1、Designation: D 5464 07Standard Test Method forpH Measurement of Water of Low Conductivity1This standard is issued under the fixed designation D 5464; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A numb
2、er in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method is applicable to determine the pH ofwater samples with a conductivity of 2 to 100 S/cm over thepH range of 3 to 11. pH me
3、asurements of water of lowconductivity are problematic. Specifically, this test methodavoids contamination of the sample with atmospheric gasesand prevents volatile components of the sample from escaping.This test method provides for pH electrodes and apparatus thataddress additional considerations
4、discussed in Annex A2. Thistest method also minimizes problems associated with thesamples pH temperature coefficient when the operator usesthis test method to calibrate an on-line pH monitor or controller(see Appendix X1).1.2 This test method covers the measurement of pH in waterof low conductivity
5、with a lower limit of 2.0 S/cm, utilizinga static grab-sample procedure where it is not practicable totake a real-time flowing sample.NOTE 1Test Method D 5128 for on-line measurement is preferredover this method whenever possible. Test Method D 5128 is not subject tothe limited conductivity range, t
6、emperature interferences, potential KClcontamination, and time limitations found with this method.1.3 For on-line measurements in water with conductivity of100 S/cm and higher, see Test Method D 6569.1.4 For laboratory measurements in water with conductivityof 100 S/cm and higher, see Test Method D
7、1193.1.5 The values stated in SI units are to be regarded asstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the
8、applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1129 Terminology Relating to WaterD 1193 Specification for Reagent WaterD 1293 Test Methods for pH of WaterD 2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee
9、 D19 on WaterD 4453 Practice for Handling of Ultra-Pure Water SamplesD 5128 Test Method for On-Line pH Measurement of Waterof Low ConductivityD 6569 Test Method for On-Line Measurement of pH3. Terminology3.1 DefinitionsFor definitions of terms used in these testmethods, refer to Terminology D 1129.3
10、.2 Definitions of Terms Specific to This Standard:3.2.1 liquid junction potentiala dc potential which ap-pears at the point of contact between the reference electrodessalt bridge (also known as reference junction or diaphragm)and the sample solution. Ideally this potential is near zero, andis stable
11、. However, in low conductivity water it may changefrom its value in buffer solution by an unknown amount, and isa zero offset (1).34. Reagents4.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents conform to the specificat
12、ions of the Committee onAnalytical Reagents of the American Chemical Society wheresuch specifications are available.4Other grades may be used,1These test methods are under the jurisdiction of ASTM Committee D19 onWater and are the direct responsibility of Subcommittee D19.03 on Sampling Waterand Wat
13、er-Formed Deposits, Analysis of Water for Power Generation and ProcessUse, On-Line Water Analysis, and Surveillance of Water.Current edition approved Aug. 1, 2007. Published August 2007. Originallyapproved in 1993. Last previous edition approved in 2001 as D 5464 93 (2001).2For referenced ASTM stand
14、ards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The boldface numbers in parentheses refer to the list of references at the end ofthis
15、standard.4Reagent Chemicals, American Chemical Society Specifications, AmericanChemical Society, Washington, DC. For suggestions on the testing of reagents notlisted by the American Chemical Society, see Analar Standards for LaboratoryChemicals, BDH Ltd., Poole, Dorset, U.K., and the United States P
16、harmacopeiaand National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,MD.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.provided it is first ascertained that the reagent is of sufficientlyhigh purity to permit its
17、 use without lessening the accuracy ofthe determination.4.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean reagent water as definedby Type III of Specification D 1193.4.3 Commercial Buffer SolutionsCommercially availableprepared buffers traceable to NIST st
18、andards should be ad-equate to perform the calibration procedures in 10.1-10.4. Theexact pH of the buffer will change with temperature and thispH versus temperature data will be provided by the purveyor ofthe specific buffer. Refer to Test Methods D 1293 for thepreparation of reference buffer soluti
19、ons if desired.4.4 Buffer ACommercially available 7.0 pH buffer.4.5 Buffer BCommercially available 4.0 pH buffer.4.6 Buffer CCommercially available 9.0 or 10.0 pHbuffer.5. Summary of Test Method5.1 The pH meter and associated electrodes are first stan-dardized with two calibration pH buffer solution
20、s.5.2 A grab sample of high purity water is taken by means ofrinsing and filling two narrow mouth bottles at the samplepoint. Once each container is filled to the top with a represen-tative sample excluding any air, the container is capped and thesamples are transported quickly to a laboratory for a
21、nalysis.5.3 pH measurement of the sample is made with high puritywater pH calibration apparatus comprised of pH and referenceelectrodes, and automatic temperature compensator (if used).The first container is used to rinse the sensors and begintemperature equilibration and the second container is use
22、d formeasurement.5.4 A trace amount of KCl electrolyte enters calibrationbuffer solutions and samples via the controlled leakage rate ofthe reference electrode liquid junction (diaphragm) to stabilizethe liquid junction potential. Excessive KCl introduction fromthe electrode liquid junction into low
23、 ionic strength sampleswill increase solution conductivity, and may alter solution pH,and should be avoided.5.5 Temperature must be measured and both Solution Tem-perature Coefficient (STC) and Nernstian electrode effectscompensated, either manually to the measured value or auto-matically by the pH
24、meter. See Appendix X1 for a discussionof temperature effects.6. Significance and Use6.1 pH measurement of low conductivity water is frequentlyapplied to power plant water and condensed steam samples forcorrosion and scale prevention. It is sometimes used in purewater treatment systems between multi
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD54642007STANDARDTESTMETHODFORPHMEASUREMENTOFWATEROFLOWCONDUCTIVITY 传导 PH 测量 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-519881.html