ASTM C1428-1999(2005) Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Single-Standard Gas Source Multiple Collector Mass Spectrometer Method《用单支架气体源多收集器质谱仪法进行.pdf
《ASTM C1428-1999(2005) Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Single-Standard Gas Source Multiple Collector Mass Spectrometer Method《用单支架气体源多收集器质谱仪法进行.pdf》由会员分享,可在线阅读,更多相关《ASTM C1428-1999(2005) Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Single-Standard Gas Source Multiple Collector Mass Spectrometer Method《用单支架气体源多收集器质谱仪法进行.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1428 99 (Reapproved 2005)Standard Test Method forIsotopic Analysis of Uranium Hexafluoride bySingleStandard Gas Source Multiple Collector MassSpectrometer Method1This standard is issued under the fixed designation C 1428; the number immediately following the designation indicates the
2、year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method is applicable to the isotopic anal
3、ysis ofuranium hexafluoride (UF6) with235U concentrations less thanor equal to 5 % and234U,236U concentrations of 0.001 to0.1 %.1.2 This test method may be applicable to the analysis of theentire range of235U isotopic compositions providing thatadequate standards are available.1.3 This standard does
4、 not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety health practices and determine the applicability ofregulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2
5、C 787 Specification for Uranium Hexafluoride for Enrich-mentC 996 Specification for Uranium Hexafluoride Enriched toLess than 5 %235U2.2 Other Document:USEC-651 Uranium Hexafluoride: A Manual of Good Han-dling Practices33. Summary of Test Method3.1 The unknown sample and a standard whose isotopiccom
6、position is close to that of the sample are introduced insequence into the mass spectrometer, and234U,235U,236U,and238U ions are focused through corresponding collector slitsto the four separate collectors. Measurements are made that areproportional to the ratios of234U,235U, or236Uto238U. Withthe k
7、nown composition of the standard, these ratios of molarratios permit calculation of the234U,235U, and236U contents.Memory corrections are applied based on the periodic mea-surement of two standards.4. Significance and Use4.1 Uranium hexafluoride is a basic material used to pro-duce nuclear reactor f
8、uel. To be suitable for this purpose, thematerial must meet criteria for isotopic composition. This testmethod is designed to determine whether the material meetsthe requirements described in Specifications C 787 and C 996.5. Apparatus5.1 A gas source multiple collector mass spectrometer withthe fol
9、lowing attributes:5.1.1 The resolving power of the mass spectrometer is notless than 500. The resolving power (R) is calculated from theregistered mass spectrum of both the235UF5+and238UF5+isotopes as follows:R 5a Mb DM(1)wherea = distance between centers of the235UF5+and238UF5+peaks,b = peak width
10、of the238UF5+isotope (10 % valley),M = 333 mass(u)238UF5+, andDM = 3 = 333 330, 330 mass(u)235UF5+5.1.2 The abundance sensitivity of the mass spectrometer isspecified as less than 13105as contribution of mass 333(238UF5+) to mass 331 (236UF5+).5.1.3 The four collectors have collector slits adjusted
11、forions of masses 329, 330, 331, and 333. Ion currents areamplified by four amplifiers, having noise level less than 0.5mV.5.1.4 The ion beams are kept within the slits by an auto-matic beam positioner circuit.1This test method is under the jurisdiction ofASTM Committee C26 on NuclearFuel Cycle and
12、is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved June 1, 2005. Published December 2005. Originallyapproved in 1999. Last previous edition approved in 1999 as C 142899.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Cu
13、stomer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from Superintendent of Documents, U.S. Government PrintingOffice, Washington, D.C. 20402.1Copyright ASTM International, 100 Barr Harbor D
14、rive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.1.5 The pumping system of the mass spectrometer analy-ser tube shall maintain a pressure less than 53108torr with asample flowing into the ion source.5.1.6 The memory correction factor of the mass spectrom-eter as defined in 9.1 sh
15、all be consistent with the requiredaccuracy and precision, and shall not exceed 1.01.5.1.7 The sample inlet system shall be equipped with amanifold, including adjustable leak, and valves for introducingthe sample and standard in sequence and for evacuatingcorresponding lines. The pumping system of t
16、he inlet systemmust maintain a pressure less than 23102torr by evacuating.6. Hazards6.1 Since UF6is radioactive, toxic, and highly reactive,especially with reducing substances and moisture (see USEC-651), appropriate facilities and practices for analysis shall beprovided.7. Calibration and Standardi
17、zation7.1 Uranium Hexafluoride Isotopic7.1.1 Two standards are used to determine the memorycorrection factor. The relative235U concentration difference ofthese standards shall not be more than 20 %.7.1.2 For memory correction factor determination for234Uand236U isotopes in concentration range 0.01 t
18、o 0.1 %, twostandards are used which differ in234U(236U) concentration.The relative difference shall not exceed two times; in concen-tration range 0.001 to 0.01 %, the relative difference shall notexceed four times.7.1.3 The standards used for measurements may differin235U concentration from a sampl
19、e, but the relative differenceshall not be more than 10 %. The relative difference in234U(236U) concentration shall not exceed two times for theconcentration range 0.001 to 0.01 %, and shall not be morethan 50 % for the concentration range 0.01 to 0.1 %.8. ProcedureA typical sequence for the analyti
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC142819992005STANDARDTESTMETHODFORISOTOPICANALYSISOFURANIUMHEXAFLUORIDEBYSINGLESTANDARDGASSOURCEMULTIPLECOLLECTORMASSSPECTROMETERMETHOD

链接地址:http://www.mydoc123.com/p-464614.html