ASTM D5996-2016 red 3961 Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography《采用在线离子色谱法测定高纯水中阴离子污染物的标准试验方法》.pdf
《ASTM D5996-2016 red 3961 Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography《采用在线离子色谱法测定高纯水中阴离子污染物的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5996-2016 red 3961 Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography《采用在线离子色谱法测定高纯水中阴离子污染物的标准试验方法》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5996 05 (Reapproved 2009)D5996 16Standard Test Method forMeasuring Anionic Contaminants in High-Purity Water byOn-Line Ion Chromatography1This standard is issued under the fixed designation D5996; the number immediately following the designation indicates the year oforiginal adoption o
2、r, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers on-line analysis of high-purity water by the ion chrom
3、atography technique. This test method isapplicable for measuring various anionic contaminants in high-purity water, typically in the range of 0.01 to 100 g/L. This testmethod is used to determine the concentration of acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in acontinuou
4、sly flowing sample. The range of the test method is only as good as the reagent water available for preparing standards.At extremely low concentrations, 1.0 g/L, preparing standards is difficult, and extra care must be taken in their preparation. Thesample may have to be conditioned from higher pres
5、sures and temperatures to conditions that are suitable for use by on-lineinstruments.1.2 Online sample analysis of flowing streams does not lend itself to collaborative studies due to the nature of the sample andthe possibility of contamination that may result from handling the sample as part of the
6、 collaborative study. Therefore this standardtest method is not based on the results of a collaborative study but is intended to provide the best possible guidance for doing thistype of analysis.1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are inclu
7、ded in this standard.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to
8、use.2. Referenced Documents2.1 ASTM Standards:2D1066 Practice for Sampling SteamD1129 Terminology Relating to WaterD1192 Guide for Equipment for Sampling Water and Steam in Closed Conduits (Withdrawn 2003)3D1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias of
9、 Applicable Test Methods of Committee D19 on WaterD3370 Practices for Sampling Water from Closed ConduitsD3864 Guide for On-Line Monitoring Systems for Water AnalysisD4453 Practice for Handling of High Purity Water SamplesD5542 Test Methods for Trace Anions in High Purity Water by Ion Chromatography
10、D5810 Guide for Spiking into Aqueous SamplesD5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis3. Terminology3.1 For definitions of terms used in this test method, refer to Terminology D1129.3.1 Definitions:3.1.1 For definitions of terms used in thi
11、s standard, refer to Terminology D1129.1 This test method is under the jurisdiction of ASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.03 on Sampling Water andWater-Formed Deposits, Analysis of Water for Power Generation and Process Use, On-Line Water Analysis, and S
12、urveillance of Water.Current edition approved Oct. 1, 2009Feb. 15, 2016. Published November 2009June 2016. Originally approved in 1996. Last previous edition approved in 20052009 asD5996 05.D5996 05 (2009). DOI: 10.1520/D5996-05R09.10.1520/D5996-16.2 For referencedASTM standards, visit theASTM websi
13、te, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 The last approved version of this historical standard is referenced on www.astm.org.This document is not an AS
14、TM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In a
15、ll cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2 Defin
16、itions of Terms Specific to This Standard:3.2.1 analytical column, na column used to separate the anions of interest.3.2.2 analytical column set, na combination of one or more guard columns followed by one or more analytical columns.3.2.3 anion suppressor device, na device that is placed between the
17、 analytical columns and the detector. Its purpose is toinhibit detector response to the ionic constituents in the eluant, so as to lower the detector background and at the same time enhancedetector response to the ions of interest.3.2.3.1 DiscussionIts purpose is to inhibit detector response to the
18、ionic constituents in the eluant, so as to lower the detector background and at thesame time enhance detector response to the ions of interest.3.2.4 breakthrough volume, nthe maximum sample volume that can be passed through a concentrator column before the leasttightly bound ion of interest is elute
19、d. All of the columns in series contribute to the overall capacity of the analytical column set.3.2.4.1 DiscussionAll of the columns in series contribute to the overall capacity of the analytical column set.3.2.5 concentrator column, nan ion exchange column used to concentrate the ions of interest a
20、nd thereby increase methodsensitivity.3.2.6 eluant, nthe ionic mobile phase used to transport the sample through the analytical column.3.2.7 guard column, na column used before the analytical column to protect it from contaminants, such as particulate matteror ionic species that may chemically foul
21、the resins and degrade their performance.3.2.8 ion chromatography, na form of liquid chromatography in which ionic constituents are separated by ion exchangefollowed by a suitable detection means.3.2.9 resolution, nthe ability of an analytical column to separate constituents under specific test cond
22、itions.4. Summary of Test Method4.1 A continuously flowing sample is injected into the instrument through a sample injection valve. The sample is pumpedthrough a concentrator column where the anions of interest are collected on ion-exchange resin.After a suitable volume of samplehas been passed thro
23、ugh the concentrator column, sample flow is diverted and an eluant is pumped through the concentratorcolumn to remove the trapped anions. This eluant then flows through an analytical column set where the anions are separated basedon the retention characteristic of each anion relative to the eluant u
24、sed. The eluant stream containing the anions of interest passesthrough a suppressor device where the cations from the eluant are exchanged for hydrogen ions, converting the anions to their acidform. After the suppressor device, the eluant solution passes through a conductivity detector where the sep
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD59962016RED3961STANDARDTESTMETHODFORMEASURINGANIONICCONTAMINANTSINHIGHPURITYWATERBYONLINEIONCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-521261.html