ASTM D5578-2004(2010) Standard Test Method for Determination of Ethylene Oxide in Workplace Atmospheres (HBr Derivatization Method)《工作场所空气中环氧乙烯测定的标准试验方法》.pdf
《ASTM D5578-2004(2010) Standard Test Method for Determination of Ethylene Oxide in Workplace Atmospheres (HBr Derivatization Method)《工作场所空气中环氧乙烯测定的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5578-2004(2010) Standard Test Method for Determination of Ethylene Oxide in Workplace Atmospheres (HBr Derivatization Method)《工作场所空气中环氧乙烯测定的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5578 04 (Reapproved 2010)Standard Test Method forDetermination of Ethylene Oxide in Workplace Atmospheres(HBr Derivatization Method)1This standard is issued under the fixed designation D5578; the number immediately following the designation indicates the year oforiginal adoption or, in
2、 the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers a method of collecting andanalyzing samples to determine th
3、e amount of ethylene oxide(ETO) present in workplace atmospheres.1.2 This test method can be used to provide a time-weightedaverage (TWA) over the concentration range from 0.2 to 4 ppm(v).1.3 This test method can be used to determine 15-minexcursions (STEL) ranging from 1 to 25 ppm (v).1.4 The value
4、s stated in SI units are to be regarded as thestandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility
5、 of regulatory limitations prior to use. See Section 9 forspecific safety hazards.2. Referenced Documents2.1 ASTM Standards:2D1356 Terminology Relating to Sampling and Analysis ofAtmospheresD3686 Practice for Sampling Atmospheres to Collect Or-ganic Compound Vapors (Activated Charcoal Tube Ad-sorpti
6、on Method)D3687 Practice for Analysis of Organic Compound VaporsCollected by the Activated Charcoal Tube AdsorptionMethodE355 Practice for Gas Chromatography Terms and Rela-tionships2.2 Other Standard:3Occupational Safety and Health Administration, U.S. De-partment of Labor, Title 29, Code of Federa
7、l Regulations,Part 1910, Subpart Z, Section 1910.1047.3. Terminology3.1 For definitions of terms used in this test method, refer toTerminology D1356, and Practice E355.4. Summary of Test Method4.1 A known volume of air is pumped through a glass tubepacked with carbon molecular sieve, surface area 40
8、0 m2/gimpregnated with hydrogen bromide (HBr) where ETO isadsorbed and converted to 2-bromoethanol.4.2 The tube contains two reactive sections for samplecollection. The backup section collects vapors that passthrough the front section and is used to determine if thecollection capacity of the front s
9、ection has been exceeded.4.3 The resultant derivative, 2-bromoethanol, is desorbedwith a mixture of acetonitrile/toluene and analyzed using a gaschromatograph equipped with an electron capture detector.4.4 Desorption efficiency is determined by spiking tubeswith known amounts of 2-bromoethanol and a
10、nalyzing with thesame procedure used for air samples.4.5 Quantitation is achieved by comparing peak areas fromsample solutions with areas from standard solutions.5. Significance and Use5.1 Ethylene oxide is a major industrial chemical withproduction volume ranked in the top 25 chemicals produced int
11、he United States. It is used in the manufacture of a greatvariety of products as well as being a sterilizing agent andfumigant.5.2 This test method provides a means of determiningexposure levels of ETO in the working environment at thepresently recommended exposure guidelines.5.2.1 OSHA Permissible
12、Exposure Limit (PEL) 1 ppm,15-min excursion limit 5 ppm (CFR, Part 1910, Subpart Z,Section 1910.1047).35.2.2 ACGIH Threshold Limit Value (TLV) 1 ppm (1.8mg/m3), suspected human carcinogen.41This test method is under the jurisdiction of ASTM Committee D22 on AirQuality and is the direct responsibilit
13、y of Subcommittee D22.04 on Workplace AirQuality.Current edition approved April 1, 2010. Published June 2010. Originallyapproved in 1994. Last previous edition approved in 2004 as D5578 - 04. DOI:10.1520/D5578-04R10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM
14、 Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from Superintendent of Documents, U.S. Government PrintingOffice, Washington, DC 20402.4Threshold Limit Values for Chemical Substances
15、 and Physical Agents andBiological Exposure Indices, American Conference of Governmental IndustrialHygienists, 6500 Glenway Avenue, Building D-7, Cincinnati, OH 45211-4438.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. Interferen
16、ces6.1 Derivatives and other compounds that have identical ornearly the same GC column retention time as 2-bromoethanolduring the gas chromatographic analysis will interfere.6.2 Interferences can sometimes be resolved by altering gaschromatographic operating conditions. The identity of sus-pected 2-
17、bromoethanol, or the presence of 2-bromoethanolmasked by a chromatographic interference, or both, can beverified by gas chromatography/mass spectrometry.7. Apparatus7.1 Carbon Molecular Sieve, surface area 400 m2/g, HBrsampling tube.7.1.1 Preparation of Collection MediumAdd 20 mL ofHBr (24 %) to 70
18、g of carbon molecular sieve, surface area 400m2/g, in a glass jar. Cap the jar and mix the contents thoroughlyfor 5 min by rotating.Allow to equilibrate and dry overnight orfor 12 h.7.1.2 Tube PreparationInsert a plug of silanized glasswool into a 10-cm by 6-mm outside diameter (4-mm insidediameter)
19、 glass tube. Pack the front section of the tube with400 mg of the reactive adsorbent (7.1.1), using gentle tappingor vibration to promote uniform packing. Insert another plug ofsilanized wool and pack 200 mg of the adsorbent in the backupsection. Hold the backup section in place by firmly inserting
20、anadditional glass wool plug. The tubes may be flame-sealed orsealed with polyethylene caps. Provide a numerical identifica-tion for each lot of tubes.7.1.3 Tube Holder, capable of preventing breakage andprotecting worker during sampling.7.1.4 High-Density Polyethylene or Polypropylene Caps,tight-fi
21、tting, for resealing used tubes.7.2 Pump and Tubing:7.2.1 Sampling Pumps, having stable low flow rates(610 % of set flow rate) within the range from 20 to 100mL/min for up to 8 h.7.2.2 Rubber or Plastic Tubing, 6-mm inside diameter, forconnecting collection tube to pump. All tubing must bedownstream
22、 (between tube and pump) of collection tube toprevent contamination or loss of sample.7.3 Vials, glass with PTFE-lined caps, 10 mL, for desorbingsamples and storing standards.7.4 Pipettes, 5 mL, for adding desorbing solution tosamples.7.5 Syringes, 10, 50, and 100-L syringes, for preparingstandards.
23、7.6 Gas-Tight Syringe, 2 L, with low dead-volume needle.7.7 Gas Chromatograph (GC):7.7.1 Gas Chromatograph, with an electron capture detectorand a suitable readout device.7.7.2 Chromatographic Column, packed or capillary col-umns in accordance with 7.7.2.1 and 7.7.2.2 have been foundsuitable for thi
24、s analysis.7.7.2.1 Packed, 3.7 m by 3 mm (12 ft by18 in.), stainlesssteel, packed with 10 % diethylene glycol succinate on diato-maceous earth, flux-calcined, silanized, 80/100 mesh.7.7.2.2 Capillary, 30-m by 0.53-mm inside diameter fusedsilica capillary column with polyethylene glycol phase.8. Reag
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD557820042010STANDARDTESTMETHODFORDETERMINATIONOFETHYLENEOXIDEINWORKPLACEATMOSPHERESHBRDERIVATIZATIONMETHOD

链接地址:http://www.mydoc123.com/p-520149.html