco-.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.NOTE 2Calibration verification standards may be instrument specific.3.2.3 in-situ nephelometera turbidimeter that
16、 determinesthe turbidity of a sample using a sensor that is placed directlyin the sample. This turbidimeter does not require transport ofthe sample to or from the sensor.3.2.4 nephelometric turbidity measurementthe measure-ment of light scatter from a sample in a direction that is at 90with respect
17、to the centerline of the incident light path. Unitsare NTU (Nephelometric Turbidity Units); when ISO 7027technology is employed, units are FNU (Formazin Nephelom-etric Units).3.2.5 ratio turbidity measurementthe measurement de-rived through the use of a nephelometric detector that serves asthe prima
18、ry detector and one or more other detectors used tocompensate for variation in incident light fluctuation, straylight, instrument noise, or sample color.3.2.6 reference turbidity standarda standard that is syn-thesized reproducibly from traceable raw materials by the user.All other standards are tra
19、ced back to this standard. Thereference standard for turbidity is formazin.3.2.7 seasoningthe process of conditioning labware withthe standard that will be diluted to a lower value. The processreduces contamination and dilution errors. See Appendix X2for suggested procedure.3.2.8 slip stream nephelo
20、meteran on-line turbidimeterthat determines the turbidity of a sample as the sample flowsthrough a sampling chamber. The sample is drawn from thesource into the turbidimeter, analyzed and then transported todrain.3.2.9 stray lightall light reaching the detector other thanthat contributed by the samp
21、le. For example: ambient lightleakage and divergent light in optical systems.3.2.10 turbidimeteran instrument that measures lightscatter using a nephelometric detector. Examples includephotoelectric nephelometers and ratio photoelectric nephelom-eters.3.2.11 turbidityan expression of the optical pro
22、perties of asample that causes light rays to be scattered and absorbedrather than transmitted in straight lines through the sample.(Turbidity of water is caused by the presence of matter such asclay, silt, finely divided organic matter, plankton, other micro-scopic organisms, organic acids, and dyes
23、.)4. Significance and Use4.1 Turbidity is undesirable in drinking water, plant effluentwaters, water for food and beverage processing, and for a largenumber of other water-dependent manufacturing processes.Removal of suspended matter is accomplished by coagulation,settling, and filtration. Measureme
24、nt of turbidity provides arapid means of process control to determine when, how, and towhat extent the water must be treated to meet specifications.4.2 This test method is suitable for the on-line monitoring ofturbidity such as that found in drinking water, process water,and high purity industrial w
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD66982007STANDARDTESTMETHODFORONLINEMEASUREMENTOFTURBIDITYBELOW5NTUINWATER 水中 混浊 低于 NTU 即时 测量 标准

链接地址:http://www.mydoc123.com/p-523272.html