ASTM D4841-1988(2013) Standard Practice for Estimation of Holding Time for Water Samples Containing Organic and Inorganic Constituents《含有机和无机成分水样保留时间估算的标准实施规程》.pdf
《ASTM D4841-1988(2013) Standard Practice for Estimation of Holding Time for Water Samples Containing Organic and Inorganic Constituents《含有机和无机成分水样保留时间估算的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4841-1988(2013) Standard Practice for Estimation of Holding Time for Water Samples Containing Organic and Inorganic Constituents《含有机和无机成分水样保留时间估算的标准实施规程》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D4841 88 (Reapproved 2013)Standard Practice forEstimation of Holding Time for Water Samples ContainingOrganic and Inorganic Constituents1This standard is issued under the fixed designation D4841; the number immediately following the designation indicates the year oforiginal adoption or,
2、 in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the means of estimating the periodof time during which a wa
3、ter sample can be stored aftercollection and preservation without significantly affecting theaccuracy of analysis.1.2 The maximum holding time is dependent upon thematrix used and the specific analyte of interest. Therefore,water samples from a specific source must be tested todetermine the period o
4、f time that sample integrity is maintainedby standard preservation practices.1.3 In the event that it is not possible to analyze the sampleimmediately at the time of collection, this practice does notprovide information regarding degradation of the constituent ofinterest or changes in the matrix tha
5、t may occur from the timeof sample collection to the time of the initial analysis.1.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated wi
6、th its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD1192 Guide for Equipment for S
7、ampling Water and Steamin Closed Conduits (Withdrawn 2003)3D1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD3694 Practices for Preparation of Sample Containers andfor Preservation of Organic ConstituentsD4
8、210 Practice for Intralaboratory Quality Control Proce-dures and a Discussion on Reporting Low-Level Data(Withdrawn 2002)3D4375 Practice for Basic Statistics in Committee D19 onWaterE178 Practice for Dealing With Outlying Observations3. Terminology3.1 Definitions:3.1.1 For definitions of terms used
9、in this practice, refer toTerminology D1129.3.1.2 criterion of detectionthe minimum quantity thatmust be observed before it can be stated that a substance hasbeen discerned with an acceptable probability that the state-ment is true (see Practice D4210).3.2 Definitions of Terms Specific to This Stand
10、ard:3.2.1 maximum holding timethe maximum period of timeduring which a properly preserved sample can be stored beforesuch degradation of the constituent of interest or change insample matrix occurs that the systematic error exceeds the99 % confidence interval (not to exceed 15 %) of the testcalculat
11、ed around the mean concentration found at zero time.3.2.2 acceptable holding timeany period of time less thanor equal to the maximum holding time.4. Summary of Practice4.1 Holding time is estimated by means of replicate analy-ses at discrete time intervals using a large volume of a watersample that
12、has been properly collected and preserved. Asufficient number of replicate analyses are performed to main-tain the 99 % confidence interval within 15 % of the concen-tration found at zero time. Concentration of the constituent ofinterest is plotted versus time. The maximum holding time isthe period
13、of time from sample collection to such time thatdegradation of the constituent of interest or change in samplematrix occurs and the systematic error exceeds the 99 %confidence interval (not to exceed 15 %) of the test calculatedaround the mean concentration at zero time. Prior to the1This practice i
14、s under the jurisdiction of ASTM Committee D19 on Water andis the direct responsibility of Subcommittee D19.02 on Quality Systems,Specification, and Statistics.Current edition approved Jan. 1, 2013. Published January 2013. Originallyapproved in 1988. Last previous edition approved in 2008 as D4841 8
15、8 (2008).DOI: 10.1520/D4841-88R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version
16、of this historical standard is referenced onwww.astm.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1determination of holding time, each laboratory must generateits own precision data in matrix water. These data are comparedto the
17、 pooled single-operator precision data on reagent waterreported in the test method and, the less precise of the two setsof data are used in the calculation.NOTE 1This practice generates only limited data which may not leadto consistent conclusions each time that the test is applied. In cases whereth
18、e concentration of the constituent of interest changes gradually over anextended period of time, the inherent variability in test results may lead tosomewhat different conclusions each time that this practice is applied.5. Significance and Use5.1 In order to obtain meaningful analytical data, sample
19、preservation techniques must be effective from the time ofsample collection to the time of analysis. A laboratory mustconfirm that sample integrity is maintained throughout maxi-mum time periods between sample collection and analysis. Inmany cases, it is useful to know the maximum holding time.Aneva
20、luation of holding time is useful also in judging the efficacyof various preservation techniques.6. Reagents6.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on An
21、alytical Reagents of the American Chemical Society,where such specifications are available.4Other grades may beused provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lowering theaccuracy of the determination.6.1.1 Refer to the specific test met
22、hod and to PracticesD3694 for information regarding necessary equipment andpreparation of reagents.6.2 Purity of Water Reference to water shall be under-stood to mean reagent water conforming to SpecificationD1193, Type II, and demonstrated to be free of specificinterference for the test being perfo
23、rmed.7. Determination of Holding Time7.1 Collection of Sample:NOTE 2In some instances, it may be of interest to determine theholding time of standard solutions prepared in water. In such cases, a largevolume of properly preserved, standard solution should be prepared andcarried through the steps of
24、the practice in the same manner as a sample.The volume of solution required can be estimated using the equation in7.1.1.7.1.1 Based on the estimated precision of the test (deter-mined from past experience or from precision data reported inthe test method), calculate the estimated total volume ofsamp
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD484119882013STANDARDPRACTICEFORESTIMATIONOFHOLDINGTIMEFORWATERSAMPLESCONTAININGORGANICANDINORGANICCONSTITUENTS

链接地址:http://www.mydoc123.com/p-518180.html