ASTM C1831 C1831M-2017 Standard Guide for Gamma Radiation Shielding Performance Testing《伽马辐射屏蔽性能试验的标准指南》.pdf
《ASTM C1831 C1831M-2017 Standard Guide for Gamma Radiation Shielding Performance Testing《伽马辐射屏蔽性能试验的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM C1831 C1831M-2017 Standard Guide for Gamma Radiation Shielding Performance Testing《伽马辐射屏蔽性能试验的标准指南》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C1831/C1831M 17Standard Guide forGamma Radiation Shielding Performance Testing1This standard is issued under the fixed designation C1831/C1831M; the number immediately following the designation indicates theyear of original adoption or, in the case of revision, the year of last revision
2、. A number in parentheses indicates the year of lastreapproval. A superscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide identifies appropriate test methods for deter-mining the sufficiency of radiological shielding for hot cellsand shielded
3、 enclosures.1.2 After constructing or modifying radiological shielding,it is necessary to verify that shielding performance meets orexceeds the shielding performance requirements. This is typi-cally accomplished using sealed test sources of much lessactivity than the design basis. This allows for mo
4、difications orcorrection of any discrepancies identified before the commis-sioning of the hot cell.1.3 The guidance and practices recommended by this guideare applicable to both new and existing shielded facilities andenclosures for evaluating shielding suitability and locating theexistence of shine
5、 paths or other shielding anomalies that resultfrom design, manufacture, or construction.1.4 Two types of testing may be performed.1.4.1 Shielding performance verification testing providesevidence that the shielding configuration is sufficient formeeting established performance criteria and for iden
6、tifyingdeficiencies in the shielding configuration or components thatmay not have been addressed during design. Test results areexpected to demonstrate that shielding performance meets orexceeds design criteria, not match the dose rates predictedanalytically.1.4.2 Shielding performance verification
7、testing identifiesshielding deficiencies (hot spots) in the installed configurationrelative to adjacent shielding but does not demonstrate com-pliance with any quantitative shielding performance require-ment.1.5 Performance testing should be specified and performedto assess shielding adequacy with s
8、ources in all criticallocations.1.6 Requirements for shielding performance testing shouldbe clearly defined in design basis or procurement documenta-tion.1.7 This guide is not applicable to neutron radiation shield-ing performance evaluations.1.8 UnitsThe values stated in either SI units or inch-pou
9、nd units are to be regarded separately as standard. Thevalues stated in each system may not be exact equivalents;therefore, each system shall be used independently of the other.Combining values from the two systems may result in noncon-formance with the standard.1.8.1 Units for total activity should
10、 be given in Becquerel(Bq) or curies (Ci).1.8.2 Units for dose rate as measured during testing shouldbe given in Sieverts (Sv/h) or rad/h.1.8.3 Distances and locations should be provided in centi-metres or inches.1.9 This standard does not purport to address all of thesafety concerns, if any, associ
11、ated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations prior to use.1.10 This international standard was developed in accor-dance with internationally rec
12、ognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2C859 Terminology Re
13、lating to Nuclear Materials3. Terminology3.1 For terms not defined in this guide, see TerminologyC859.3.2 Definitions:3.2.1 angular response sensitivity, nthe ability of aninstrument or detector to detect radiation based on the angle ofincidence of the radiation on the instrument or detector.1This g
14、uide is under the jurisdiction of ASTM Committee C26 on Nuclear FuelCycle and is the direct responsibility of Subcommittee C26.14 on Remote Systems.Current edition approved Nov. 1, 2017. Published November 2017. DOI:10.1520/C1831_C1831M-17.2For referenced ASTM standards, visit the ASTM website, www.
15、astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis int
16、ernational standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) C
17、ommittee.13.2.2 Bremsstrahlung radiation, nelectromagnetic radia-tion produced by the deceleration of a charged particle whendeflected by another charged particle, typically an electron byan atomic nucleus.3.2.2.1 DiscussionThe moving particle loses kineticenergy, which is converted into a photon be
18、cause energy isconserved.3.2.3 buildup factor, nfor radiation passing through amedium, the buildup factor is the ratio of the total value of aspecific radiation quantity (direct and scattered) measured asabsorbed dose at any point within that medium to the contri-bution to that quantity from the inc
19、ident uncollided radiationreaching that point.3.2.3.1 DiscussionThe buildup factor increases with in-creased shielding thickness and is higher for low atomicnumber materials.3.2.4 encapsulated source, nsealed source strong enoughto maintain leak tightness under the conditions of use for whichthe sou
20、rce was designed and also under foreseeable mishaps.3.2.4.1 DiscussionIt is radioactive material that is perma-nently sealed in a capsule or closely bonded and in a solidform.3.2.5 Geiger-Mller counter, GM, ntype of particle detec-tor that measures ionizing radiation.3.2.5.1 DiscussionThe radiation-
21、sensing element is aninert gas-filled Geiger-Mller tube (usually containing helium,neon, or argon with halogens added) at a low pressure thatbriefly conducts an electrical charge when a particle or photonof radiation makes the gas conductive by ionization. GMsdisplay the number of ionization events,
22、 typically in “counts-per-second.” The GM tube can detect the presence of radiationbut not its energy, which determines the ionizing effect.Instruments that make use of an energy compensated GM tubeare capable of displaying absorbed dose.3.2.6 ionization chamber, nbeta-gamma radiation meteruseful fo
23、r direct measurement of exposure and dose rates,determining shielding effectiveness, checking sourcecontainers, monitoring radiation areas, and checking resultsfollowing decontamination procedures.3.2.6.1 DiscussionThe unit contains an ion chamber thatconverts directly from ion chamber current to do
24、se rate.3.2.7 manipulator through tube, npenetration providedspecifically for installation of a master-slave manipulator.3.2.7.1 DiscussionTypically, a hole with a diameter of acertain size depending on the manipulator being used thatpasses horizontally though the shield wall above the operators.3.2
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC1831C1831M2017STANDARDGUIDEFORGAMMARADIATIONSHIELDINGPERFORMANCETESTING 辐射 屏蔽 性能 试验 标准 指南 PDF

链接地址:http://www.mydoc123.com/p-465898.html