ASTM D5110-1998(2010) Standard Practice for Calibration of Ozone Monitors and Certification of Ozone Transfer Standards Using Ultraviolet Photometry《使用紫外光度法进行臭氧监测器校准和臭氧转变标准认证的标准操作规.pdf
《ASTM D5110-1998(2010) Standard Practice for Calibration of Ozone Monitors and Certification of Ozone Transfer Standards Using Ultraviolet Photometry《使用紫外光度法进行臭氧监测器校准和臭氧转变标准认证的标准操作规.pdf》由会员分享,可在线阅读,更多相关《ASTM D5110-1998(2010) Standard Practice for Calibration of Ozone Monitors and Certification of Ozone Transfer Standards Using Ultraviolet Photometry《使用紫外光度法进行臭氧监测器校准和臭氧转变标准认证的标准操作规.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5110 98 (Reapproved 2010)Standard Practice forCalibration of Ozone Monitors and Certification of OzoneTransfer Standards Using Ultraviolet Photometry1This standard is issued under the fixed designation D5110; the number immediately following the designation indicates the year oforigina
2、l adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers a means for calibrating ambient,workplace, or
3、 indoor ozone monitors, and for certifyingtransfer standards to be used for that purpose.1.2 This practice describes means by which dynamicstreams of ozone in air can be designated as primary ozonestandards.1.3 The values stated in SI units are to be regarded asstandard. No other units of measuremen
4、t are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitatio
5、ns prior to use. See Section 8 forspecific precautionary statements.2. Referenced Documents2.1 ASTM Standards:2D1356 Terminology Relating to Sampling and Analysis ofAtmospheresD3195 Practice for Rotameter CalibrationD3249 Practice for General Ambient Air Analyzer Proce-duresD3631 Test Methods for Me
6、asuring Surface AtmosphericPressureD5011 Practices for Calibration of Ozone Monitors UsingTransfer Standards3E220 Test Method for Calibration of Thermocouples ByComparison TechniquesE591 Practice for Safety and Health Requirements Relatingto Occupational Exposure to Ozone3E644 Test Methods for Testi
7、ng Industrial Resistance Ther-mometers3. Terminology3.1 For definitions of terms used in this practice, refer toTerminology D1356.3.2 Definitions of Terms Specific to This Standard:3.2.1 primary standarda standard directly defined andestablished by some authority, against which all secondarystandard
8、s are compared.3.2.2 secondary standarda standard used as a means ofcomparison, but checked against a primary standard.3.2.3 standardan accepted reference sample or deviceused for establishing measurement of a physical quantity.3.2.4 transfer standarda type of secondary standard. It isa transportabl
9、e device or apparatus that, together with opera-tional procedures, is capable of reproducing pollutant concen-tration or producing acceptable assays of pollutant concentra-tions.3.2.5 zero airpurified air that does not contain ozone, anddoes not contain any other component that may interfere withthe
10、 measurement (see 7.1).4. Summary of Practice4.1 This practice is based on the photometric assay of ozone(O3) concentrations in a dynamic flow system. The concentra-tion of O3in an absorption cell is determined from a measure-ment of the amount of 253.7 nm light absorbed by the sample.This determina
11、tion requires knowledge of (1)4the absorptioncoefficient of O3at 253.7 nm, (2) the optical path lengththrough the sample, (3) the transmittance of the sample at a1This practice is under the jurisdiction ofASTM Committee D22 onAir Qualityand is the direct responsibility of Subcommittee D22.03 on Ambi
12、ent Atmospheresand Source Emissions.Current edition approved Oct. 1, 2010. Published November 2010. Originallyapproved in 1990. Last previous edition approved in 2004 as D5110 - 98(2004).DOI: 10.1520/D5110-98R10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Cus
13、tomer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn.4The boldface numbers in parentheses refer to the references listed at the end ofthis practice.1Copyright ASTM International, 100 Barr Har
14、bor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.wavelength of 253.7 nm, and (4) the temperature and pressureof the sample. The transmittance is defined as the ratio:I/Iowhere:I = the intensity of light that passes through the cell and issensed by the detector when the cell co
15、ntains an O3sample, andIo= the intensity of light that passes through the cell and issensed by the detector when the cell contains zero air.It is assumed that all conditions of the system, except for thecontents of the absorption cell, are identical during measure-ments of I and Io. The quantities d
16、efined above are related bythe Beer-Lambert absorption law:Transmittance5I/Io5e2acd(1)where:a = absorption coefficient of O3at 253.7 nm,(308 6 4) 3 106ppm1cm1at 0C and 101.3 kPa (1atm) (1, 2, 3, 4, 5, 6, 7, 8)c =O3concentration, ppm, andd = optical path length, cm.4.1.1 In practice, a stable O3gener
17、ator (see 6.1.4) is used toproduce O3concentrations over the required range. Each O3concentration is determined from the measurement of thetransmittance of the sample at 253.7 nm, and is calculated fromthe equation:c52lnIIoad!(2)The calculated O3concentrations must be corrected for O3losses, which m
18、ay occur in the photometer, and for thetemperature and pressure of the sample.5. Significance and Use5.1 The reactivity and instability of O3preclude the storageof O3concentration standards for any practical length of time,and precludes direct certification of O3concentrations asStandard Reference M
19、aterials (SRMs). Moreover, there is noavailable SRM that can be readily and directly adapted to thegeneration of O3standards analogous to permeation devicesand standard gas cylinders for sulfur dioxide and nitrogenoxides. Dynamic generation of O3concentrations is relativelyeasy with a source of ultr
20、aviolet (UV) radiation. However,accurately certifying an O3concentration as a primary standardrequires assay of the concentration by a comprehensivelyspecified analytical procedure, which must be performed everytime a standard is needed (9).5.2 This practice is not designed for the routine calibrati
21、onof O3monitors at remote locations (see Practices D5011).6. Apparatus6.1 Atypical complete UV calibration system consists of anO3generator, an output port or manifold, a photometer, asource of zero air, and other components as necessary. Theconfiguration must provide a stable O3concentration at the
22、system output and allow the photometer to assay accurately theoutput concentration to the precision specified for the photom-eter. Fig. 1 shows the system, and illustrates the calibrationsystem. Ozone is highly reactive and subject to losses uponcontact with surfaces. All components between the O3ge
23、nera-tor and the photometer absorption cell shall be of inertmaterial, such as glass or TFE-fluorocarbon. Lines and inter-connections shall be as short as possible, and all surfaces shallbe chemically clean. For certification of transfer standards thatprovide their own source of O3, the generator an
24、d possiblyother components shown in Fig. 1 may not be required (seePractices D5011).6.1.1 UV Photometer, consisting of a low-pressure mercurydischarge lamp, collimation optics (optional), an absorptioncell, a detector, and signal-processing electronics, as shown inFig. 1. It shall be capable of meas
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD511019982010STANDARDPRACTICEFORCALIBRATIONOFOZONEMONITORSANDCERTIFICATIONOFOZONETRANSFERSTANDARDSUSINGULTRAVIOLETPHOTOMETRY

链接地址:http://www.mydoc123.com/p-518891.html