ASTM D4821-2007 Standard Guide for Carbon Black-Validation of Test Method Precision and Bias《炭黑的标准指南 试验方法精确度和偏差的验证》.pdf
《ASTM D4821-2007 Standard Guide for Carbon Black-Validation of Test Method Precision and Bias《炭黑的标准指南 试验方法精确度和偏差的验证》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4821-2007 Standard Guide for Carbon Black-Validation of Test Method Precision and Bias《炭黑的标准指南 试验方法精确度和偏差的验证》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 4821 07Standard Guide forCarbon BlackValidation of Test Method Precision andBias1This standard is issued under the fixed designation D 4821; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision.
2、A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers a procedure for usingASTM StandardReference Blacks2(SRBs) to continuously monitor the preci-sion of those carbon b
3、lack test methods for which standardvalues have been established. It also offers guidelines fortroubleshooting various test methods.1.2 This guide establishes the x-chart control limits to beused when continuously monitoring those tests listed in Section2. Alternatively, these control limits may be
4、used as a basis forcomparison to testing precision computed within a laboratory.1.3 This guide uses statistical control chart methodology asdiscussed in STP-15-D3to determine if a laboratorys testresults differ significantly from the accepted values of theSRBs.1.4 This guide provides a statistical p
5、rocedure for improv-ing test reproducibility when a laboratory cannot physicallycalibrate its apparatus to obtain the standard values of theASTM SRBs, within the ranges given in the precision state-ment of the test method.1.5 The values stated in SI units are to be regarded as thestandard. The value
6、s given in parentheses are for informationonly.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regu
7、latory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:4D 1510 Test Method for Carbon BlackIodine AdsorptionNumberD 1513 Test Method for Carbon Black, PelletedPourDensityD 1765 Classification System for Carbon Blacks Used inRubber ProductsD 2414 Test Method for Carbon BlackOil Abs
8、orptionNumber (OAN)D 3037 Test Methods for Carbon Black-Surface Area byNitrogen Adsorption5D 3191 Test Methods for Carbon Black in SBR (Styrene-Butadiene Rubber)Recipe and Evaluation ProceduresD 3192 Test Methods for Carbon Black Evaluation in NR(Natural Rubber)D 3265 Test Method for Carbon BlackTin
9、t StrengthD 3493 Test Method for Carbon BlackOil AbsorptionNumber of Compressed Sample (COAN)D 3765 Test Method for Carbon BlackCTAB (Cetyltrim-ethylammonium Bromide) Surface Area5D 4820 Test Methods for Carbon BlackSurface Area byMultipoint B.E.T. Nitrogen Adsorption (Discontinued2000) Replaced by
10、D 65565D 5816 Test Methods for Carbon BlackExternal SurfaceArea by Multipoint Nitrogen Adsorption (Discontinued2000) Replaced by D 65565D 6556 Test Method for Carbon BlackTotal and ExternalSurface Area by Nitrogen Adsorption3. Significance and Use3.1 One of the major causes of poor test precision is
11、 the lackof calibration or standardization of instruments, apparatus,reagents, and technique among laboratories. The sum of allsources of testing error is unique for an individual laboratory.A least-squares regression of a laboratorys actual test valuesfor reference materials to the established mean
12、 values willresult in a unique least-squares regression line (and equation)for that laboratory. Generally, there are two reasons for usingthe SRBs in testing: (1) to monitor testing performance (seeSection 4) to ensure that no systematic error or bias is affecting1This guide is under the jurisdictio
13、n ofASTM Committee D24 on Carbon Blackand is the direct responsibility of Subcommittee D24.61 on Carbon Black Samplingand Statistical Analysis.Current edition approved Oct. 1, 2007. Published November 2007. Originallyapproved in 1988. Last previous edition approved in 2006 as D 4821 06.2Standard Ref
14、erence Blacks are available from Laboratory Standards notto all the tests in Table 1 or Table 2.5.10 When it becomes necessary to use a statistical correc-tion factor, all the SRBs for the range of interest (A-F or A-G)must be used to develop the linear regression equation, even ifsome of the SRBs f
15、all outside the range of interest. At leastfour values (six preferred) should be used for each SRB. Anequal number of results must be used for each SRB. If theseinstructions are not followed, the correction equation will notbe comparable to the corrections made for other instrumentsand other laborat
16、ories. For the oil absorption test (Test MethodD 2414), it is acceptable to develop a separate linear regressionequation for the tread (A-C) and carcass (D-F) SRBs.5.11 Continuously monitor the selected SRBs and use themost current data to construct the correction equation.5.12 Use the correction eq
17、uation to obtain corrected valuesfor the SRBs and plot the corrected values until the controllimit is exceeded. When a test value exceeds the control limit,perform a retest immediately, and if the average of the tworemains outside the control limit, then recalculate the statisticalcorrection equatio
18、n using the most recent test data, excludingthe value that was outside the control limit, for each selectedSRB.5.13 If the corrected data using the new correction equationstill fall outside the control limits, then no further use should bemade of the test method until corrective action (see Section
19、6)brings the SRBs back inside the three standard deviationcontrol chart limits.5.14 Apply this procedure only to the interpretation of SRBtest data. If problems exist, correct them before making furtheruse of the test method. “Local reference” materials do not haveestablished “industry average” valu
20、es to be used in a linearregression equation.5.15 This procedure prescribes only the minimum actionneeded to continuously monitor test precision.Additional SRBtesting may be done and correction equations regularly recal-culated at individual discretion.5.16 Discontinue using a statistical correction
21、 equationwhen new or repaired test equipment is put into service, a newlot of SRBs are purchased and put into use, or other correctiveactions yield uncorrected test results that fall within the controllimits. Resume using a statistical correction equation onlywhen the conditions of Section 4 and 5.1
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD48212007STANDARDGUIDEFORCARBONBLACKVALIDATIONOFTESTMETHODPRECISIONANDBIAS 炭黑 标准 指南 试验 方法 精确度 偏差

链接地址:http://www.mydoc123.com/p-518122.html