ASTM D5153-2010(2016) Standard Test Method for Palladium in Molecular Sieve Catalyst by Atomic Absorption《采用原子吸收法测定分子筛催化剂中钯的标准试验方法》.pdf
《ASTM D5153-2010(2016) Standard Test Method for Palladium in Molecular Sieve Catalyst by Atomic Absorption《采用原子吸收法测定分子筛催化剂中钯的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5153-2010(2016) Standard Test Method for Palladium in Molecular Sieve Catalyst by Atomic Absorption《采用原子吸收法测定分子筛催化剂中钯的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5153 10 (Reapproved 2016)Standard Test Method forPalladium in Molecular Sieve Catalyst by AtomicAbsorption1This standard is issued under the fixed designation D5153; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the
2、 year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of palladiumin molecular sieve-containing fresh catalysts with abou
3、t 0.5weight % of palladium.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this
4、 standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterE177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE456 Termin
5、ology Relating to Quality and StatisticsE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test MethodD7442 Practice for Sample Preparation of Fluid CatalyticCracking Catalysts and Zeolites for ElementalAnalysis byInductively Coupled Plasma Atomic Emission Spectros-
6、copy2.2 U.S. Federal Specification:Federal Spec. NNN-P-395C Tolerance for Class A Pipets33. Summary of Test Method3.1 The test sample is treated with a mixture of sulfuric andhydrofluoric acids. Upon dissolution, the excess hydrofluoricacid is expelled. Aqua regia and lanthanum chloride are addedand
7、 the solution is diluted to a specific volume. Palladiumconcentration is determined by atomic absorption spectropho-tometry. Absorbance of the samples is bracketed using a set ofnarrow range, matrix matched standards. A second sample,taken at the same time as the analysis sample, is used todetermine
8、 loss on ignition.4. Significance and Use4.1 This test method provides a means of determining thepalladium content in fresh catalysts containing molecularsieves.4.2 This test method is not intended to cover samplescontaining precious metals other than palladium.5. Apparatus5.1 Analytical Balance, ca
9、pable of weighing to nearest0.1 mg.5.2 Atomic Absorption Spectrophotometer.5.3 Beakers, TFE-fluorocarbon, 100-mL.5.4 Crucibles, porcelain, 10-mL.5.5 Crucible Cover, porcelain, for 10-mL crucible.5.6 Desiccator.5.7 Flasks, Erlenmeyer, 2000-mL.5.8 Graduated Cylinders, glass, 5-mL, 10-mL, 25-mL,50-mL,
10、250-mL, 500-mL, 1000-mL.5.9 Graduated Cylinder, plastic, 10-mL.5.10 Hot Plate.5.11 Muffle Furnace, electrically heated, capable of 1000C.5.12 Pipets, 4-mL, 6-mL, 8-mL, 10-mL.5.13 Volumetric Flasks, 100-mL, 500-mL.5.14 Watch Glasses, TFE-fluorocarbon, for 100-mL beaker.5.15 Weighing Papers.6. Reagent
11、s and Materials6.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended that1This test method is under the jurisdiction of ASTM Committee D32 onCatalysts and is the direct responsibility of Subcommittee D32.03 on ChemicalComposition.Current
12、edition approved Jan. 1, 2016. Published January 2016. Originallyapproved in 1991. Last previous edition approved in 2010 as D515310. DOI:10.1520/D5153-10R16.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of A
13、STMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from Standardization Documents Order Desk, DODSSP, Bldg. 4,Section D, 700 Robbins Ave., Philadelphia, PA 19111-5098.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Consh
14、ohocken, PA 19428-2959. United States1all reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are available.4Other grades may beused provided it is first ascertained that the reagent is ofsufficiently high p
15、urity to permit its use without lessening theaccuracy of the determination.6.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean reagent water as definedby Specification D1193.6.3 Hydrochloric Acid, concentrated, 36.538.0 wt. % or 12M, sp gr 1.18.6.4 Hydrofluo
16、ric Acid, concentrated, 48.051.0 wt. % or28.9 M, sp gr 1.17.NOTE 1Refer to Section 8 of Practice D7442-08 for hazards associ-ated with handling of acids.6.5 Nitric Acid, concentrated, 69.071.0 wt. % or 15.7 M, spgr 1.41.6.6 Sulfuric Acid, concentrated, 95.098.0 wt. % or 18 M,sp gr 1.84.6.7 Aluminum
17、Oxide (Al2O3).6.8 Lanthanum Chloride (LaCl37H2O).6.9 Palladium Wire, 99.99 %.6.10 Aqua RegiaMix three parts by volume of concen-trated hydrochloric acid (12 M) and one part by volume ofconcentrated nitric acid (15.7 M) immediately before use.6.11 Sulfuric Acid, 15.816.3 wt. % or 3M. Cautiously add25
18、0 mL of concentrated sulfuric acid (18 M) to 1250 mL ofwatermix well and allow to cool.6.12 Palladium Standard Solution, 500 mg/L. Dissolve0.2500 6 0.0001 g of palladium wire (99.99 %) in 25 mL ofaqua regia. Evaporate the solution to dryness on a steam bath.Dissolve the remaining salts by addition o
19、f 25 mL of concen-trated hydrochloric acid (12 M) and 25 mL of distilled water.Transfer the solution to a 500 mL volumetric flask and dilute tovolume when cool.NOTE 2A commercially available atomic absorption palladium refer-ence solution may be used if it is known to be reliable.6.13 Lanthanum Chlo
20、ride SolutionDissolve 25.5 g oflanthanum chloride (LaCl37H2O) in distilled water, dilute to100 mL and mix well. This solution serves as an ionizationsuppressor in atomic absorption.6.14 Desiccant, molecular sieve, type 4A.7. Procedure7.1 Preparation of Calibration Standards:7.1.1 If the concentratio
21、n of aluminum oxide in the sampleis unknown, digest a representative sample in acid and deter-mine the aluminum concentration by atomic absorption spec-trophotometry.7.1.2 Transfer 30 mL of sulfuric acid (3M) to five TFE-fluorocarbon beakers.7.1.3 To each beaker add the equivalent weight of aluminum
22、oxide that would be present in 0.8000 g of sample asdetermined in 7.1.1.NOTE 3If it is available, 0.800 g of catalyst base material may besubstituted for the aluminum oxide.7.1.4 Transfer 0, 4, 6, 8 and 10 mL of the 500 mg/Lpalladium standard to the five TFE-fluorocarbon beakers.7.1.5 Add 10 mL of c
23、oncentrated hydrofluoric acid to eachbeaker.7.1.6 Cover the beakers with TFE-fluorocarbon lids anddigest on a hot plate at medium heat until all solid material isin solution (including any brown stains that may appear on thebeaker walls) and light fumes are evolved.NOTE 4If brown stains do not redis
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD515320102016STANDARDTESTMETHODFORPALLADIUMINMOLECULARSIEVECATALYSTBYATOMICABSORPTION 采用 原子 吸收 测定

链接地址:http://www.mydoc123.com/p-519005.html