《ASTM D4230-2002(2007) Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4230-2002(2007) Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D4230 02 (Reapproved 2007)Standard Test Method ofMeasuring Humidity with Cooled-Surface Condensation(Dew-Point) Hygrometer1This standard is issued under the fixed designation D4230; the number immediately following the designation indicates the year oforiginal adoption or, in the case o
2、f revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of the ther-modynamic dew- or frost-point temperatu
3、re of ambient air bythe condensation of water vapor on a cooled surface. Forbrevity this is referred to in this method as the condensationtemperature.1.2 This test method is applicable for the range of conden-sation temperatures from 60C to 70C.1.3 This test method includes a general description of
4、theinstrumentation and operational procedures, including siteselection, to be used for obtaining the measurements and adescription of the procedures to be used for calculating theresults.1.4 This test method is applicable for the continuous mea-surement of ambient humidity in the natural atmosphere
5、on astationary platform.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prio
6、r to use. For specificprecautionary statements, see Section 8.2. Referenced Documents2.1 ASTM Standards:2D 1356 Terminology Relating to Sampling and Analysis ofAtmospheresD 3631 Test Methods for Measuring Surface AtmosphericPressure3. Terminology3.1 Definitions:3.1.1 For definitions of other terms i
7、n this method, refer toTerminology D 1356.3.2 Definitions of Terms Specific to This Standard:3.2.1 nonhygroscopic materialmaterial that neither ab-sorbs nor retains water vapor.3.2.2 mirror (front surface)a polished surface, usually ametallic surface, on which condensates are deposited.3.3 Symbols:e
8、 = vapor pressure of water vapor in moist air.ei= saturation pressure of water vapor in equilibriumwith the plane surface of ice.ew= saturation pressure of water vapor in equilibriumwith the plane surface of water.P = ambient pressure.r = mixing ratio.T = ambient air temperature.Td= thermodynamic de
9、w- or frost-point temperature.RHi= relative humidity with respect to ice.RHw= relative humidity with respect to water.4. Summary of Test Method4.1 The ambient humidity is measured with a dew- andfrost-point hygrometer.4.2 The mirror or some other surface on which the conden-sate is deposited is prov
10、ided with the means for cooling andheating, detection of condensate, and the measurement of thetemperature of the mirror surface.4.3 Calculations of saturation vapor pressure over water andice as functions of temperature are provided.5. Significance and Use5.1 Humidity information is important for t
11、he understandingof atmospheric phenomena and industrial processes. Measure-ments of the dew-point and calculations of related vaporpressures are important to quantify the humidity information.1This test method is under the jurisdiction of ASTM Committee D22 on AirQuality and is the direct responsibi
12、lity of Subcommittee D22.11 on Meteorology.Current edition approved Oct. 1, 2007. Published December 2007. Originallyapproved in 1983. Last previous edition approved in 2002 as D4230 - 02e1.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at servi
13、ceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.Copyright by ASTM Intl (all rights reserved); Fri Dec 26
14、 01:13:18 EST 2008Downloaded/printed byGuo Dehua (CNIS) pursuant to License Agreement. No further reproductions authorized.6. Interferences6.1 This method is not applicable if other constituents in theatmosphere condense before water vapor.7. Apparatus7.1 Dew-point hygrometers, specifically designed
15、 for me-teorological observations are available commercially. A sche-matic arrangement of a typical optical dew-point hygrometer isshown in Fig. 1.7.1.1 The sample air flows through a small chamber.7.1.2 Within the chamber is a mirror or surface on which thecondensate can be deposited.7.1.3 A beam o
16、f light from an incandescent lamp, lightemitting diode or other suitable light source shines on themirror.7.1.4 Dew or frost is detected with an electro-optic device.7.1.5 The mirror is cooled by a Peltier thermoelectricelement. Peltier cooling is a convenient method for unattendedand automatic inst
17、ruments.7.1.6 Preferred devices of sensing mirror temperature areresistance thermometers, thermistors, and thermocouples.7.1.6.1 The temperature sensors shall be attached to orembedded in the mirror to measure the temperature of thesurface of the mirror.7.1.7 Suitable control circuitry shall be prov
18、ided to maintaina constant quantity of condensate on the mirror.7.1.8 Suitable provisions shall be provided to compensatefor the contamination of the surface of the mirror.7.2 Auxiliary Equipment:7.2.1 Provision shall be provided for assuring air flow pastthe dewpoint mirror without changing the pre
19、ssure in themirror chamber more than 0.5 % from the ambient pressuresurrounding the sensor. An air flow of approximately 1.1 litresper minute is recommended for typical chambers.7.2.2 Readout instrumentation is available with the dew-point hygrometer.8. Precautions8.1 Safety Precautions:8.1.1 The hy
20、grometer shall be packaged in a suitableenclosure for application in industrial or outdoor environment.8.1.2 Electrical connectors and cables shall be suitable forthe industrial or outdoor environment.8.1.3 Appropriate voltage surge protection circuitry must beincorporated.8.2 Technical Precautions:
21、8.2.1 The accuracy of a cooled-surface condensation hy-grometer is degraded by the presence of water-soluble materi-als. A mirror-cleaning schedule, consistent with the contami-nation rate, is necessary to maintain the initial calibrationaccuracy. The user must determine the required maintenancesche
22、dule for the specific site, by comparison of calibrationsmade before and after cleaning.8.2.2 Caution in performing this method should be taken ifthe indicated mirror temperature is between 0C and 30C.Below freezing, the initial formation of the condensate on thesurface of a mirror may be either dew
23、 or frost. In the case ofnonfiltered atmospheric air, the supercooled water usually doesnot persist long on a mirror surface and quickly changes tofrost. The only positive method for determining the state of thecondensate is by visual observation of the mirror surface.8.2.2.1 The following illustrat
24、es the magnitude of the errorinvolved when dew or frost is not differentiated: The saturationvapor pressure of supercooled water at 30C corresponds tosaturation vapor pressure of ice at 27.2C; dew pointof 20C corresponds to frost point of 18.0C; 10C dewpoint corresponds to frost point of 8.9C. (The
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD423020022007STANDARDTESTMETHODOFMEASURINGHUMIDITYWITHCOOLEDSURFACECONDENSATIONDEWPOINTHYGROMETER

链接地址:http://www.mydoc123.com/p-516593.html