ASTM D1005-1995(2007) Standard Test Method for Measurement of Dry-Film Thickness of Organic Coatings Using Micrometers《用千分尺测量有机涂层干膜厚度的标准试验方法》.pdf
《ASTM D1005-1995(2007) Standard Test Method for Measurement of Dry-Film Thickness of Organic Coatings Using Micrometers《用千分尺测量有机涂层干膜厚度的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D1005-1995(2007) Standard Test Method for Measurement of Dry-Film Thickness of Organic Coatings Using Micrometers《用千分尺测量有机涂层干膜厚度的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 1005 95 (Reapproved 2007)Standard Test Method forMeasurement of Dry-Film Thickness of Organic CoatingsUsing Micrometers1This standard is issued under the fixed designation D 1005; the number immediately following the designation indicates the year oforiginal adoption or, in the case o
2、f revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 This test me
3、thod covers the measurement of film thick-ness of dried films of paint, varnish, lacquer, and relatedproducts using micrometers. Procedures A and B utilize sta-tionary micrometers and Procedures C and D, hand-heldmicrometers. Procedures A and C are not recommended forfilms less than 12.5 m (0.5 mils
4、) in thickness. The minimumthickness required for Procedures B and D is a function of thatrequired to enable removal of the sample as a free film.1.2 The procedures appear as follows:1.2.1 Procedure AStationary micrometer for measuringcoatings applied to plane rigid surfaces.1.2.2 Procedure BStation
5、ary micrometer for measuringfree films.1.2.3 Procedure CHand-held micrometer for measuringcoatings applied to plane rigid surfaces.1.2.4 Procedure DHand-held micrometer for measuringfree films.1.3 The values stated in SI units are to be regarded as thestandard. The values given in parentheses are fo
6、r informationonly.1.4 This standard does not purport to address the safetyconcerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety andhealth practices and determine the applicability of regulatorylimitations prior to use.2. Refer
7、enced Documents2.1 ASTM Standards:2D 823 Practices for Producing Films of Uniform Thicknessof Paint, Varnish, and Related Products on Test PanelsD 2370 Test Method for Tensile Properties of OrganicCoatings3. Significance and Use3.1 This test method is particularly applicable to the mea-surement of f
8、ree films and is also satisfactory for the measure-ment of films on laboratory test panels.3.2 The accuracy and precision of the thickness measure-ments may be influenced by the deformability of the coating.This test method is not applicable to coatings that are readilydeformable under the load of t
9、he measuring instrument.3.3 The accuracy and precision of the thickness measure-ments are also influenced by the uniformity of the substratewhen the coatings are applied to laboratory test panels.4. Apparatus4.1 Procedures A and B:4.1.1 The apparatus shall consist of a dial comparator, dialindicator
10、, or micrometer. A rigid base is required for mountingthe dial comparator or dial indicator gages. The presser foot ofthe micrometer or dial indicator shall be circular, from 1.5 to3.0 mm (116 to18 in.) in diameter, and shall be flat on thebottom. The presser foot shall be fixed to an indicator that
11、 readsto 2.5 m (0.1 mil). The load on the presser foot shall bebetween 140 and 275 kPa (20 and 40 psi ). For Procedure B, asmooth uncoated test plate is also required.4.1.2 Verify the accuracy of instrument calibration by set-ting to zero with the anvils closed followed by measuringshims of known th
12、icknesses or standards specifically designedfor this purpose. Record the standard thickness gage measure-ment and the micrometer reading. Use these results to constructa calibration curve.4.2 Procedures C and DThe apparatus shall consist of ahand-held micrometer. The anvils of the micrometer shall b
13、ecircular, from 1.5 to 3.0 mm (116 to18 in.) in diameter, withflat bottoms. Verify the accuracy of instrument calibration bysetting to zero with the anvils closed followed by measuring1This test method is under the jurisdiction of ASTM Committee D01 on Paintand Related Coatings, Materials, and Appli
14、cations and is the direct responsibility ofSubcommittee D01.23 on Physical Properties of Applied Paint Films.Current edition approved Nov. 1, 2007. Published November 2007. Originallyapproved in 1949. Last previous edition approved in 2001 as D 1005 95 (2001).2For referenced ASTM standards, visit th
15、e ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
16、United States.shims of known thicknesses or standards specifically designedfor this purpose. Record the standard thickness gage measure-ment and the micrometer reading. Use these results to constructa calibration curve.5. Test Specimen5.1 Procedures A and CApply test films to a suitableplane, rigid
17、base material from which the dried film may besatisfactorily removed. The panels shall be of sufficient size topermit film thickness measurements to be made 25 mm (1 in.)from any edge.5.1.1 Coatings should be applied in accordance with Prac-tices D 823 or as agreed upon between the purchaser and the
18、seller.5.2 Procedures B and DFree films of the test material arerequired. Alternatively, the test materials can be applied to anappropriate substrate in order that they can be removed as freefilms without deformation. If the method of specimen prepa-ration affects the film forming properties of the
19、test material orrequires cutting or scraping to remove the free film, useProcedures A or C instead.5.2.1 Free films may be prepared in accordance with TestMethod D 2370.6. Procedure6.1 Procedure A:6.1.1 Mount the test panel rigidly on a suitable base. Clampor hold it to the base in such a way that t
20、here will be nomovement or spring of the panel during the film thicknessmeasurement.6.1.2 Close the gage slowly until contact is made, butwithout visible distortion of the film. Read the gage, estimatingto 2.5 m (0.1 mil), and record the reading.6.1.3 Open the gage and remove the film carefully from
21、 thearea where the measurement was taken. Any suitable means,chemical or mechanical, may be used to remove the film,taking care not to distort the panel. Close the gage, slowly onthe area from which the film was removed, and take a reading,estimating to 2.5 m (0.1 mil).6.1.4 The difference in the ga
22、ge readings before and afterthe removal of the film is the thickness of the film. Record to2.5 m (0.1 mil).6.1.5 As an alternative to the above, the panels may bemeasured in distinct locations prior to the application of thecoating, and again in the identical locations after the coatinghas cured. Th
23、e difference represents the coating thickness.6.1.6 Take a sufficient number of readings to characterizethe panel. A recommended minimum is three determinationsfor a 75 by 150-mm (3 by 6-in.) panel and more in proportionto size.6.2 Procedure B:6.2.1 Mount a smooth uncoated test panel rigidly on asui
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD100519952007STANDARDTESTMETHODFORMEASUREMENTOFDRYFILMTHICKNESSOFORGANICCOATINGSUSINGMICROMETERS

链接地址:http://www.mydoc123.com/p-509117.html