《ASTM C1742-2010 Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Double Standard Single-Collector Gas Mass Spectrometer Method《用双标准单收集器气体质谱仪法进行六氟化铀同位素分析的标准试验方法.pdf》由会员分享,可在线阅读,更多相关《ASTM C1742-2010 Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Double Standard Single-Collector Gas Mass Spectrometer Method《用双标准单收集器气体质谱仪法进行六氟化铀同位素分析的标准试验方法.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C1742 10Standard Test Method forIsotopic Analysis of Uranium Hexafluoride by DoubleStandard Single-Collector Gas Mass Spectrometer Method1This standard is issued under the fixed designation C1742; the number immediately following the designation indicates the year oforiginal adoption or
2、, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This is a quantitative test method applicable to deter-mining the mass percent
3、of uranium isotopes in uraniumhexafluoride (UF6) samples with235U concentrations between0.1 and 5.0 mass %.1.2 This test method may be applicable for the entire rangeof235U concentrations for which adequate standards areavailable.21.3 This test method is for analysis by a gas magnetic sectormass spe
4、ctrometer with a single collector using interpolation todetermine the isotopic concentration of an unknown samplebetween two characterized UF6standards.1.4 This test method is to replace the existing test methodcurrently published in Test Methods C761 and is used in thenuclear fuel cycle for UF6isot
5、opic analyses.1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to
6、establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3C761 Test Methods for Chemical, Mass Spectrometric,Spectrochemical, Nuclear, and Radiochemical Analysis ofUranium HexafluorideC787 Spe
7、cification for Uranium Hexafluoride for Enrich-mentC996 Specification for Uranium Hexafluoride Enriched toLess Than 5 %235U2.2 Other Document:USEC-651 Uranium Hexafluoride: A Manual of Good Han-dling Practices, Latest Revisions3. Summary of Test Method3.1 The unknown sample and two standards, whose2
8、35Umass percent bracket that of the unknown, are introduced insequence into the mass spectrometer and measurements aremade that are a function of the mol ratio of235U to the total ofthe other isotopes of uranium. These measurements, togetherwith the known composition of the standards, permit calcula
9、-tion of the235U composition of the sample by linear interpo-lation (1-6).3.2 The symmetrical sequence of sample-standard introduc-tions from low standard, sample, high standard, and thenreversed is designed to minimize biases resulting from instru-ment drift, sample interaction or memory, and the n
10、onlinearityof the relationship between the measured resistance ratios andthe true sample mol ratios. Corrections generally are notrequired for instrument memory.4. Significance and Use4.1 Uranium hexafluoride is a basic material used to preparenuclear reactor fuel. To be suitable for this purpose, t
11、hematerial shall meet the criteria for isotopic composition. Thistest method is designed to determine whether the materialmeets the requirements described in Specifications C787 andC996.4.2 ASTM Committee C26 Safeguards Statement:4.2.1 The material (uranium hexafluoride) to which this testmethod app
12、lies is subject to the nuclear safeguards regulationsgoverning its possession and use. The analytical procedure inthis test method has been designated as technically acceptablefor generating safeguards accountability data.4.2.2 When used in conjunction with the appropriate certi-fied reference mater
13、ials (CRMs), this procedure can demon-strate traceability to the national measurement base. However,1This test method is under the jurisdiction of ASTM Committee C26 on NuclearFuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved Oct. 1, 2010.
14、Published November 2010. DOI:10.1520/C174210.2This test method applies to the measurement of235U. It would also be able tomeasure minor isotopes using the same principle subject to appropriate validation bythe measurement laboratory.3For referenced ASTM standards, visit the ASTM website, www.astm.or
15、g, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.adherence to
16、this procedure does not automatically guaranteeregulatory acceptance of the regulatory safeguards measure-ments. It remains the sole responsibility of the user of this testmethod to ensure that its application to safeguards has theapproval of the proper regulatory authorities.5. Apparatus5.1 A mass
17、spectrometer is used with the following featuresand capabilities (2-7):5.1.1 An ion source with an accelerating voltage of approxi-mately 1.75kv.5.1.2 The sample inlet system shall have a minimum ofthree points for attachment of standards and samples, thenecessary valves to evacuate the sample lines
18、 and admit theUF6into the ion source, and a variable leak to control the flowof UF6into the ion source.5.1.3 A single ion collector (8) consisting of a high and lowsignal shall be used. The high-side current signal contains acentral slit, preferably adjustable, to permit passage of greaterthan 90 %
19、of the ions of the235U isotope (mass 330). The235U ions passing through the slit is called the low-side currentsignal.5.1.4 The measuring system shall provide a mole ratiomeasurement calculation from the low- and high-current sig-nal. Such a system usually consists of an amplifier and counterfor eac
20、h V-to-F convertor.5.1.5 The resolving power of the instrument should beshould be less than 5 % between Mass 330/333. This resolutionrequirement should be met with the collector slit widthadjusted to pass at least 90 % of the235U ion beam.5.2 Computer control for opening and closing valves, tim-ing
21、sequence and integrations of signals and data acquisitions.6. Procedure6.1 Prepare Sample and Standards:6.1.1 Select a high and low standard that bracket the235Umass percent of the sample.6.1.2 Attach sample and standard containers to the spec-trometer.6.1.3 Open and close the appropriate valves to
22、evacuate theair from the inlet system.6.1.4 Open the sample and standard containers individuallyto remove impurities that may bias the results or interfere withthe ionization. If necessary, freeze the UF6with ice water or amixture of crushed dry ice and isopropyl alcohol to permitlonger venting with
23、out losing large amounts of UF6.6.1.5 Permit the exhaust system to recover.6.1.6 Check to see if impurities have been sufficientlyremoved by introducing UF6into the ion source and observingpressure.6.1.7 If necessary, repeat 6.1.4 to 6.1.6 until samples havebeen sufficiently purified.6.2 Prepare Ins
24、trument:6.2.1 Ensure the instrument parameters are set so the235UF5+ion beam is focused through the slit on the high-sideplate onto the low side plate.6.2.2 Enter standard values and other information if neededfor calculations performed by the computer.6.2.3 For purposes of programming entry sequenc
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC17422010STANDARDTESTMETHODFORISOTOPICANALYSISOFURANIUMHEXAFLUORIDEBYDOUBLESTANDARDSINGLECOLLECTORGASMASSSPECTROMETERMETHOD

链接地址:http://www.mydoc123.com/p-465725.html