ASTM D3760-2002e1 Standard Test Method for Analysis of Isopropylbenzene (Cumene) by Gas Chromatography《用气相色谱法分析异丙苯的标准试验方法》.pdf
《ASTM D3760-2002e1 Standard Test Method for Analysis of Isopropylbenzene (Cumene) by Gas Chromatography《用气相色谱法分析异丙苯的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D3760-2002e1 Standard Test Method for Analysis of Isopropylbenzene (Cumene) by Gas Chromatography《用气相色谱法分析异丙苯的标准试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 3760 02e1Standard Test Method forAnalysis of Isopropylbenzene (Cumene) by GasChromatography1This standard is issued under the fixed designation D 3760; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.e1NOTEResearch report was added editorially in September 2004.1. Scope1.1 This test method covers the determination of the purityof i
3、sopropylbenzene (cumene) by gas chromatography.1.2 This test method has been found applicable to themeasurement of impurities such as nonaromatic hydrocarbons,benzene, ethylbenzene, t-butylbenzene, n-propylbenzene,alpha-methylstyrene, sec-butylbenzene, and diisopropylben-zene, which are common to th
4、e manufacturing process ofisopropylbenzene. Limit of detection for these impurities is 10mg/kg (see 5.1).1.3 The following applies to all specified limits in thisstandard: for purposes of determining conformance with thisstandard, an observed value or a calculated value shall berounded off “to the n
5、earest unit” in the last right-hand digitused in expressing the specification limit, in accordance withthe rounding-off method of Practice E 29.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standa
6、rd to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific hazardstatements, see Section 7.2. Referenced Documents2.1 ASTM Standards:2D 3437 Practice for Sampling and Handling Liquid CyclicProductsE 29 Practice for U
7、sing Significant Digits in Test Data toDetermine Conformance with SpecificationsE 260 Practice for Packed Gas ChromatographyE 355 Practice for Gas Chromatography Terms and Rela-tionships2.2 Other Document:OSHA Regulations, 29CFR, paragraphs 1910.1000 and1910.120033. Summary of Test Method3.1 A known
8、 amount of internal standard is added to asample of isopropylbenzene. The prepared sample is mixedand analyzed by a gas chromatograph equipped with a flameionization detector (FID). The peak area of each impurity andthe internal standard is measured and the amount of eachimpurity is calculated from
9、the ratio of the peak area of theinternal standard versus the peak area of the impurity. Purity byGC (the isopropylbenzene content) is calculated by subtractingthe sum of the impurities found from 100.00. Results arereported in weight percent.4. Significance and Use4.1 This test method is suitable f
10、or setting specifications onthe materials referenced in 1.2 and for use as an internal qualitycontrol tool where isopropylbenzene is produced or is used ina manufacturing process. It may also be used in developmentor research work involving isopropylbenzene.4.2 This test method is useful in determin
11、ing the purity ofisopropylbenzene with normal impurities present includingdiisopropylbenzenes. If extremely high boiling or unusualimpurities are present in the isopropylbenzene, this test methodwould not necessarily detect them and the purity calculationwould be erroneous.4.3 Cumene hydroperoxide,
12、if present, will yield decompo-sition products that will elute in the chromatogram therebygiving incorrect results.4.4 The nonaromatic hydrocarbons commonly present fromthe isopropylbenzene manufacturing process will interfere withthe determination of benzene when Column A in Table 1 isused.1This te
13、st method is under the jurisdiction of ASTM Committee D16 onAromatic Hydrocarbons and Related Chemicals and is the direct responsibility ofSubcommittee D16.07 on Styrene, Ethylbenzene, and C9and C10Aromatic Hydro-carbons.Current edition approved Jan. 10, 2002. Published March 2002. Originallypublish
14、ed as D 3760 79. Last previous edition D 3760 98.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available f
15、rom U.S. Government Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.4.5 The internal standard must be sufficiently resolved from
16、any impurity and the isopropylbenzene peak.5. Apparatus5.1 Gas ChromatographAny instrument having a flameionization detector that can be operated at the conditions givenin Table 1. The system should have sufficient sensitivity toobtain a minimum peak height response for 10 mg/kgn-butylbenzene of twi
17、ce the height of the signal backgroundnoise.5.2 ColumnsThe choice of column is based on resolutionrequirements. Any column may be used that is capable ofresolving all significant impurities from isopropylbenzene andfrom the internal standard. The columns described in Table 1have been used successful
18、ly.5.3 RecorderElectronic integration is recommended.6. Reagents and Materials6.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the Ameri
19、can Chemical Society,where such specifications are available.4Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.6.1.1 Internal StandardNormal Butylbenzene (nBB) isthe recommend
20、ed internal standard of choice. Other com-pounds may be found acceptable provided they meet thecriteria as defined in 4.5 and 6.1.6.2 Carrier GasChromatographic grade helium or hydro-gen is recommended.6.3 Compressed AirChromatographic grade.6.4 HydrogenHigh purity.7. Hazards7.1 Consult current OSHA
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD37602002E1STANDARDTESTMETHODFORANALYSISOFISOPROPYLBENZENECUMENEBYGASCHROMATOGRAPHY 用气相 色谱 分析 异丙苯

链接地址:http://www.mydoc123.com/p-515424.html