ASTM C1380-2004 Standard Test Method for Determination of Uranium Content and Isotopic Composition by Isotope Dilution Mass Spectrometry《用同位素稀释质谱法测定铀含量和同位素成分的标准试验方法》.pdf
《ASTM C1380-2004 Standard Test Method for Determination of Uranium Content and Isotopic Composition by Isotope Dilution Mass Spectrometry《用同位素稀释质谱法测定铀含量和同位素成分的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM C1380-2004 Standard Test Method for Determination of Uranium Content and Isotopic Composition by Isotope Dilution Mass Spectrometry《用同位素稀释质谱法测定铀含量和同位素成分的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1380 04Standard Test Method forthe Determination of Uranium Content and IsotopicComposition by Isotope Dilution Mass Spectrometry1This standard is issued under the fixed designation C 1380; the number immediately following the designation indicates the year oforiginal adoption or, in
2、the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers a method for the determinationof the uranium concentration
3、in uranium oxides by isotopedilution mass spectrometry (IDMS). The isotopic compositionof the oxide is measured simultaneously.1.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish a
4、ppro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:D 1193 Specification for Reagent Water23. Summary of Test Method3.1 For measurement of the elemental uranium concentra-tion of uranium oxides by I
5、DMS, a representative and accu-rately measured aliquot of the sample is prepared. A knownquantity of233U (“spike”) is added to an aliquot of the sample.The sample aliquot and spike are taken to dryness, redissolvedin dilute nitric acid, and loaded on a filament for analysis in athermal ionization ma
6、ss spectrometer (TIMS). After measure-ment of the isotopic ratios in the spiked sample, the uraniumcontent and isotopic composition of sample are calculated.4. Significance and Use4.1 Determination of percent uranium content and235Uabundance in oxides and other materials containing highconcentration
7、s of uranium is required for special nuclearmaterials accountability, regulatory requirements, and processcontrol.5. Interferences5.1 The calculations assume any233U in the sample isnegligible. If the sample contains significant233U, the samplemust be analyzed for isotopic composition with and witho
8、utadded spike, and the calculations adjusted accordingly.6. Apparatus6.1 Thermal ionization mass spectrometer (TIMS) config-ured with Faraday cup detectors and an automated operatingsystem.6.2 Preconditioning unit for the TIMS.6.3 Filament loading assembly for the TIMS.6.4 Balance, analytical, with
9、five-place range.6.5 Vials, glass, disposable with plastic caps.6.6 Pipet, automatic, Ranin or equivalent, variable to 1000L.6.7 Pipet tips, disposable plastic, 1001000 L.6.8 Liquid dispenser, Repipettey or equivalent.7. Reagents and Materials7.1 Purity of MaterialsReagent grade chemicals shall beus
10、ed in all tests. Unless otherwise indicated, it is intended thatall reagents conform to the specification of the Committee onAnalytical Reagents of the American Chemical Society wheresuch specifications are available.3Other grades may be usedprovided it is first ascertained that the reagent is of su
11、fficientlyhigh purity to permit its use without lessening the accuracy ofthe determination.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall mean reagent water in conformance withSpecification D 1193.7.3 Nitric Acid, HNO3, concentrated (70 %).7.4 Nitric Acid, 0.8 M (5 % v/v)Cau
12、tiously add 50 mL ofconcentrated nitric acid to 950 mL of water.1This test method is under the jurisdiction of ASTM Committee C26 on NuclearFuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved June 1, 2004. Published July 2004. Originally appr
13、ovedin 1997. Last previous edition approved in 1997 as C 1380 97.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM webs
14、ite.3Reagent Chemicals, American Chemical Society Specifications, AmericanChemical Society, Washington, D. C. For suggestions on the testing of reagents notlisted by the American Chemical Society, Washington, D. C. For suggestions on thetesting of reagents not listed by the American Chemical Society
15、, see AnalarStandards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U. K., and theUnited States Pharmacopeia and National Formulary, U.S. Pharmaceutical Con-vention (USPC), Rockville, MD.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United
16、States.7.5 Nitric Acid, 0.1 MAdd 6.5 mL concentrated nitric acidto 900 mL of water, mix, and bring to 1000 mL with water.7.6 Hydrogen Peroxide H2O2,30%.7.7 Elemental and Isotopic Uranium Standards (New Brun-swick Laboratory CRM 114, CRM 116, CRM 129, or equiva-lent4).7.8 Uranium-233 spike assay and
17、isotopic standard (NBLCRM 111A4or equivalent)Dilute NBL CRM 111A 1:50by weight with a 5 % nitric acid to give a233U concentration ofapproximately 10 g233U/g solution.NOTE 1An exact 1:50 dilution is not required. The requirement is forprecise and accurate weights of standard and diluent. Calculate th
18、e exactconcentration of each isotope in the diluted spike standard in accordancewith 11.2. Other dilutions of CRM 111-A may be used if accurate weightsare known, and the aliquot of CRM 111A used in the measurement of testsamples yields a233U/23XU ratio of at least 0.02, where X is the majoruranium i
19、sotope.8. Preparation of Apparatus8.1 Prepare the thermal ionization mass spectrometer(TIMS) in accordance with manufacturers recommendations.9. Calibration and Standardization9.1 Standardization of233U Spike Solution:9.1.1 Prior to using a new diluted spike solution, verify theconcentration of the
20、solution with CRM 129 uranium oxide orits equivalent. If the new spike solution does not give resultswithin control limits, its standardized concentration must beverified with another NBL certified reference material forelemental uranium, such as CRM 114 or equivalent.9.2 Calibration of TIMS:9.2.1 C
21、alibrate the TIMS in accordance with manufacturersrecommendations to achieve the users performance and qual-ity assurance criteria.10. Procedure10.1 Weigh, to the nearest 0.01 mg, a labeled vial and capfor each sample; record the weight.10.2 Tare the balance and weigh, to the nearest 0.01 mg, a0.350
22、.45 g aliquot of the sample into the vial. Record theweight.NOTE 2Smaller or larger aliquots of sample may be used. However,the sample size should be sufficient to obtain a representative sample ofthe material and the ratio of the measured233U to major uranium isotopein the material must be greater
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC13802004STANDARDTESTMETHODFORDETERMINATIONOFURANIUMCONTENTANDISOTOPICCOMPOSITIONBYISOTOPEDILUTIONMASSSPECTROMETRY

链接地址:http://www.mydoc123.com/p-464427.html