ASTM D4754-2018 Standard Test Method for Two-Sided Liquid Extraction of Plastic Materials Using FDA Migration Cell.pdf
《ASTM D4754-2018 Standard Test Method for Two-Sided Liquid Extraction of Plastic Materials Using FDA Migration Cell.pdf》由会员分享,可在线阅读,更多相关《ASTM D4754-2018 Standard Test Method for Two-Sided Liquid Extraction of Plastic Materials Using FDA Migration Cell.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D4754 11D4754 18Standard Test Method forTwo-Sided Liquid Extraction of Plastic Materials Using FDAMigration Cell1This standard is issued under the fixed designation D4754; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision
2、, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the use of the FDA migration cell in the extraction of components and permits q
3、uantitation ofindividual migrants from plastic materials by suitable extracting liquids, including liquid foods and food-stimulating solvents.1.2 This test method provides a two-sided, liquid extraction test for plastic materials that can be formed into film, sheet, or disks.1.3 This test method has
4、 been applied to a variety of migrant/polymer systems in contact with numerous foods and foodsimulants.2 Though most of the migrants examined were radiolabeled, the use of the FDA cell has been validated for migrationstudies of unlabeled sytrene from polystyrene.31.4 This test method has been shown
5、to yield reproducible results under the conditions for migration tests requested by the FDA.However, if the data is to be submitted to the FDA, it is suggested that their guidelines be consulted.1.5 Because it employs two-sided extraction, this test method may not be suitable for multi-layered plast
6、ics intended forsingle-sided food contact use.1.6 The size of the FDA migration cell as described may preclude its use in determining total nonvolatile extractives in somecases.NOTE 1For more information, see Practice D1898, the AOAC Methods of Analysis on Flexible Barrier Materials Exposed for Extr
7、action, and theGuidance for Industry: Preparation of Premarket Submissions for Food Contact Substances: Chemistry Recommendations, December 2007.1.7 Analytical procedures must be available to quantitate the migrant(s) generated by this test method.1.8 The values stated in SI units are to be regarded
8、 as the standard.1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.
9、 Specific hazards statements are given in Section 8.NOTE 2There is no known ISO equivalent to this test method.1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safe
10、ty, health, and environmental practices and determine the applicability ofregulatory limitations prior to use. Specific hazards statements are given in Section 8.NOTE 2There is no known ISO equivalent to this test method.1.10 This international standard was developed in accordance with international
11、ly recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:4D883 Terminolog
12、y Relating to Plastics1 This test method is under the jurisdiction of ASTM Committee D20 on Plastics and is the direct responsibility of Subcommittee D20.70 on Analytical Methods.Current edition approved Dec. 1, 2011May 1, 2018. Published December 2011May 2018. Originally approved in 1987. Last prev
13、ious edition approved in 20032011 asD4754 98D4754 11.(2003). DOI: 10.1520/D4754-11.10.1520/D4754-18.2 “A Study of Indirect Food Additive Migration,” Arthur D. Little, Inc., FDA Contract No. 223-77-2360.3 Supporting data have been filed at ASTM International Headquarters and may be obtained by reques
14、ting Research Report RR:D20-1141.4 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM
15、 standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all
16、 cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1D1898 Pract
17、ice for Sampling of Plastics (Withdrawn 1998)5E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test MethodIEEE/ASTM SI 10 Standard for Use of the International System of Units (SI): The Modernized Metric System2.2 Association of Offcial Analytical Chemists (AOAC)
18、 Methods of Analysis:Flexible Barrier Materials Exposed for Extraction62.3 Federal Document:Guidance for Industry: Preparation of Premarket Submissions for Food Contact Substances: Chemistry Recommendations,December 200773. Terminology3.1 GeneralThe units, symbols, and abbreviations used in this tes
19、t method are in accordance with Terminology D883 andPractice IEEE/ASTM SI 10.4. Summary of Test Method4.1 Specimens of plastic materials, formed in the shape of disks, are threaded onto a stainless steel wire with alternating glassbead spacers and placed in a glass vial. Solvent is added to the vial
20、 and the vial is capped and maintained at the desired extractiontemperature. Aliquots of the liquid are removed at various times and the migrant(s) in the liquid determined by suitable analyticalmethods.NOTE 3 Significant migration loss due to volatility may occur if migration is carried out at temp
21、eratures exceeding 50C for periods greater than 2weeks.5. Significance and Use5.1 Knowledge of migrants from plastic materials may serve many useful purposes, such as testing for compliance with foodadditive regulations. The procedure described in this test method is recommended as suitable for obta
22、ining such data on manymigrant(s)/plastic(s) combinations.6. Apparatus6.1 FDA Migration Cell (Fig. 1), consisting of:6.1.1 Glass Vials, 23-mL,6.1.2 Mininert Slide Valve Caps,6.1.3 Stainless Steel Wire (20-gage), and5 The last approved version of this historical standard is referenced on www.astm.org
23、.6 Available through the Association of Official Analytical Chemists, 481 North Frederick Avenue, Suite 500, Gaithersburg, Maryland 20877-2417 USA.7 Available from Division of Food Contact Notifications, Office of Food Additive Safety, Center for Food Safety and Applied Nutrition, Food and Drug Admi
24、nistration,College Park, MD 20740, USA. https:/www.fda.gov/Food/GuidanceRegulation/GuidanceDocumentsRegulatoryInformation/IngredientsAdditivesGRASPackaging/ucm081818.htmFIG. 1 FDA Migration CellD4754 1826.1.4 Glass Bead (5-mm diameter), containing hole slightly larger than diameter of stainless stee
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD47542018STANDARDTESTMETHODFORTWOSIDEDLIQUIDEXTRACTIONOFPLASTICMATERIALSUSINGFDAMIGRATIONCELLPDF

链接地址:http://www.mydoc123.com/p-517897.html