ASTM C1379-2004 Standard Test Method for Analysis of Urine for Uranium-235 and Uranium-238 Isotopes by Inductively Coupled Plasma-Mass Spectrometry《用感应耦合等离子体质谱法分析铀235和铀238同位素用尿的标准试.pdf
《ASTM C1379-2004 Standard Test Method for Analysis of Urine for Uranium-235 and Uranium-238 Isotopes by Inductively Coupled Plasma-Mass Spectrometry《用感应耦合等离子体质谱法分析铀235和铀238同位素用尿的标准试.pdf》由会员分享,可在线阅读,更多相关《ASTM C1379-2004 Standard Test Method for Analysis of Urine for Uranium-235 and Uranium-238 Isotopes by Inductively Coupled Plasma-Mass Spectrometry《用感应耦合等离子体质谱法分析铀235和铀238同位素用尿的标准试.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1379 04Standard Test Method forAnalysis of Urine for Uranium-235 and Uranium-238Isotopes by Inductively Coupled Plasma-MassSpectrometry1This standard is issued under the fixed designation C 1379; the number immediately following the designation indicates the year oforiginal adoption o
2、r, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of the con-centration of uranium-23
3、5 and uranium-238 in urine usingInductively Coupled Plasma-Mass Spectrometry. This testmethod can be used to support uranium facility bioassayprograms.1.2 This method detection limits for235U and238U are 6ng/L. To meet the requirements of ANSI N13.30, the minimumdetectable activity (MDA) of each rad
4、ionuclide measured mustbe at least 0.1 pCi/L (0.0037 Bq/L). The MDA translates to 47ng/L for235U and 300 ng/L for238U. Uranium 234 cannot bedetermined at the MDA with this test method because of its lowmass concentration level equivalent to 0.1 pCi/L.1.3 The digestion and anion separation of urine m
5、ay not benecessary when uranium concentrations of more than 100 ng/Lare present.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and det
6、ermine the applica-bility of regulatory limitations prior to use.NOTE 1Warning: The ICP-MS is a source of intense ultravioletradiation from the radio frequency induced plasma. Protection from radiofrequency radiation and UV radiation is provided by the instrument undernormal operation.2. Referenced
7、Documents2.1 ASTM Standards:2C 1310 Test Method for Determining Radionuclides inSoils by Inductively Coupled Plasma-Mass SpectrometryUsing Flow Injection PreconcentrationC 1345 Test Method for Analysis of Total and IsotopicUranium and Total Thorium in Soils by InductivelyCoupled Plasma-Mass Spectrom
8、etryD 1193 Specification for Reagent Water2.2 Other Documents:ANSI N13.30 Radiological Measurement QualityDOE Order 5480.11 Radiological Measurements Quality3. Terminology3.1 Definitions:3.1.1 isobar, nany atom that has the same atomic massnumber as another atom but a different atomic number.3.2 Acr
9、onyms:Acronyms:3.2.1 AMU, natomic mass unit3.2.2 CB, ncalibration blank3.2.3 COC, nchain of custody3.2.4 CVS, ncalibration verification standard3.2.5 ICS, ninstrument check standard3.2.6 IDL, ninstrument detection limit3.2.7 LCS, nlaboratory control sample3.2.8 MDA, nminimum detectable activity3.2.9
10、 m/e, nmass/charge ratio3.2.10 RMDA, nrequired minimum detectable activity3.2.11 % RDS, npercent relative standard deviation(1Standard Deviation / Mean) * 1004. Summary of Test Method4.1 A urine sample is digested and wet oxidized withconcentrated nitric and hydrochloric acids to solubilize ura-nium
11、 and to destroy the organic matter. Uranium is selectivelyseparated from the chloride salts by an anion exchange resinand is eluted with dilute nitric acid. The235U and238U isotopesare determined by ICP-MS.5. Significance and Use5.1 DOE Order 5480.11 and ANSI N13.30 require thatinternal dose assessm
12、ents be made as part of the bioassayprogram for nuclear facility workers. For indirect bioassay ofuranium workers, the uranium isotopes must be measured1This test method is under the jurisdiction of ASTM Committee C26 on NuclearFuel Cycle and is the direct responsibility of Subcommittee D26.05 on Me
13、thods ofTest.Current edition approved Jan. 1, 2004. Published February 2004. Originallyapproved in 1997. Last previous edition approved in 1997 as C137997.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTM
14、Standardsvolume information, refer to the standards Document Summary page onthe ASTM website1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.along with the total uranium in urine samples. The RMDA foreach uranium isotope is 0.1 pCi/L.
15、5.2 This method is applicable for measuring235U and238Uat the RMDA. Because of extremely low mass concentration(because of the high specific activity),234U cannot be measuredwithout additional sample preconcentration.NOTE 2Column chromatography separations and concentration of234U using manual or fl
16、ow-injection preconcentration followed byICP-MS isotopic determination are described in Test Methods C 1310 andC 1345. These methods focus on environmental soil sample analysis, butwith some development, may be applicable to digested urine samples. The234U concentration can be calculated based on an
17、 enrichment gradient forworkers in uranium enrichment plants, and internal dose assessments canbe made.NOTE 3Use of high resolution ICP-MS may also be used to obtainlower detection limits, see 1.6. Interferences6.1 No known isobaric elemental interferences occur fordetermining235U and238U using this
18、 test method.NOTE 4Bismuth, such as found in antacids (for example, PeptoBismol) may interferes with the analysis by using binding sites on theresin or biasing the internal standard measurement on the ICP-MSanalysis.7. Apparatus7.1 Inductively Coupled Plasma-Mass Spectrometer,computer-controlled, mu
19、lti-channel peristaltic pump, and anautosampler (2, 3, 4).7.2 Appropriate sized disposable graduated test tube withcap that will accommodate the autosampler.7.3 Twelve-mL disposable polyethylene column or suitablesize with frit.7.4 Vacuum manifold chamber with regulator valve,vacuum gage, vacuum rel
20、ief valve, and a vacuum manifoldbeaker rack (optional).8. Reagents and Materials8.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents conform to the specifications of the Committee onAnalytical Reagents of the American Ch
21、emical Society wheresuch specifications are available.3Other grades of reagentsmay be used, provided it is first ascertained that the reagent isof sufficiently high purity to permit its use without lesseningthe accuracy of the determination.8.2 Purity of WaterUnless otherwise indicated, referencesto
22、 water shall be understood to mean reagent water, as ASTM,Type I water as defined in Specification D 1193.8.3 Nitric Acid (sp gr 1.42)concentrated nitric acid(HNO3).8.4 Hydrochloric Acid (sp gr 1.18)concentrated hydro-chloric acid (HCl).8.5 Nitric Acid (8M)Add 500 mL of concentrated HNO3to 500 mL of
23、 water, and mix.8.6 Nitric Acid (0.8 M)Add 50 mL of concentrated HNO3to 950 mL of water, and mix.8.7 Nitric Acid (0.01 M)Add 1.25 mL of 8M HNO3to950 mL of water, and dilute to a final volume of 1000 mL, andmix.8.8 Hydrochloric Acid (6 M)Add 500 mL of concentratedHCl to 500 mL of water, and mix.8.9 A
24、rgon Gaspurity 99.99 % or better recommended.8.10 Standard Metals Stock Solutions Prepare or purchasecertified traceable stock or equivalent certified solutions ofberyllium, bismuth, cobalt, indium, lanthanum, lead, magne-sium, and uranium to be used for the tuning solution, calibra-tion, spiking, m
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC13792004STANDARDTESTMETHODFORANALYSISOFURINEFORURANIUM235ANDURANIUM238ISOTOPESBYINDUCTIVELYCOUPLEDPLASMAMASSSPECTROMETRY

链接地址:http://www.mydoc123.com/p-464411.html