《ASTM D2162-2017 Standard Practice for Basic Calibration of Master Viscometers and Viscosity Oil Standards《主粘度计和标准粘度油基本校准的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM D2162-2017 Standard Practice for Basic Calibration of Master Viscometers and Viscosity Oil Standards《主粘度计和标准粘度油基本校准的标准实施规程》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D2162 14D2162 17Standard Practice forBasic Calibration of Master Viscometers and Viscosity OilStandards1This standard is issued under the fixed designation D2162; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This practice covers the calibration of master viscometers and viscosity oil standards, both of which may be use
3、d to calibrateroutine viscometers as described in Test Method D445 and Specifications D446 over the temperature range from 15 C to 100 C.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.2.1 The SI-based units for calibrat
4、ion constants and kinematic viscosities are mm2/s2 and mm2/s, respectively.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determin
5、e the applicability of regulatorylimitations prior to use. For specific warning statements, see Section 7.1.4 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of Internat
6、ional Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D445 Test Method for Kinematic Viscosity of Transparent and Opaque Liquids (and Calculation of Dynamic Viscosity)D446 Specification
7、s and Operating Instructions for Glass Capillary Kinematic ViscometersD1193 Specification for Reagent WaterD1250 Guide for Use of the Petroleum Measurement TablesD1480 Test Method for Density and Relative Density (Specific Gravity) of Viscous Materials by Bingham PycnometerD1590 Test Method for Surf
8、ace Tension of WaterE1 Specification for ASTM Liquid-in-Glass ThermometersE563 Practice for Preparation and Use of an Ice-Point Bath as a Reference TemperatureE644 Test Methods for Testing Industrial Resistance ThermometersE1137 Specification for Industrial Platinum Resistance ThermometersE1750 Guid
9、e for Use of Water Triple Point CellsE2593 Guide for Accuracy Verification of Industrial Platinum Resistance ThermometersE2877 Guide for Digital Contact Thermometers2.2 ISO Standard:3ISO 3666 Viscosity of Water3. Terminology3.1 Definitions:3.1.1 digital contact thermometer (DCT) , nan electronic dev
10、ice consisting of temperature measuring sensor in contact withthe material that provides an output to a digital display of the measured value.and associated temperature sensing probe.1 This practice is under the jurisdiction of ASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants a
11、nd is the direct responsibility of SubcommitteeD02.07 on Flow Properties.Current edition approved June 1, 2014May 1, 2017. Published July 2014May 2017. Originally approved in 1963. Last previous edition approved in 20132014 asD2162 13.D2162 14. DOI: 10.1520/D2162-14.10.1520/D2162-17.2 For referenced
12、ASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 Available from American National Standards Institute (ANSI), 25 W. 43rd St.
13、, 4th Floor, New York, NY 10036.10036, http:/www.ansi.org.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all change
14、s accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr
15、Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.1.1.1 DiscussionThis device consists of a temperature sensor connected to a measuring instrument; this instrument measures the temperature-dependent quantity of the sensor, computes the temperature from the measured quantit
16、y, and provides a digital output. This digitaloutput goes to a digital display and/or recording device that may be internal or external to the device. These devices are sometimesreferred to as a “digital thermometer”.thermometers.”NOTE 1Portable electronic thermometers (PET) is an acronym sometimes
17、used to refer to a subset of the devices covered by this definition.3.1.1.2 DiscussionPET is an acronym for portable electronic thermometers, a subset of digital contact thermometers (DCT).3.2 Definitions of Terms Specific to This Standard:3.2.1 basic calibration, ncalibration based on the primary s
18、tandard, water.3.2.1.1 DiscussionPure water has a kinematic viscosity of 1.0034 mm2/s at 20 C. See ISO 3666.3.2.2 master viscometer, nglass capillary viscometer with a liquid driving head of at least 400 mm.3.2.2.1 DiscussionIt is specially designed to minimize errors due to surface tension, kinetic
19、 energy, and capillary end effects.3.2.3 viscosity oil standard, nstable Newtonian liquid, the kinematic viscosity of which has been related to the kinematicviscosity of water through the step-up procedure described in this practice.4. Summary of Practice4.1 Two or more master viscometers, having ca
20、libration constants in the 0.001 mm2/s2 to 0.003 mm2/s2 range, are calibratedwith water at 20 C. The kinematic viscosities of two or more oil standards are measured at 40 C in these two master viscometers.Corrections are made for buoyancy and, where necessary, for temperature and surface tension.4.2
21、 Athird master viscometer, with a calibration constant of 0.003 mm2/s2 to 0.009 mm2/s2 , is then calibrated at 40 C with thetwo standard oils and its calibration factor calculated at standard conditions for water at 20 C. In like manner additional viscosityoil standards and additional master viscome
22、ters are calibrated at 40 C using the average results from at least two masterviscometers or two oil standards. Steps between successive calibration constants or viscosities increase by a factor of three or lessuntil the desired viscosity range is covered.4.3 Oils are calibrated at other temperature
23、s using the average result from at least two master viscometers.5. Significance and Use5.1 Because there are surface tension or kinematic viscosity differences, or both, between the primary standard (7.4) andkinematic viscosity standards (7.5), special procedures using master viscometers are require
24、d to “step-up” from the kinematicviscosity of the primary standard to the kinematic viscosities of oil standards.5.2 Using master viscometers calibrated according to this practice, an operator can calibrate kinematic viscometers inaccordance with Specifications D446.5.3 Using viscosity oil standards
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD21622017STANDARDPRACTICEFORBASICCALIBRATIONOFMASTERVISCOMETERSANDVISCOSITYOILSTANDARDS 粘度计 标准 粘度

链接地址:http://www.mydoc123.com/p-511116.html