ASTM D6481-2014 7731 Standard Test Method for Determination of Phosphorus Sulfur Calcium and Zinc in Lubrication Oils by Energy Dispersive X-ray Fluorescence Spectroscopy《采用能量扩散X射线.pdf
《ASTM D6481-2014 7731 Standard Test Method for Determination of Phosphorus Sulfur Calcium and Zinc in Lubrication Oils by Energy Dispersive X-ray Fluorescence Spectroscopy《采用能量扩散X射线.pdf》由会员分享,可在线阅读,更多相关《ASTM D6481-2014 7731 Standard Test Method for Determination of Phosphorus Sulfur Calcium and Zinc in Lubrication Oils by Energy Dispersive X-ray Fluorescence Spectroscopy《采用能量扩散X射线.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D6481 14Standard Test Method forDetermination of Phosphorus, Sulfur, Calcium, and Zinc inLubrication Oils by Energy Dispersive X-ray FluorescenceSpectroscopy1This standard is issued under the fixed designation D6481; the number immediately following the designation indicates the year of
2、original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the quantitative determinationof a
3、dditive elements in unused lubricating oils, as shown inTable 1.1.2 This test method is limited to the use of energydispersive X-ray fluorescence (EDXRF) spectrometers em-ploying an X-ray tube for excitation in conjunction with theability to separate the signals of adjacent elements.1.3 This test me
4、thod uses interelement correction factorscalculated from empirical calibration data.1.4 This test method is not suitable for the determination ofmagnesium and copper at the concentrations present in lubri-cating oils.1.5 This test method excludes lubricating oils that containchlorine or barium as an
5、 additive element.1.6 This test method can be used by persons who are notskilled in X-ray spectrometry. It is intended to be used as aroutine test method for production control analysis.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is th
6、eresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations to use.2. Summary of Test Method2.1 A specimen is placed in the X-ray beam, and theappropriate regions of its spectrum are measured to give th
7、efluorescent intensities of phosphorus, sulfur, calcium, and zinc.Other regions of the spectrum are measured to compensate forvarying background. If the detector does not completelyresolve all the elements in a single measurement, then toimprove selectivity, there is a combination of sequential ands
8、imultaneous measurements employing primary and secondarybeam filters. There can be correction of measured intensitiesfor spectral overlap. Concentrations of the elements of interestare determined by comparison of these intensities against acalibration curve using empirical interelement correction fa
9、c-tors and ratio to backscatter.2.2 The EDXRF spectrometer is initially calibrated using aset of prepared standards to collect the necessary intensity data.Each calibration line and any correction coefficient are ob-tained by a regression of this data, using the program suppliedwith the spectrometer
10、.3. Significance and Use3.1 Some oils are formulated with organo-metallicadditives, which act, for example, as detergents, antioxidants,and antiwear agents. Some of these additives contain one ormore of these elements: calcium, phosphorus, sulfur, and zinc.This test method provides a means of determ
11、ining the concen-trations of these elements, which in turn provides an indicationof the additive content of these oils.3.2 Several additive elements and their compounds areadded to the lubricating oils to give beneficial performance(Table 2).3.3 This test method is primarily intended to be used at a
12、manufacturing location for monitoring of additive elements inlubricating oils. It can also be used in central and researchlaboratories.4. Interferences4.1 The additive elements found in lubricating oils willaffect the measured intensities from the elements of interest toa varying degree. In general,
13、 for lubricating oils, theX-radiation emitted by the element of interest can be absorbedby itself (self-absorption) or by the other elements present inthe sample matrix. Also the X-radiation emitted from oneelement can further excite (enhance) another element. Theseinterelement effects are significa
14、nt at concentrations varyingfrom 0.03 mass %, due to the heavier elements, to 1 mass %,for the lighter elements. Enhancement effects can be minimisedby selective excitation. The measured concentration for a given1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products,
15、Liquid Fuels, and Lubricantsand is the direct responsibility ofSubcommittee D02.03 on Elemental Analysis.Current edition approved Dec. 1, 2014. Published January 2015. Originallyapproved in 1999. Last previous edition approved in 2010 as D6481 99 (2010).DOI: 10.1520/D6481-14.*A Summary of Changes se
16、ction appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1element can be mathematically corrected for self-absorptionand for interelement effects by other elements present in thesample matrix. If an elem
17、ent is present at significant concen-trations and an inter-element correction for that element is notemployed, the results can be low due to absorption or high dueto enhancement.4.2 If a sample containing barium as an additive above 0.03mass % is measured against a calibration derived from stan-dard
18、s without barium, then results will be low.4.3 If a sample containing chlorine as an impurity above0.03 mass % is measured against a calibration derived fromstandards without chlorine, then the results can be affected.4.4 There can be spectral overlap of one element ontoanother, especially for phosp
19、horus on sulfur, and the instru-ment must include correction procedures for any such overlaps.5. Apparatus5.1 Energy Dispersive X-ray Fluorescent AnalyzerAnyenergy dispersive X-ray fluorescent analyzer can be used if itsdesign incorporates at least the following features.5.1.1 Source of X-ray Excita
20、tion , X-ray tube withpalladium, silver, or rhodium target programmable between 4and at least 25 keV for preferential excitation to simplify thesample spectra. (WarningOperation of an analyzer using anX-ray tube source is to be conducted in accordance with themanufacturers safety instructions and fe
21、deral, state, and localregulations)5.1.2 X-ray Detector, gas filled proportional counter withhigh sensitivity and a resolution value not to exceed 1300 eVat 5.9 keV.NOTE 1The limited data from instruments with solid state detectors inthe inter-laboratory precision study did not support their inclusi
22、on in themethod.5.1.3 Primary Beam Filters, to make the excitation moreselective.5.1.4 Secondary Beam FiltersWhen a proportional coun-ter is used, these are necessary as a means of discriminatingbetween an analytes X rays and other analytes and thespectrum from the X-ray tube.5.1.5 Multi-Channel Ana
23、lyzer, for discrimination betweenan analytes X rays and background X rays.5.1.6 Optional Helium Purgeable Optical Path.5.2 Sample Cells, providing a depth of at least 6 mm andequipped with replaceable X-ray transparent film. Suitablefilms include polypropylene and polycarbonate with thicknessfrom 3.
24、5 to 8 m.5.3 Instrument Setting-Up Samples (Elemental ReferenceSamples), to quantify spectral overlaps. These are requiredwhen the instruments software does not include referencespectra to deconvolute spectra.5.4 Drift Correction Monitors, to correct for instrumentaldrift. At least two samples are n
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD648120147731STANDARDTESTMETHODFORDETERMINATIONOFPHOSPHORUSSULFURCALCIUMANDZINCINLUBRICATIONOILSBYENERGYDISPERSIVEXRAYFLUORESCENCESPECTROSCOPY

链接地址:http://www.mydoc123.com/p-522636.html