1、Designation: D 4291 04An American National StandardStandard Test Method forTrace Ethylene Glycol in Used Engine Oil1This standard is issued under the fixed designation D 4291; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of ethyleneglycol as a contaminant in used engine oil. This test meth
3、od isdesigned to quantitate ethylene glycol in the range from 5 to200 mass ppm.1.2 The values stated in SI units are to be regarded as thestandard. The values given in parentheses are for informationonly.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with
4、 its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.For specific warningstatements, see Section 6.NOTE 1A qualitative determination of glycol-base antifreeze is pr
5、o-vided in Test Methods D 2982. ProcedureAis sensitive to about 100 ppm.2. Referenced Documents2.1 ASTM Standards:2D 1193 Specification for Reagent WaterD 2982 Test Methods for Detecting Glycol-Base Antifreezein Used Lubricating OilsD 4057 Practice for Manual Sampling of Petroleum andPetroleum Produ
6、cts3. Summary of Test Method3.1 The sample of oil is extracted with water and theanalysis is performed on the water extract. A reproduciblevolume of the extract is injected into a gas chromatographusing on-column injection and the eluting compounds aredetected by a flame ionization detector. The eth
7、ylene glycolpeak area is determined and compared with areas obtainedfrom the injection of freshly prepared known standards.4. Significance and Use4.1 Leakage of aqueous engine coolant into the crank caseweakens the ability of the oil to lubricate. If ethylene glycol ispresent, it promotes varnish an
8、d deposit formation. This testmethod is designed for early detection to prevent coolant fromaccumulating and seriously damaging the engine.5. Apparatus5.1 Gas ChromatographAny gas chromatographequipped with the following:5.1.1 Flame Ionization Detector, capable of operating con-tinuously at a temper
9、ature equivalent to the maximum columntemperature employed, and connected to the column so as toavoid any cold spots.5.1.2 Sample Inlet System, providing for on-column injec-tion and capable of operating continuously at a temperatureequivalent to the maximum column temperature employed.5.2 RecorderR
10、ecording potentiometer with a full-scaleresponse time of2sorless may be used.5.3 Columns1.2-m (4-ft) by 6.4-mm (14-in.) copper tubepacked with 5 mass % Carbowax 20-M liquid phase on 30/60mesh Chromosorb T solid support. As an alternative, a fusedsilica capillary column, 15 m long with a 0.53mm ID an
11、d2.0micron film thickness of a bonded polyethylene glycol canbe used.5.4 IntegratorManual, mechanical, or electronic integra-tion is required to determine the peak area. However, bestprecision and automated operation can be achieved withelectronic integration.5.5 CentrifugeRCF 600 minimum and centri
12、fuge tubeswith stoppers.5.6 SyringeA microsyringe, 10 L is needed for sampleintroduction.5.7 Pasteur Pipets.5.8 Vials, 2 mL, with crimped septum caps.1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of Subcommittee
13、D02.04 on Hydrocarbon Analysis.Current edition approved Nov. 1, 2004. Published November 2004. Originallyapproved in 1983. Last previous edition approved in 1998 as D 429193(1998).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org
14、. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
15、6. Reagents and Materials6.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents conform to the specifications of the Committee onAnalytical Reagents of the American Chemical Society wheresuch specifications are available.3
16、Other grades may be used,provided it is first ascertained that the reagent is of sufficientlyhigh purity to permit its use without lessening the accuracy ofthe determination.6.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean reagent water as definedby Type
17、II of Specification D 1193.6.3 Air and Hydrogen(WarningThe air supply may befrom a cylinder under high pressure. Hydrogen is an extremelyflammable gas under pressure.)6.4 Calibration MixturesA minimum of three mixtures ofwater and ethylene glycol are prepared to cover the range from5 to 200 mass ppm
18、. Prepare one blend of approximately 2000mass ppm ethylene glycol in water to provide for accurateweighing; then, prepare dilutions of that solution.6.5 Carrier Gas, helium or nitrogen may be used with theflame ionization detector. (WarningHelium and nitrogen arecompressed gases under high pressure.
19、)6.6 Ethylene Glycol, 99 mass % pure.6.7 n-Hexane, 99 mol % pure. (Warningn-Hexane isextremely flammable, harmful if inhaled, may produce nervecell damage.)6.8 Liquid Phase and Solid Support, 5 mass % Carbowax20-M liquid phase on 30/60 mesh Chromosorb T solid support.6.9 Tubing, 6.4 mm (14 in.) in o
20、utside diameter, 1.2 m (4 ft)long of copper.6.10 Water, deionized or distilled.7. Preparation of Apparatus7.1 Packed Column PreparationIf a packed column isused, prepare it using the following steps:7.1.1 Prepare the packing, 5 mass % Carbowax 20-M liquidphase on 30/60 mesh Chromosorb T solid suppor
21、t, by anysatisfactory method used in the practice of gas chromatogra-phy.NOTE 2Care should be taken in handling Chromosorb T solid supportbecause of its static charge and softness. Chilling may be helpful inimproving its handling properties.7.1.2 Add the prepared packing to the copper tubing usingon
22、ly gentle tapping. Do not use vacuum or mechanical vibra-tion to pack the column. Chromosorb T solid support is a resinwhich will deform under pressure or severe vibration.7.2 Column InstallationThe column must be attached tothe injection port in such a way as to allow on-columninjection.7.3 Column
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD42912004STANDARDTESTMETHODFORTRACEETHYLENEGLYCOLINUSEDENGINEOIL 发动 机油 微量 乙二醇 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-516725.html