ASTM D4878-2008 Standard Test Methods for Polyurethane Raw Materials Determination of Viscosity of Polyols《聚氨酯原料标准试验方法 测定多醇的粘度》.pdf
《ASTM D4878-2008 Standard Test Methods for Polyurethane Raw Materials Determination of Viscosity of Polyols《聚氨酯原料标准试验方法 测定多醇的粘度》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4878-2008 Standard Test Methods for Polyurethane Raw Materials Determination of Viscosity of Polyols《聚氨酯原料标准试验方法 测定多醇的粘度》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 4878 08Standard Test Methods forPolyurethane Raw Materials: Determination of Viscosity ofPolyols1This standard is issued under the fixed designation D 4878; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of
2、 last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods (A and B) determine the viscosity ofpolyols in the range from 10 to 100 000 mPas(cP) at 25C orat 5
3、0C. Test Method A also applies to more viscous samplesthat are soluble in n-butyl acetate. Test Method B is simply areference to a general procedure for kinematic viscosity,D 445. (See Note 1.)1.2 The values stated in SI units are to be regarded as thestandard. Other equivalent units are provided be
4、cause ofcurrent common usage.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations
5、 prior to use.NOTE 1There is no equivalent ISO standard for Test Method Aalthough ISO 3219 is similar. Test Method B is equivalent to ISO 3104.2. Referenced Documents2.1 ASTM Standards:2D 445 Test Method for Kinematic Viscosity of Transparentand Opaque Liquids (and Calculation of Dynamic Viscos-ity)
6、D 883 Terminology Relating to PlasticsE1 Specification for ASTM Liquid-in-Glass Thermometers2.2 ISO Standards:3ISO 3104ISO 32193. Terminology3.1 For definitions of terms used in these test methods seeTerminology D 883.4. Significance and Use4.1 These test methods are suitable for research or as qual
7、itycontrol or specification tests.4.2 Viscosity measures the resistance of a fluid to uniformlycontinuous flow without turbulence or other forces.5. Sampling5.1 Polyesters and polyethers usually contain moleculescovering an appreciable range of molecular weights. Thesehave a tendency to fractionate
8、during solidification. Unless thematerial is a finely ground solid it is necessary to melt (usingno higher temperature than necessary) and mix the resin wellbefore removing a sample for analysis. Many polyols arehygroscopic and care should be taken to provide minimumexposure to atmospheric moisture
9、during the sampling.TEST METHOD ABROOKFIELD VISCOSITY6. Summary of Test Method6.1 The viscosity of resins is measured by determining thetorque on a spindle rotating at constant speed in the liquidsample which is adjusted to 25 6 0.1C. Samples with viscosi-ties exceeding 100 000 mPas(cP) at 50C are d
10、issolved inn-butyl acetate (or other solvent) and the viscosity is deter-mined at 25 6 0.1C.7. Apparatus7.1 Constant-Temperature Bath, capable of maintainingtemperatures of 25 6 0.1C and 50 6 0.1C should be used.Water, water and glycerin, or oil may be used as the heatingmedium and the bath should b
11、e provided with heating, stirring,and thermostatting devices.7.2 Bath and Sample Thermometers, graduated in 0.1Csubdivisions and standardized for the range of use to thenearest 0.01C. ASTM Saybolt Viscosity Thermometers hav-ing ranges from 19 to 27C and 49 to 57C, as specified, and1These test method
12、s are under the jurisdiction of ASTM Committee D20 onPlastics and is the direct responsibility of Subcommittee D20.22 on CellularMaterials - Plastics and Elastomers.Current edition approved Nov. 1, 2008. Published November 2008. Originallyapproved in 1988. Last previous edition approved in 2003 as D
13、 4878 - 03.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Instit
14、ute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.conforming to the requirements for Thermometers 17C and19C, respectively, as prescribed in Specification E1a
15、re rec-ommended.7.3 Brookfield Synchrolectric Viscometer4Model LVFwith speeds of 60, 30, 12, and 6 r/min is to be used whenavailable. It is applicable to the range of 10 to 100 000mPas(cP). If this model is not available, Model RVF or HAFmay be substituted. However, samples should be heated ordissol
16、ved in the standard way to keep the measured viscositybelow 100 000 mPas(cP) so that the test may be repeated inother laboratories under similar conditions with Model LVF.The calibration of the instrument should be checked periodi-cally by measuring the viscosity of Brookfield EngineeringLaboratorie
17、s viscosity standard fluids.4Standard fluids L-2,L-3, R-1, R-2, H-1 are suitable for the usual range. Thecalibration corrections should be applied to sample measure-ments.8. Solvent8.1 n-Butyl Acetate, reagent grade.9. Preparation of Sample9.1 The preparation of a homogeneous sample is of primaryimp
18、ortance in viscosity measurements. A nonuniform tempera-ture distribution as well as the presence of air bubbles andtraces of extraneous material should be avoided. Resins are noteasily made homogeneous with respect to temperature, there-fore, the sample should be thoroughly mixed and the tempera-tu
19、re measured at several locations in the sample vessel beforedetermining the viscosity.10. Preparation of Apparatus10.1 Attach the viscometer with an adjustable clamp to aring stand. Adjust the legs at the base of the ring stand until thebubble is in the center of the spirit level on the viscometer.A
20、ttach the spindle that applies to the range expected for thesample (see Section 12).11. Choice of Temperature11.1 Samples that are liquid and have a viscosity of less than100 000 mPas(cP) at 25C should be measured at that tem-perature. Materials that fulfill this requirement only whenheated from 25
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD48782008STANDARDTESTMETHODSFORPOLYURETHANERAWMATERIALSDETERMINATIONOFVISCOSITYOFPOLYOLS 聚氨酯 原料 标准

链接地址:http://www.mydoc123.com/p-518291.html