ASTM D2503-1992(2002)e1 Standard Test Method for Relative Molecular Mass (Molecular Weight) of Hydrocarbons by Thermoelectric Measurement of Vapor Pressure《蒸气压力热电测量法测定烃类相对分子质量(分子量).pdf
《ASTM D2503-1992(2002)e1 Standard Test Method for Relative Molecular Mass (Molecular Weight) of Hydrocarbons by Thermoelectric Measurement of Vapor Pressure《蒸气压力热电测量法测定烃类相对分子质量(分子量).pdf》由会员分享,可在线阅读,更多相关《ASTM D2503-1992(2002)e1 Standard Test Method for Relative Molecular Mass (Molecular Weight) of Hydrocarbons by Thermoelectric Measurement of Vapor Pressure《蒸气压力热电测量法测定烃类相对分子质量(分子量).pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 2503 92 (Reapproved 2002)e1An American National StandardStandard Test Method forRelative Molecular Mass (Molecular Weight) ofHydrocarbons by Thermoelectric Measurement of VaporPressure1This standard is issued under the fixed designation D 2503; the number immediately following the des
2、ignation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use
3、by agencies of the Department of Defense.e1NOTEWarnings were moved from notes to section text editorially December 2002.1. Scope1.1 This test method covers the determination of the averagerelative molecular mass (molecular weight) of hydrocarbonoils. It can be applied to petroleum fractions with mol
4、ecularweights (relative molecular mass) up to 3000; however, theprecision of the method has not been established above 800molecular weight (relative molecular mass). The methodshould not be applied to oils having initial boiling points lowerthan 220C.1.2 Values stated in SI units are to be regarded
5、as thestandard. The values given in parentheses are provided forinformation purposes only.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practic
6、es and determine the applica-bility of regulatory limitations prior to use. For specific hazardstatements, 5.2.1, 5.2.3, and 5.2.3.2. Summary of Test Method2.1 Aweighed portion of the sample is dissolved in a knownquantity of appropriate solvent. A drop of this solution and adrop of solvent are susp
7、ended, side by side, on separatethermistors in a closed chamber saturated with solvent vapor.Since the vapor pressure of the solution is lower than that of thesolvent, solvent condenses on the sample drop and causes atemperature difference between the two drops. The resultantchange in temperature is
8、 measured and used to determine therelative molecular mass (molecular weight) of the sample byreference to a previously prepared calibration curve.3. Significance and Use3.1 Relative molecular mass (molecular weight) is a funda-mental physical constant that can be used in conjunction withother physi
9、cal properties to characterize pure hydrocarbons andtheir mixtures.3.2 A knowledge of the relative molecular mass (molecularweight) is required for the application of a number of correla-tive methods that are useful in determining the gross compo-sition of the heavier fractions of petroleum.4. Appar
10、atus4.1 Vapor Pressure Osmometer, with operating diagram.25. Reagents and Materials5.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the
11、American Chemical Society,where such specifications are available.3Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.5.2 SolventsSolvents that do not react with the samplemust
12、be used. Since many organic materials exhibit a tendencyto associate or dissociate in solution, it is desirable to use polarsolvents for polar samples and nonpolar solvents for nonpolar1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the di
13、rect responsibility of SubcommitteeD02.04 on Hydrocarbon Analysis.Current edition approved Aug. 15, 1992. Published October 1992. Originallypublished as D 2503 66T. Last previous edition D 2503 82 (1987)e1.2A vapor pressure osmometer is available from H. Knauer and Co., Berlin, WestGermany. The manu
14、facture of the Mechrolab instrument previously referred to inthis footnote has been discontinued. However, some models may be available fromstocks on hand at laboratory supply houses, or as used equipment from laboratoryinstrument exchanges.3“Reagent Chemicals, American Chemical Society Specificatio
15、ns,” Am. Chemi-cal Soc., Washington, DC. For suggestions on the testing of reagents not listed bythe American Chemical Society, see “Analar Standards for Laboratory U.K.Chemicals,” BDH Ltd., Poole, Dorset, and the “United States Pharmacopeia.”1Copyright ASTM International, 100 Barr Harbor Drive, PO
16、Box C700, West Conshohocken, PA 19428-2959, United States.samples. The solvents listed have been found suitable forhydrocarbons and petroleum fractions.5.2.1 Benzene (WarningPoison. Carcinogen. Harmful ifswallowed. Extremely flammable. Vapors may cause flash fire.Vapor harmful, may be absorbed throu
17、gh skin.)5.2.2 Chloroform (WarningMay be fatal if swallowed.Harmful if inhaled. May produce toxic vapors if burned.)5.2.3 1,1,1-Trichloroethane (WarningHarmful if inhaled.High concentrations may cause unconsciousness or death.Contact may cause skin irritation and dermatitis.)NOTE 1The precision data
18、 given in 10.1 and 10.2 will apply onlywhen benzene is used as the solvent. There is also some evidence thatdeterminations on the same oil sample carried out in different solvents willproduce results that differ somewhat in absolute magnitude of apparentmolecular weight (relative molecular mass).5.3
19、 Reference StandardsA calibration curve must be con-structed for each new lot of solvent using a pure compoundwhose relative molecular mass (molecular weight) is accu-rately known. Compounds that have been used successfullyinclude benzil (210.2), n-octadecane (254.5), and squalane(422.8).6. Sampling
20、6.1 The sample must be completely soluble in the selectedsolvent at concentrations of at least 0.10 M, and it must nothave an appreciable vapor pressure at the test temperature.7. Preparation and Calibration of Apparatus7.1 Prepare standard 0.01, 0.02, 0.04, 0.06, 0.08, and 0.1 Msolutions of the cal
21、ibrating compound in the solvent selected.7.2 Remove the upper sample chamber assembly. Rinse thesolvent cap with the solvent to be used. Install a vapor wick inthe cup and fill with solvent to the bottom of the notches in theinner wick. Place the cup in the chamber base recess, align thevapor wick
22、openings with the viewing tubes, and replace theupper assembly. Take care that the guide pins properly engagematching holes in the thermal block and that the matchingsurfaces of the base and block are clean. Be careful not to allowthe thermistor beads to touch the cup or wicks as they may bebent out
23、 of alignment. Turn on the thermostat and allow thetemperature of the sample chamber to reach equilibrium at37C.NOTE 2If the block is at room temperature, 2 to 3 h will be required.To avoid such delay, it is desirable to always leave the thermostat switchin the “on” position. If the chamber is at eq
24、uilibrium and is opened briefly,30 to 45 min will generally be required before temperature stabilization isregained. The exchange or refilling of syringes does not necessitate anywaiting period.7.3 Thoroughly rinse all syringes with the solvent beingused and allow to dry.7.4 Fill the syringes from g
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD250319922002E1STANDARDTESTMETHODFORRELATIVEMOLECULARMASSMOLECULARWEIGHTOFHYDROCARBONSBYTHERMOELECTRICMEASUREMENTOFVAPORPRESSURE

链接地址:http://www.mydoc123.com/p-512980.html